Bismuth shield affecting CT image quality and radiation dose in adjacent and distant zones relative to shielding surface: A phantom study
BACKGROUND: To quantify image quality and radiation doses in regions adjacent to and distant from bismuth shields in computed tomography (CT). METHODS: An American College of Radiology accreditation phantom with four solid rods embedded in a water-like background was scanned to verify CT number (CTN...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Chang Gung University
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6888723/ https://www.ncbi.nlm.nih.gov/pubmed/31783995 http://dx.doi.org/10.1016/j.bj.2019.04.004 |
Sumario: | BACKGROUND: To quantify image quality and radiation doses in regions adjacent to and distant from bismuth shields in computed tomography (CT). METHODS: An American College of Radiology accreditation phantom with four solid rods embedded in a water-like background was scanned to verify CT number (CTN) accuracy when using bismuth shields. CTNs, image noise, and contrast-to-noise ratios (CNRs) were determined in the phantom at 80–140 kVp. Image quality was investigated on image portions in the zones adjacent (A zone) to and distant (D zone) from a bismuth shield. Surface radiation doses were measured using thermoluminescent dosimeters. Streak artefacts were graded on a 3-point-scale. RESULTS: Changes in CTN caused by a bismuth shield resulted in changes in X-ray spectra. CTN changes were more apparent in the A zone than in the D zone, particularly for a low tube voltage. The degrees of CTN changes and image noise were proportional to the thickness of the bismuth shields. A 1-ply bismuth shield reduced surface radiation doses by 7.2%–15.5%. The overall CNRs were slightly degraded, and streak artefacts were acceptable. CONCLUSIONS: Using a bismuth shield could result in significant CTN changes and perceivable artefacts, particularly for a superficial organ close to the shield, and is not recommended for quantification CT examinations or follow-up CT examinations. |
---|