Correcting Artifacts in Single Molecule Localization Microscopy Analysis Arising from Pixel Quantum Efficiency Differences in sCMOS Cameras
Optimal analysis of single molecule localization microscopy (SMLM) data acquired with a scientific Complementary Metal-Oxide-Semiconductor (sCMOS) camera relies on statistical compensation for its pixel-dependent gain, offset and readout noise. In this work we show that it is also necessary to compe...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Nature Publishing Group UK
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6889274/ https://www.ncbi.nlm.nih.gov/pubmed/31792238 http://dx.doi.org/10.1038/s41598-019-53698-x |
Sumario: | Optimal analysis of single molecule localization microscopy (SMLM) data acquired with a scientific Complementary Metal-Oxide-Semiconductor (sCMOS) camera relies on statistical compensation for its pixel-dependent gain, offset and readout noise. In this work we show that it is also necessary to compensate for differences in the relative quantum efficiency (RQE) of each pixel. We found differences in RQE on the order of 4% in our tested sCMOS sensors. These differences were large enough to have a noticeable effect on analysis algorithm results, as seen both in simulations and biological imaging data. We discuss how the RQE differences manifest themselves in the analysis results and present the modifications to the Poisson maximum likelihood estimation (MLE) sCMOS analysis algorithm that are needed to correct for the RQE differences. |
---|