Micromechanical modeling of the contact stiffness of an osseointegrated bone–implant interface

BACKGROUND: The surgical success of cementless implants is determined by the evolution of the biomechanical properties of the bone–implant interface (BII). One difficulty to model the biomechanical behavior of the BII comes from the implant surface roughness and from the partial contact between bone...

Descripción completa

Detalles Bibliográficos
Autores principales: Raffa, Maria Letizia, Nguyen, Vu-Hieu, Haiat, Guillaume
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6889538/
https://www.ncbi.nlm.nih.gov/pubmed/31796076
http://dx.doi.org/10.1186/s12938-019-0733-3
_version_ 1783475439027617792
author Raffa, Maria Letizia
Nguyen, Vu-Hieu
Haiat, Guillaume
author_facet Raffa, Maria Letizia
Nguyen, Vu-Hieu
Haiat, Guillaume
author_sort Raffa, Maria Letizia
collection PubMed
description BACKGROUND: The surgical success of cementless implants is determined by the evolution of the biomechanical properties of the bone–implant interface (BII). One difficulty to model the biomechanical behavior of the BII comes from the implant surface roughness and from the partial contact between bone tissue and the implant. The determination of the constitutive law of the BII would be of interest in the context of implant finite element (FE) modeling to take into account the imperfect characteristics of the BII. The aim of the present study is to determine an effective contact stiffness [Formula: see text] of an osseointegrated BII accounting for its micromechanical features such as surface roughness, bone–implant contact ratio (BIC) and periprosthetic bone properties. To do so, a 2D FE model of the BII under normal contact conditions was developed and was used to determine the behavior of [Formula: see text] . RESULTS: The model is validated by comparison with three analytical schemes based on micromechanical homogenization including two Lekesiz’s models (considering interacting and non-interacting micro-cracks) and a Kachanov’s model. [Formula: see text] is found to be comprised between 10(13) and 10(15) N/m(3) according to the properties of the BII. [Formula: see text] is shown to increase nonlinearly as a function of the BIC and to decrease as a function of the roughness amplitude for high BIC values (above around 20%). Moreover, [Formula: see text] decreases as a function of the roughness wavelength and increases linearly as a function of the Young’s modulus of periprosthetic bone tissue. CONCLUSIONS: These results open new paths in implant biomechanical modeling since this model may be used in future macroscopic finite element models modeling the bone–implant system to replace perfectly rigid BII conditions. [Image: see text]
format Online
Article
Text
id pubmed-6889538
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-68895382019-12-11 Micromechanical modeling of the contact stiffness of an osseointegrated bone–implant interface Raffa, Maria Letizia Nguyen, Vu-Hieu Haiat, Guillaume Biomed Eng Online Research BACKGROUND: The surgical success of cementless implants is determined by the evolution of the biomechanical properties of the bone–implant interface (BII). One difficulty to model the biomechanical behavior of the BII comes from the implant surface roughness and from the partial contact between bone tissue and the implant. The determination of the constitutive law of the BII would be of interest in the context of implant finite element (FE) modeling to take into account the imperfect characteristics of the BII. The aim of the present study is to determine an effective contact stiffness [Formula: see text] of an osseointegrated BII accounting for its micromechanical features such as surface roughness, bone–implant contact ratio (BIC) and periprosthetic bone properties. To do so, a 2D FE model of the BII under normal contact conditions was developed and was used to determine the behavior of [Formula: see text] . RESULTS: The model is validated by comparison with three analytical schemes based on micromechanical homogenization including two Lekesiz’s models (considering interacting and non-interacting micro-cracks) and a Kachanov’s model. [Formula: see text] is found to be comprised between 10(13) and 10(15) N/m(3) according to the properties of the BII. [Formula: see text] is shown to increase nonlinearly as a function of the BIC and to decrease as a function of the roughness amplitude for high BIC values (above around 20%). Moreover, [Formula: see text] decreases as a function of the roughness wavelength and increases linearly as a function of the Young’s modulus of periprosthetic bone tissue. CONCLUSIONS: These results open new paths in implant biomechanical modeling since this model may be used in future macroscopic finite element models modeling the bone–implant system to replace perfectly rigid BII conditions. [Image: see text] BioMed Central 2019-12-03 /pmc/articles/PMC6889538/ /pubmed/31796076 http://dx.doi.org/10.1186/s12938-019-0733-3 Text en © The Author(s) 2019 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research
Raffa, Maria Letizia
Nguyen, Vu-Hieu
Haiat, Guillaume
Micromechanical modeling of the contact stiffness of an osseointegrated bone–implant interface
title Micromechanical modeling of the contact stiffness of an osseointegrated bone–implant interface
title_full Micromechanical modeling of the contact stiffness of an osseointegrated bone–implant interface
title_fullStr Micromechanical modeling of the contact stiffness of an osseointegrated bone–implant interface
title_full_unstemmed Micromechanical modeling of the contact stiffness of an osseointegrated bone–implant interface
title_short Micromechanical modeling of the contact stiffness of an osseointegrated bone–implant interface
title_sort micromechanical modeling of the contact stiffness of an osseointegrated bone–implant interface
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6889538/
https://www.ncbi.nlm.nih.gov/pubmed/31796076
http://dx.doi.org/10.1186/s12938-019-0733-3
work_keys_str_mv AT raffamarialetizia micromechanicalmodelingofthecontactstiffnessofanosseointegratedboneimplantinterface
AT nguyenvuhieu micromechanicalmodelingofthecontactstiffnessofanosseointegratedboneimplantinterface
AT haiatguillaume micromechanicalmodelingofthecontactstiffnessofanosseointegratedboneimplantinterface