Cargando…

Use of three points to determine the accuracy of guided implantation

AIM: 'This study aims to establish an open-source algorithm using Python to analyze the accuracy of guided implantation, which simplifies interstudy comparisons. METHODS: Given ≥3 landmark pairs, this Tri-Point (TriP) method can register images. With ≥4 landmark pairs, TriP can calculate system...

Descripción completa

Detalles Bibliográficos
Autores principales: Liang, Ye, Yuan, ShanShan, Huan, JingJing, Wang, HuiXin, Zhang, YiYi, Fang, ChangYun, Li, Jia-Da
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6890217/
https://www.ncbi.nlm.nih.gov/pubmed/31794595
http://dx.doi.org/10.1371/journal.pone.0225823
_version_ 1783475567303065600
author Liang, Ye
Yuan, ShanShan
Huan, JingJing
Wang, HuiXin
Zhang, YiYi
Fang, ChangYun
Li, Jia-Da
author_facet Liang, Ye
Yuan, ShanShan
Huan, JingJing
Wang, HuiXin
Zhang, YiYi
Fang, ChangYun
Li, Jia-Da
author_sort Liang, Ye
collection PubMed
description AIM: 'This study aims to establish an open-source algorithm using Python to analyze the accuracy of guided implantation, which simplifies interstudy comparisons. METHODS: Given ≥3 landmark pairs, this Tri-Point (TriP) method can register images. With ≥4 landmark pairs, TriP can calculate system errors for image registration. We selected 8 indicators from the literature. Considering development errors in new bone on cone beam computed tomography (CBCT), we added the indicators of apical rectified deviation (ARD) and coronal rectified deviation (CRD), providing accurate references but neglecting depth deviations. Our program can calculate and output these indicators. To evaluate the TriP method’s feasibility, an implantation group assisted by a Visual Direction-INdicating Guide (VDING) was analyzed. Accuracy was measured with the traditional and proposed TriP methods. Factors affecting the system error of the method were then analyzed. RESULTS: Comparisons with paired-sample t-tests showed that our TriP method was similar to the traditional method in evaluating implantation accuracy, with no significant difference (P>0.05). The average system error was 0.30±0.10 mm when the TriP method evaluated the VDING template. The results showed that increasing the provided landmarks from 4 to 5 pairs decreased the between-group differences significantly (P<0.05). With ≥6 pairs of landmarks, the system error tended to be stable, and the groups showed no statistically significant differences (P>0.05). Large distances between landmarks are helpful in reducing system error, as demonstrated with a geometric method. CONCLUSIONS: This study established an open-source algorithm to analyze the accuracy of guided implantation with system errors reported.
format Online
Article
Text
id pubmed-6890217
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-68902172019-12-13 Use of three points to determine the accuracy of guided implantation Liang, Ye Yuan, ShanShan Huan, JingJing Wang, HuiXin Zhang, YiYi Fang, ChangYun Li, Jia-Da PLoS One Research Article AIM: 'This study aims to establish an open-source algorithm using Python to analyze the accuracy of guided implantation, which simplifies interstudy comparisons. METHODS: Given ≥3 landmark pairs, this Tri-Point (TriP) method can register images. With ≥4 landmark pairs, TriP can calculate system errors for image registration. We selected 8 indicators from the literature. Considering development errors in new bone on cone beam computed tomography (CBCT), we added the indicators of apical rectified deviation (ARD) and coronal rectified deviation (CRD), providing accurate references but neglecting depth deviations. Our program can calculate and output these indicators. To evaluate the TriP method’s feasibility, an implantation group assisted by a Visual Direction-INdicating Guide (VDING) was analyzed. Accuracy was measured with the traditional and proposed TriP methods. Factors affecting the system error of the method were then analyzed. RESULTS: Comparisons with paired-sample t-tests showed that our TriP method was similar to the traditional method in evaluating implantation accuracy, with no significant difference (P>0.05). The average system error was 0.30±0.10 mm when the TriP method evaluated the VDING template. The results showed that increasing the provided landmarks from 4 to 5 pairs decreased the between-group differences significantly (P<0.05). With ≥6 pairs of landmarks, the system error tended to be stable, and the groups showed no statistically significant differences (P>0.05). Large distances between landmarks are helpful in reducing system error, as demonstrated with a geometric method. CONCLUSIONS: This study established an open-source algorithm to analyze the accuracy of guided implantation with system errors reported. Public Library of Science 2019-12-03 /pmc/articles/PMC6890217/ /pubmed/31794595 http://dx.doi.org/10.1371/journal.pone.0225823 Text en © 2019 Liang et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Research Article
Liang, Ye
Yuan, ShanShan
Huan, JingJing
Wang, HuiXin
Zhang, YiYi
Fang, ChangYun
Li, Jia-Da
Use of three points to determine the accuracy of guided implantation
title Use of three points to determine the accuracy of guided implantation
title_full Use of three points to determine the accuracy of guided implantation
title_fullStr Use of three points to determine the accuracy of guided implantation
title_full_unstemmed Use of three points to determine the accuracy of guided implantation
title_short Use of three points to determine the accuracy of guided implantation
title_sort use of three points to determine the accuracy of guided implantation
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6890217/
https://www.ncbi.nlm.nih.gov/pubmed/31794595
http://dx.doi.org/10.1371/journal.pone.0225823
work_keys_str_mv AT liangye useofthreepointstodeterminetheaccuracyofguidedimplantation
AT yuanshanshan useofthreepointstodeterminetheaccuracyofguidedimplantation
AT huanjingjing useofthreepointstodeterminetheaccuracyofguidedimplantation
AT wanghuixin useofthreepointstodeterminetheaccuracyofguidedimplantation
AT zhangyiyi useofthreepointstodeterminetheaccuracyofguidedimplantation
AT fangchangyun useofthreepointstodeterminetheaccuracyofguidedimplantation
AT lijiada useofthreepointstodeterminetheaccuracyofguidedimplantation