Design and Validation of a Holographic Particle Counter
An in-line holographic particle counter concept is presented and validated where multiple micrometer sized particles are detected in a three dimensional sampling volume, all at once. The proposed Particle Imaging Unit is capable of detecting holograms of particles which sizes are in the lower [Formu...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6891361/ https://www.ncbi.nlm.nih.gov/pubmed/31717587 http://dx.doi.org/10.3390/s19224899 |
_version_ | 1783475795584352256 |
---|---|
author | Brunnhofer, Georg Bergmann, Alexander Klug, Andreas Kraft, Martin |
author_facet | Brunnhofer, Georg Bergmann, Alexander Klug, Andreas Kraft, Martin |
author_sort | Brunnhofer, Georg |
collection | PubMed |
description | An in-line holographic particle counter concept is presented and validated where multiple micrometer sized particles are detected in a three dimensional sampling volume, all at once. The proposed Particle Imaging Unit is capable of detecting holograms of particles which sizes are in the lower [Formula: see text] m- range. The detection and counting principle is based on common image processing techniques using a customized Hough Transform with a result directly relating to the particle number concentration in the recorded sampling volume. The proposed counting unit is mounted ontop of a Condensation Nucleus Magnifier for comparison with a commercial TSI-3775 Condensation Particle Counter (CPC). The concept does not only allow for a precise in-situ determination of low particle number concentrations but also enables easy upscaling to higher particle densities (e.g., [Formula: see text]) through its linear expandability and option of cascading. The impact of coincidence at higher particle densities is shown and two coincidence correction approaches are presented where, at last, its analogy to the coincidence correction methods used in state-of-the-art CPCs is identified. |
format | Online Article Text |
id | pubmed-6891361 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | MDPI |
record_format | MEDLINE/PubMed |
spelling | pubmed-68913612019-12-12 Design and Validation of a Holographic Particle Counter Brunnhofer, Georg Bergmann, Alexander Klug, Andreas Kraft, Martin Sensors (Basel) Article An in-line holographic particle counter concept is presented and validated where multiple micrometer sized particles are detected in a three dimensional sampling volume, all at once. The proposed Particle Imaging Unit is capable of detecting holograms of particles which sizes are in the lower [Formula: see text] m- range. The detection and counting principle is based on common image processing techniques using a customized Hough Transform with a result directly relating to the particle number concentration in the recorded sampling volume. The proposed counting unit is mounted ontop of a Condensation Nucleus Magnifier for comparison with a commercial TSI-3775 Condensation Particle Counter (CPC). The concept does not only allow for a precise in-situ determination of low particle number concentrations but also enables easy upscaling to higher particle densities (e.g., [Formula: see text]) through its linear expandability and option of cascading. The impact of coincidence at higher particle densities is shown and two coincidence correction approaches are presented where, at last, its analogy to the coincidence correction methods used in state-of-the-art CPCs is identified. MDPI 2019-11-09 /pmc/articles/PMC6891361/ /pubmed/31717587 http://dx.doi.org/10.3390/s19224899 Text en © 2019 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Brunnhofer, Georg Bergmann, Alexander Klug, Andreas Kraft, Martin Design and Validation of a Holographic Particle Counter |
title | Design and Validation of a Holographic Particle Counter |
title_full | Design and Validation of a Holographic Particle Counter |
title_fullStr | Design and Validation of a Holographic Particle Counter |
title_full_unstemmed | Design and Validation of a Holographic Particle Counter |
title_short | Design and Validation of a Holographic Particle Counter |
title_sort | design and validation of a holographic particle counter |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6891361/ https://www.ncbi.nlm.nih.gov/pubmed/31717587 http://dx.doi.org/10.3390/s19224899 |
work_keys_str_mv | AT brunnhofergeorg designandvalidationofaholographicparticlecounter AT bergmannalexander designandvalidationofaholographicparticlecounter AT klugandreas designandvalidationofaholographicparticlecounter AT kraftmartin designandvalidationofaholographicparticlecounter |