Cargando…

Hemocompatibility and safety of the Carmat Total Artifical Heart hybrid membrane

The Carmat bioprosthetic total artificial heart (C-TAH) is a biventricular pump developed to minimize drawbacks of current mechanical assist devices and improve quality of life during support. This study aims to evaluate the safety of the hybrid membrane, which plays a pivotal role in this artificia...

Descripción completa

Detalles Bibliográficos
Autores principales: Richez, Ulysse, De Castilla, Hector, Guerin, Coralie L., Gendron, Nicolas, Luraghi, Giulia, Grimme, Marc, Wu, Wei, Taverna, Myriam, Jansen, Piet, Latremouille, Christian, Migliavacca, Francesco, Dubini, Gabriele, Capel, Antoine, Carpentier, Alain, Smadja, David M.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6906674/
https://www.ncbi.nlm.nih.gov/pubmed/31867454
http://dx.doi.org/10.1016/j.heliyon.2019.e02914
_version_ 1783478392362893312
author Richez, Ulysse
De Castilla, Hector
Guerin, Coralie L.
Gendron, Nicolas
Luraghi, Giulia
Grimme, Marc
Wu, Wei
Taverna, Myriam
Jansen, Piet
Latremouille, Christian
Migliavacca, Francesco
Dubini, Gabriele
Capel, Antoine
Carpentier, Alain
Smadja, David M.
author_facet Richez, Ulysse
De Castilla, Hector
Guerin, Coralie L.
Gendron, Nicolas
Luraghi, Giulia
Grimme, Marc
Wu, Wei
Taverna, Myriam
Jansen, Piet
Latremouille, Christian
Migliavacca, Francesco
Dubini, Gabriele
Capel, Antoine
Carpentier, Alain
Smadja, David M.
author_sort Richez, Ulysse
collection PubMed
description The Carmat bioprosthetic total artificial heart (C-TAH) is a biventricular pump developed to minimize drawbacks of current mechanical assist devices and improve quality of life during support. This study aims to evaluate the safety of the hybrid membrane, which plays a pivotal role in this artificial heart. We investigated in particular its blood-contacting surface layer of bovine pericardial tissue, in terms of mechanical aging, risks of calcification, and impact of the hemodynamics shear stress inside the ventricles on blood components. Hybrid membranes were aged in a custom-designed endurance bench. Mechanical, physical and chemical properties were not significantly modified from 9 months up to 4 years of aging using a simulating process. Exploration of erosion areas did not show no risk of oil diffusion through the membrane. Blood contacting materials in the ventricular cavities were subcutaneously implanted in Wistar rats for 30 days as a model for calcification and demonstrated that the in-house anti-calcification pretreatment with Formaldehyde-Ethanol-Tween 80 was able to significantly reduce the calcium concentration from 132 μg/mg to 4.42 μg/mg (p < 0.001). Hemodynamic simulations with a computational model were used to reproduce shear stress in left and right ventricles and no significant stress was able to trigger hemolysis, platelet activation nor degradation of the von Willebrand factor multimers. Moreover, explanted hybrid membranes from patients included in the feasibility clinical study were analyzed confirming preclinical results with the absence of significant membrane calcification. At last, blood plasma bank analysis from the four patients implanted with C-TAH during the feasibility study showed no residual glutaraldehyde increase in plasma and confirmed hemodynamic simulation-based results with the absence of hemolysis and platelet activation associated with normal levels of plasma free hemoglobin and platelet microparticles after C-TAH implantation. These results on mechanical aging, calcification model and hemodynamic simulations predicted the safety of the hybrid membrane used in the C-TAH, and were confirmed in the feasibility study.
format Online
Article
Text
id pubmed-6906674
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-69066742019-12-20 Hemocompatibility and safety of the Carmat Total Artifical Heart hybrid membrane Richez, Ulysse De Castilla, Hector Guerin, Coralie L. Gendron, Nicolas Luraghi, Giulia Grimme, Marc Wu, Wei Taverna, Myriam Jansen, Piet Latremouille, Christian Migliavacca, Francesco Dubini, Gabriele Capel, Antoine Carpentier, Alain Smadja, David M. Heliyon Article The Carmat bioprosthetic total artificial heart (C-TAH) is a biventricular pump developed to minimize drawbacks of current mechanical assist devices and improve quality of life during support. This study aims to evaluate the safety of the hybrid membrane, which plays a pivotal role in this artificial heart. We investigated in particular its blood-contacting surface layer of bovine pericardial tissue, in terms of mechanical aging, risks of calcification, and impact of the hemodynamics shear stress inside the ventricles on blood components. Hybrid membranes were aged in a custom-designed endurance bench. Mechanical, physical and chemical properties were not significantly modified from 9 months up to 4 years of aging using a simulating process. Exploration of erosion areas did not show no risk of oil diffusion through the membrane. Blood contacting materials in the ventricular cavities were subcutaneously implanted in Wistar rats for 30 days as a model for calcification and demonstrated that the in-house anti-calcification pretreatment with Formaldehyde-Ethanol-Tween 80 was able to significantly reduce the calcium concentration from 132 μg/mg to 4.42 μg/mg (p < 0.001). Hemodynamic simulations with a computational model were used to reproduce shear stress in left and right ventricles and no significant stress was able to trigger hemolysis, platelet activation nor degradation of the von Willebrand factor multimers. Moreover, explanted hybrid membranes from patients included in the feasibility clinical study were analyzed confirming preclinical results with the absence of significant membrane calcification. At last, blood plasma bank analysis from the four patients implanted with C-TAH during the feasibility study showed no residual glutaraldehyde increase in plasma and confirmed hemodynamic simulation-based results with the absence of hemolysis and platelet activation associated with normal levels of plasma free hemoglobin and platelet microparticles after C-TAH implantation. These results on mechanical aging, calcification model and hemodynamic simulations predicted the safety of the hybrid membrane used in the C-TAH, and were confirmed in the feasibility study. Elsevier 2019-12-08 /pmc/articles/PMC6906674/ /pubmed/31867454 http://dx.doi.org/10.1016/j.heliyon.2019.e02914 Text en © 2019 Published by Elsevier Ltd. http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Article
Richez, Ulysse
De Castilla, Hector
Guerin, Coralie L.
Gendron, Nicolas
Luraghi, Giulia
Grimme, Marc
Wu, Wei
Taverna, Myriam
Jansen, Piet
Latremouille, Christian
Migliavacca, Francesco
Dubini, Gabriele
Capel, Antoine
Carpentier, Alain
Smadja, David M.
Hemocompatibility and safety of the Carmat Total Artifical Heart hybrid membrane
title Hemocompatibility and safety of the Carmat Total Artifical Heart hybrid membrane
title_full Hemocompatibility and safety of the Carmat Total Artifical Heart hybrid membrane
title_fullStr Hemocompatibility and safety of the Carmat Total Artifical Heart hybrid membrane
title_full_unstemmed Hemocompatibility and safety of the Carmat Total Artifical Heart hybrid membrane
title_short Hemocompatibility and safety of the Carmat Total Artifical Heart hybrid membrane
title_sort hemocompatibility and safety of the carmat total artifical heart hybrid membrane
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6906674/
https://www.ncbi.nlm.nih.gov/pubmed/31867454
http://dx.doi.org/10.1016/j.heliyon.2019.e02914
work_keys_str_mv AT richezulysse hemocompatibilityandsafetyofthecarmattotalartificalhearthybridmembrane
AT decastillahector hemocompatibilityandsafetyofthecarmattotalartificalhearthybridmembrane
AT guerincoraliel hemocompatibilityandsafetyofthecarmattotalartificalhearthybridmembrane
AT gendronnicolas hemocompatibilityandsafetyofthecarmattotalartificalhearthybridmembrane
AT luraghigiulia hemocompatibilityandsafetyofthecarmattotalartificalhearthybridmembrane
AT grimmemarc hemocompatibilityandsafetyofthecarmattotalartificalhearthybridmembrane
AT wuwei hemocompatibilityandsafetyofthecarmattotalartificalhearthybridmembrane
AT tavernamyriam hemocompatibilityandsafetyofthecarmattotalartificalhearthybridmembrane
AT jansenpiet hemocompatibilityandsafetyofthecarmattotalartificalhearthybridmembrane
AT latremouillechristian hemocompatibilityandsafetyofthecarmattotalartificalhearthybridmembrane
AT migliavaccafrancesco hemocompatibilityandsafetyofthecarmattotalartificalhearthybridmembrane
AT dubinigabriele hemocompatibilityandsafetyofthecarmattotalartificalhearthybridmembrane
AT capelantoine hemocompatibilityandsafetyofthecarmattotalartificalhearthybridmembrane
AT carpentieralain hemocompatibilityandsafetyofthecarmattotalartificalhearthybridmembrane
AT smadjadavidm hemocompatibilityandsafetyofthecarmattotalartificalhearthybridmembrane