A novel approach to Verify air gap and SSD for proton radiotherapy using surface imaging

PURPOSE: To develop a novel approach to accurately verify patient set up in proton radiotherapy, especially for the verification of the nozzle – body surface air gap and source-to-skin distance (SSD), the consistency and accuracy of which is extremely important in proton treatment. METHODS: Patient...

Descripción completa

Detalles Bibliográficos
Autores principales: Wang, Xiao, Ma, Chi, Davis, Rihan, Parikh, Rahul R., Jabbour, Salma K., Haffty, Bruce G., Yue, Ning J., Nie, Ke, Zhang, Yin
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6907344/
https://www.ncbi.nlm.nih.gov/pubmed/31829246
http://dx.doi.org/10.1186/s13014-019-1436-4
_version_ 1783478525821452288
author Wang, Xiao
Ma, Chi
Davis, Rihan
Parikh, Rahul R.
Jabbour, Salma K.
Haffty, Bruce G.
Yue, Ning J.
Nie, Ke
Zhang, Yin
author_facet Wang, Xiao
Ma, Chi
Davis, Rihan
Parikh, Rahul R.
Jabbour, Salma K.
Haffty, Bruce G.
Yue, Ning J.
Nie, Ke
Zhang, Yin
author_sort Wang, Xiao
collection PubMed
description PURPOSE: To develop a novel approach to accurately verify patient set up in proton radiotherapy, especially for the verification of the nozzle – body surface air gap and source-to-skin distance (SSD), the consistency and accuracy of which is extremely important in proton treatment. METHODS: Patient body surfaces can be captured and monitored with the optical surface imaging system during radiation treatment for improved intrafraction accuracy. An in-house software package was developed to reconstruct the patient body surface in the treatment position from the optical surface imaging reference capture and to calculate the corresponding nozzle – body surface air gap and SSD. To validate this method, a mannequin was scanned on a CT simulator and proton plans were generated for a Mevion S250 Proton machine with 20 gantry/couch angle combinations, as well as two different snout sizes, in the Varian Eclipse Treatment Planning Systems (TPS). The surface generated in the TPS from the CT scan was imported into the optical imaging system as an RT Structure for the purpose of validating and establishing a benchmark for ground truth comparison. The optical imaging surface reference capture was acquired at the treatment setup position after orthogonal kV imaging to confirm the positioning. The air gaps and SSDs calculated with the developed method from the surface captured at the treatment setup position (VRT surface) and the CT based surface imported from the TPS were compared to those calculated in TPS. The same approach was also applied to 14 clinical treatment fields for 10 patients to further validate the methodology. RESULTS: The air gaps and SSDs calculated from our program agreed well with the corresponding values derived from the TPS. For the phantom results, using the CT surface, the absolute differences in the air gap were 0.45 mm ± 0.33 mm for the small snout, and 0.51 mm ± 0.49 mm for the large snout, and the absolute differences in SSD were 0.68 mm ± 0.42 mm regardless of snout size. Using the VRT surface, the absolute differences in air gap were 1.17 mm ± 1.17 mm and 2.1 mm ± 3.09 mm for the small and large snouts, respectively, and the absolute differences in SSD were 0.81 mm ± 0.45 mm. Similarly, for patient data, using the CT surface, the absolute differences in air gap were 0.42 mm ± 0.49 mm, and the absolute differences in SSD were 1.92 mm ± 1.4 mm. Using the VRT surface, the absolute differences in the air gap were 2.35 mm ± 2.3 mm, and the absolute differences in SSD were 2.7 mm ± 2.17 mm. CONCLUSION: These results showed the feasibility and robustness of using an optical surface imaging approach to conveniently determine the air gap and SSD in proton treatment, providing an accurate and efficient way to confirm the target depth at treatment.
format Online
Article
Text
id pubmed-6907344
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-69073442019-12-19 A novel approach to Verify air gap and SSD for proton radiotherapy using surface imaging Wang, Xiao Ma, Chi Davis, Rihan Parikh, Rahul R. Jabbour, Salma K. Haffty, Bruce G. Yue, Ning J. Nie, Ke Zhang, Yin Radiat Oncol Research PURPOSE: To develop a novel approach to accurately verify patient set up in proton radiotherapy, especially for the verification of the nozzle – body surface air gap and source-to-skin distance (SSD), the consistency and accuracy of which is extremely important in proton treatment. METHODS: Patient body surfaces can be captured and monitored with the optical surface imaging system during radiation treatment for improved intrafraction accuracy. An in-house software package was developed to reconstruct the patient body surface in the treatment position from the optical surface imaging reference capture and to calculate the corresponding nozzle – body surface air gap and SSD. To validate this method, a mannequin was scanned on a CT simulator and proton plans were generated for a Mevion S250 Proton machine with 20 gantry/couch angle combinations, as well as two different snout sizes, in the Varian Eclipse Treatment Planning Systems (TPS). The surface generated in the TPS from the CT scan was imported into the optical imaging system as an RT Structure for the purpose of validating and establishing a benchmark for ground truth comparison. The optical imaging surface reference capture was acquired at the treatment setup position after orthogonal kV imaging to confirm the positioning. The air gaps and SSDs calculated with the developed method from the surface captured at the treatment setup position (VRT surface) and the CT based surface imported from the TPS were compared to those calculated in TPS. The same approach was also applied to 14 clinical treatment fields for 10 patients to further validate the methodology. RESULTS: The air gaps and SSDs calculated from our program agreed well with the corresponding values derived from the TPS. For the phantom results, using the CT surface, the absolute differences in the air gap were 0.45 mm ± 0.33 mm for the small snout, and 0.51 mm ± 0.49 mm for the large snout, and the absolute differences in SSD were 0.68 mm ± 0.42 mm regardless of snout size. Using the VRT surface, the absolute differences in air gap were 1.17 mm ± 1.17 mm and 2.1 mm ± 3.09 mm for the small and large snouts, respectively, and the absolute differences in SSD were 0.81 mm ± 0.45 mm. Similarly, for patient data, using the CT surface, the absolute differences in air gap were 0.42 mm ± 0.49 mm, and the absolute differences in SSD were 1.92 mm ± 1.4 mm. Using the VRT surface, the absolute differences in the air gap were 2.35 mm ± 2.3 mm, and the absolute differences in SSD were 2.7 mm ± 2.17 mm. CONCLUSION: These results showed the feasibility and robustness of using an optical surface imaging approach to conveniently determine the air gap and SSD in proton treatment, providing an accurate and efficient way to confirm the target depth at treatment. BioMed Central 2019-12-11 /pmc/articles/PMC6907344/ /pubmed/31829246 http://dx.doi.org/10.1186/s13014-019-1436-4 Text en © The Author(s). 2019 Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research
Wang, Xiao
Ma, Chi
Davis, Rihan
Parikh, Rahul R.
Jabbour, Salma K.
Haffty, Bruce G.
Yue, Ning J.
Nie, Ke
Zhang, Yin
A novel approach to Verify air gap and SSD for proton radiotherapy using surface imaging
title A novel approach to Verify air gap and SSD for proton radiotherapy using surface imaging
title_full A novel approach to Verify air gap and SSD for proton radiotherapy using surface imaging
title_fullStr A novel approach to Verify air gap and SSD for proton radiotherapy using surface imaging
title_full_unstemmed A novel approach to Verify air gap and SSD for proton radiotherapy using surface imaging
title_short A novel approach to Verify air gap and SSD for proton radiotherapy using surface imaging
title_sort novel approach to verify air gap and ssd for proton radiotherapy using surface imaging
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6907344/
https://www.ncbi.nlm.nih.gov/pubmed/31829246
http://dx.doi.org/10.1186/s13014-019-1436-4
work_keys_str_mv AT wangxiao anovelapproachtoverifyairgapandssdforprotonradiotherapyusingsurfaceimaging
AT machi anovelapproachtoverifyairgapandssdforprotonradiotherapyusingsurfaceimaging
AT davisrihan anovelapproachtoverifyairgapandssdforprotonradiotherapyusingsurfaceimaging
AT parikhrahulr anovelapproachtoverifyairgapandssdforprotonradiotherapyusingsurfaceimaging
AT jabboursalmak anovelapproachtoverifyairgapandssdforprotonradiotherapyusingsurfaceimaging
AT hafftybruceg anovelapproachtoverifyairgapandssdforprotonradiotherapyusingsurfaceimaging
AT yueningj anovelapproachtoverifyairgapandssdforprotonradiotherapyusingsurfaceimaging
AT nieke anovelapproachtoverifyairgapandssdforprotonradiotherapyusingsurfaceimaging
AT zhangyin anovelapproachtoverifyairgapandssdforprotonradiotherapyusingsurfaceimaging
AT wangxiao novelapproachtoverifyairgapandssdforprotonradiotherapyusingsurfaceimaging
AT machi novelapproachtoverifyairgapandssdforprotonradiotherapyusingsurfaceimaging
AT davisrihan novelapproachtoverifyairgapandssdforprotonradiotherapyusingsurfaceimaging
AT parikhrahulr novelapproachtoverifyairgapandssdforprotonradiotherapyusingsurfaceimaging
AT jabboursalmak novelapproachtoverifyairgapandssdforprotonradiotherapyusingsurfaceimaging
AT hafftybruceg novelapproachtoverifyairgapandssdforprotonradiotherapyusingsurfaceimaging
AT yueningj novelapproachtoverifyairgapandssdforprotonradiotherapyusingsurfaceimaging
AT nieke novelapproachtoverifyairgapandssdforprotonradiotherapyusingsurfaceimaging
AT zhangyin novelapproachtoverifyairgapandssdforprotonradiotherapyusingsurfaceimaging