Evaluation of the Beckman Coulter DxC 700 AU chemistry analyzer
OBJECTIVES: We evaluated and defined the analytical performance of the Beckman Coulter DxC 700 AU analyzer compared to the Siemens Dimension Vista 500 analyzer. DESIGN: and Methods: Performance characteristics included intra and inter-run precision, linearity/analytical measurement range, method cor...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Elsevier
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6909053/ https://www.ncbi.nlm.nih.gov/pubmed/31872016 http://dx.doi.org/10.1016/j.plabm.2019.e00148 |
_version_ | 1783478873321635840 |
---|---|
author | Bush, V.J. Smola, C. Schmitt, P. |
author_facet | Bush, V.J. Smola, C. Schmitt, P. |
author_sort | Bush, V.J. |
collection | PubMed |
description | OBJECTIVES: We evaluated and defined the analytical performance of the Beckman Coulter DxC 700 AU analyzer compared to the Siemens Dimension Vista 500 analyzer. DESIGN: and Methods: Performance characteristics included intra and inter-run precision, linearity/analytical measurement range, method correlation, and reference range verification. A total of 53 assays including 11 critical care, 19 general chemistries, 11 proteins, 10 urines, and 2 CSF analytes were tested. We also evaluated similarities and differences in assay methodologies between the 2 systems. RESULTS: The DxC 700 AU demonstrated excellent precision, comparable analytical measurement ranges and strong method correlation with the Dimension Vista 500 for most serum/plasma assays. 95% of the intra-run and 95% of the inter-run precision QC levels showed <3.0%CV and <6.0%CV, respectively. None of the deviations were clinically significant. The AMRs for all analytes except 5 met the manufacturer’s stated range. ALP, Lactate, U-glucose and CSF-glucose all recovered above the stated upper limit range, while prealbumin showed a smaller range. All analytes, except 14, showed slopes between 0.9 and 1.1 and/or biases <10%. Only ammonia, ferritin and lipase required significant reference range changes. The urine and CSF assays correlated very well with no adjustments in reference ranges required. CONCLUSIONS: The analytical performance of the DxC 700 AU analyzer was acceptable with only a small number of analytes requiring significant reference range changes. |
format | Online Article Text |
id | pubmed-6909053 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | Elsevier |
record_format | MEDLINE/PubMed |
spelling | pubmed-69090532019-12-23 Evaluation of the Beckman Coulter DxC 700 AU chemistry analyzer Bush, V.J. Smola, C. Schmitt, P. Pract Lab Med Article OBJECTIVES: We evaluated and defined the analytical performance of the Beckman Coulter DxC 700 AU analyzer compared to the Siemens Dimension Vista 500 analyzer. DESIGN: and Methods: Performance characteristics included intra and inter-run precision, linearity/analytical measurement range, method correlation, and reference range verification. A total of 53 assays including 11 critical care, 19 general chemistries, 11 proteins, 10 urines, and 2 CSF analytes were tested. We also evaluated similarities and differences in assay methodologies between the 2 systems. RESULTS: The DxC 700 AU demonstrated excellent precision, comparable analytical measurement ranges and strong method correlation with the Dimension Vista 500 for most serum/plasma assays. 95% of the intra-run and 95% of the inter-run precision QC levels showed <3.0%CV and <6.0%CV, respectively. None of the deviations were clinically significant. The AMRs for all analytes except 5 met the manufacturer’s stated range. ALP, Lactate, U-glucose and CSF-glucose all recovered above the stated upper limit range, while prealbumin showed a smaller range. All analytes, except 14, showed slopes between 0.9 and 1.1 and/or biases <10%. Only ammonia, ferritin and lipase required significant reference range changes. The urine and CSF assays correlated very well with no adjustments in reference ranges required. CONCLUSIONS: The analytical performance of the DxC 700 AU analyzer was acceptable with only a small number of analytes requiring significant reference range changes. Elsevier 2019-11-20 /pmc/articles/PMC6909053/ /pubmed/31872016 http://dx.doi.org/10.1016/j.plabm.2019.e00148 Text en © 2019 The Authors http://creativecommons.org/licenses/by/4.0/ This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/). |
spellingShingle | Article Bush, V.J. Smola, C. Schmitt, P. Evaluation of the Beckman Coulter DxC 700 AU chemistry analyzer |
title | Evaluation of the Beckman Coulter DxC 700 AU chemistry analyzer |
title_full | Evaluation of the Beckman Coulter DxC 700 AU chemistry analyzer |
title_fullStr | Evaluation of the Beckman Coulter DxC 700 AU chemistry analyzer |
title_full_unstemmed | Evaluation of the Beckman Coulter DxC 700 AU chemistry analyzer |
title_short | Evaluation of the Beckman Coulter DxC 700 AU chemistry analyzer |
title_sort | evaluation of the beckman coulter dxc 700 au chemistry analyzer |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6909053/ https://www.ncbi.nlm.nih.gov/pubmed/31872016 http://dx.doi.org/10.1016/j.plabm.2019.e00148 |
work_keys_str_mv | AT bushvj evaluationofthebeckmancoulterdxc700auchemistryanalyzer AT smolac evaluationofthebeckmancoulterdxc700auchemistryanalyzer AT schmittp evaluationofthebeckmancoulterdxc700auchemistryanalyzer |