Evaluation of lead equivalence of radiation protection apparatuses as a function of tube potential and spectral shaping filter

PURPOSE: This study aims to evaluate the lead equivalence (LE) of radiation protective apparatuses under various combinations of tube potentials and spectral shaping filter. METHOD: In this study, the commercially available 3M™ Lead Foil Tape 421, with nominal lead thickness of 0.1 mm, was employed...

Descripción completa

Detalles Bibliográficos
Autores principales: Aljabal, Areej Fawzi, Wargo, Richard Ryan, Lin, Pei‐Jan Paul
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6909127/
https://www.ncbi.nlm.nih.gov/pubmed/31738469
http://dx.doi.org/10.1002/acm2.12768
_version_ 1783478890838097920
author Aljabal, Areej Fawzi
Wargo, Richard Ryan
Lin, Pei‐Jan Paul
author_facet Aljabal, Areej Fawzi
Wargo, Richard Ryan
Lin, Pei‐Jan Paul
author_sort Aljabal, Areej Fawzi
collection PubMed
description PURPOSE: This study aims to evaluate the lead equivalence (LE) of radiation protective apparatuses under various combinations of tube potentials and spectral shaping filter. METHOD: In this study, the commercially available 3M™ Lead Foil Tape 421, with nominal lead thickness of 0.1 mm, was employed to determine the LE of four different radiation protective apparatuses. The LE of protective apparatus was determined by utilizing the X‐ray transmission curves obtained with the lead foil tape at 60–120 kVp in combination with the spectral shaping filters of 0.1, 0.2, 0.3, 0.6, and 0.9 mmCu. The experimental setup and test method, for the transmission measurements with narrow beam geometry, was performed in accordance to ASTM Designation F2547‐18 Standards. All measurements were obtained using cardiovascular interventional angiography system. RESULTS: A much larger discrepancies between the measured LE and stated (nominal) LE were observed at low tube potential (<70 kVp) for non‐lead protective apparatus. At higher tube potentials (>80 kVp) and thicker spectral shaping filters, the measured LE appears to be more consistent with the manufacturer specified nominal thickness for the protective apparatus investigated. On the other hand, for the lead protective eyeglasses, the measured lead equivalence of both the lead side shield and the lens of eyeglasses (0.38 and 0.85 mmPb respectively) are consistent across all tube voltage. CONCLUSION: The conventional specification of LE without considering spectral shaping filter is a valid measure for tube voltages at and above 80 kVp. The measured LE generally exceed the specifications. The difference is most significant at lower tube potentials, and especially with thicker spectral shaping filters. At higher voltages (>100 kVp), the measured LE and the nominal LE are in good agreement with each other irrespective of the spectral shaping filter thickness.
format Online
Article
Text
id pubmed-6909127
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-69091272019-12-20 Evaluation of lead equivalence of radiation protection apparatuses as a function of tube potential and spectral shaping filter Aljabal, Areej Fawzi Wargo, Richard Ryan Lin, Pei‐Jan Paul J Appl Clin Med Phys Technical Note PURPOSE: This study aims to evaluate the lead equivalence (LE) of radiation protective apparatuses under various combinations of tube potentials and spectral shaping filter. METHOD: In this study, the commercially available 3M™ Lead Foil Tape 421, with nominal lead thickness of 0.1 mm, was employed to determine the LE of four different radiation protective apparatuses. The LE of protective apparatus was determined by utilizing the X‐ray transmission curves obtained with the lead foil tape at 60–120 kVp in combination with the spectral shaping filters of 0.1, 0.2, 0.3, 0.6, and 0.9 mmCu. The experimental setup and test method, for the transmission measurements with narrow beam geometry, was performed in accordance to ASTM Designation F2547‐18 Standards. All measurements were obtained using cardiovascular interventional angiography system. RESULTS: A much larger discrepancies between the measured LE and stated (nominal) LE were observed at low tube potential (<70 kVp) for non‐lead protective apparatus. At higher tube potentials (>80 kVp) and thicker spectral shaping filters, the measured LE appears to be more consistent with the manufacturer specified nominal thickness for the protective apparatus investigated. On the other hand, for the lead protective eyeglasses, the measured lead equivalence of both the lead side shield and the lens of eyeglasses (0.38 and 0.85 mmPb respectively) are consistent across all tube voltage. CONCLUSION: The conventional specification of LE without considering spectral shaping filter is a valid measure for tube voltages at and above 80 kVp. The measured LE generally exceed the specifications. The difference is most significant at lower tube potentials, and especially with thicker spectral shaping filters. At higher voltages (>100 kVp), the measured LE and the nominal LE are in good agreement with each other irrespective of the spectral shaping filter thickness. John Wiley and Sons Inc. 2019-11-18 /pmc/articles/PMC6909127/ /pubmed/31738469 http://dx.doi.org/10.1002/acm2.12768 Text en © 2019 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Technical Note
Aljabal, Areej Fawzi
Wargo, Richard Ryan
Lin, Pei‐Jan Paul
Evaluation of lead equivalence of radiation protection apparatuses as a function of tube potential and spectral shaping filter
title Evaluation of lead equivalence of radiation protection apparatuses as a function of tube potential and spectral shaping filter
title_full Evaluation of lead equivalence of radiation protection apparatuses as a function of tube potential and spectral shaping filter
title_fullStr Evaluation of lead equivalence of radiation protection apparatuses as a function of tube potential and spectral shaping filter
title_full_unstemmed Evaluation of lead equivalence of radiation protection apparatuses as a function of tube potential and spectral shaping filter
title_short Evaluation of lead equivalence of radiation protection apparatuses as a function of tube potential and spectral shaping filter
title_sort evaluation of lead equivalence of radiation protection apparatuses as a function of tube potential and spectral shaping filter
topic Technical Note
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6909127/
https://www.ncbi.nlm.nih.gov/pubmed/31738469
http://dx.doi.org/10.1002/acm2.12768
work_keys_str_mv AT aljabalareejfawzi evaluationofleadequivalenceofradiationprotectionapparatusesasafunctionoftubepotentialandspectralshapingfilter
AT wargorichardryan evaluationofleadequivalenceofradiationprotectionapparatusesasafunctionoftubepotentialandspectralshapingfilter
AT linpeijanpaul evaluationofleadequivalenceofradiationprotectionapparatusesasafunctionoftubepotentialandspectralshapingfilter