Cargando…

Shunt-Bronchial Fistula with Coughing Up and Swallowing of Cerebrospinal Fluid: Rare Complication of Ventriculopleural Shunt

BACKGROUND: Erosion of the distal catheter into lung parenchyma is an extremely rare complication of ventriculopleural shunt placement. CASE DESCRIPTION: We report a 51-year-old woman with a history of parasagittal meningioma invading the sagittal sinus who presented with recurrent pneumonia after p...

Descripción completa

Detalles Bibliográficos
Autores principales: Katsevman, Gennadiy A., Harron, Raymond, Bhatia, Sanjay
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Elsevier 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6909168/
https://www.ncbi.nlm.nih.gov/pubmed/31872190
http://dx.doi.org/10.1016/j.wnsx.2019.100065
_version_ 1783478900711489536
author Katsevman, Gennadiy A.
Harron, Raymond
Bhatia, Sanjay
author_facet Katsevman, Gennadiy A.
Harron, Raymond
Bhatia, Sanjay
author_sort Katsevman, Gennadiy A.
collection PubMed
description BACKGROUND: Erosion of the distal catheter into lung parenchyma is an extremely rare complication of ventriculopleural shunt placement. CASE DESCRIPTION: We report a 51-year-old woman with a history of parasagittal meningioma invading the sagittal sinus who presented with recurrent pneumonia after placement of a ventriculopleural shunt. A nuclear study revealed accumulation of radiotracer material sequentially in the right hemithorax, trachea, mainstem bronchi, stomach, and bowel. The ventriculopleural shunt had eroded into the patient's lung parenchyma, with the effect of cerebrospinal fluid draining into the respiratory system and then being coughed up and swallowed into the gastrointestinal system. CONCLUSION: Surgeons should be aware of the potential complication of a ventriculopleural shunt eroding through the lung parenchyma to cause a shunt-bronchial fistula with persistent coughing and recurrent pneumonias. Shuntogram nuclear imaging may be useful in the diagnosis of the complication.
format Online
Article
Text
id pubmed-6909168
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Elsevier
record_format MEDLINE/PubMed
spelling pubmed-69091682019-12-23 Shunt-Bronchial Fistula with Coughing Up and Swallowing of Cerebrospinal Fluid: Rare Complication of Ventriculopleural Shunt Katsevman, Gennadiy A. Harron, Raymond Bhatia, Sanjay World Neurosurg X Case Report BACKGROUND: Erosion of the distal catheter into lung parenchyma is an extremely rare complication of ventriculopleural shunt placement. CASE DESCRIPTION: We report a 51-year-old woman with a history of parasagittal meningioma invading the sagittal sinus who presented with recurrent pneumonia after placement of a ventriculopleural shunt. A nuclear study revealed accumulation of radiotracer material sequentially in the right hemithorax, trachea, mainstem bronchi, stomach, and bowel. The ventriculopleural shunt had eroded into the patient's lung parenchyma, with the effect of cerebrospinal fluid draining into the respiratory system and then being coughed up and swallowed into the gastrointestinal system. CONCLUSION: Surgeons should be aware of the potential complication of a ventriculopleural shunt eroding through the lung parenchyma to cause a shunt-bronchial fistula with persistent coughing and recurrent pneumonias. Shuntogram nuclear imaging may be useful in the diagnosis of the complication. Elsevier 2019-11-01 /pmc/articles/PMC6909168/ /pubmed/31872190 http://dx.doi.org/10.1016/j.wnsx.2019.100065 Text en © 2019 The Author(s) http://creativecommons.org/licenses/by-nc-nd/4.0/ This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
spellingShingle Case Report
Katsevman, Gennadiy A.
Harron, Raymond
Bhatia, Sanjay
Shunt-Bronchial Fistula with Coughing Up and Swallowing of Cerebrospinal Fluid: Rare Complication of Ventriculopleural Shunt
title Shunt-Bronchial Fistula with Coughing Up and Swallowing of Cerebrospinal Fluid: Rare Complication of Ventriculopleural Shunt
title_full Shunt-Bronchial Fistula with Coughing Up and Swallowing of Cerebrospinal Fluid: Rare Complication of Ventriculopleural Shunt
title_fullStr Shunt-Bronchial Fistula with Coughing Up and Swallowing of Cerebrospinal Fluid: Rare Complication of Ventriculopleural Shunt
title_full_unstemmed Shunt-Bronchial Fistula with Coughing Up and Swallowing of Cerebrospinal Fluid: Rare Complication of Ventriculopleural Shunt
title_short Shunt-Bronchial Fistula with Coughing Up and Swallowing of Cerebrospinal Fluid: Rare Complication of Ventriculopleural Shunt
title_sort shunt-bronchial fistula with coughing up and swallowing of cerebrospinal fluid: rare complication of ventriculopleural shunt
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6909168/
https://www.ncbi.nlm.nih.gov/pubmed/31872190
http://dx.doi.org/10.1016/j.wnsx.2019.100065
work_keys_str_mv AT katsevmangennadiya shuntbronchialfistulawithcoughingupandswallowingofcerebrospinalfluidrarecomplicationofventriculopleuralshunt
AT harronraymond shuntbronchialfistulawithcoughingupandswallowingofcerebrospinalfluidrarecomplicationofventriculopleuralshunt
AT bhatiasanjay shuntbronchialfistulawithcoughingupandswallowingofcerebrospinalfluidrarecomplicationofventriculopleuralshunt