Cargando…

Fracture of the Femoral Component Caused by Insufficient Crimping after Modern Unicompartmental Knee Arthroplasty

Fracture of a femoral component after modern unicompartmental knee arthroplasty is very rare. Although this is not the first case on this subject, no study has reported insufficient crimping as the cause of femoral component loosening that led to breakage of a metallic component. A 69-year-old man u...

Descripción completa

Detalles Bibliográficos
Autores principales: Ishibashi, Yoichi, Mashiba, Tasuku, Mori, Masaki, Yamamoto, Tetsuji
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6914986/
https://www.ncbi.nlm.nih.gov/pubmed/31885982
http://dx.doi.org/10.1155/2019/5938598
_version_ 1783479928531976192
author Ishibashi, Yoichi
Mashiba, Tasuku
Mori, Masaki
Yamamoto, Tetsuji
author_facet Ishibashi, Yoichi
Mashiba, Tasuku
Mori, Masaki
Yamamoto, Tetsuji
author_sort Ishibashi, Yoichi
collection PubMed
description Fracture of a femoral component after modern unicompartmental knee arthroplasty is very rare. Although this is not the first case on this subject, no study has reported insufficient crimping as the cause of femoral component loosening that led to breakage of a metallic component. A 69-year-old man underwent medial unicompartmental knee arthroplasty for right medial knee osteoarthritis. His early postoperative course was good; however, the 1-year postoperative radiograph showed an apparent radiolucent line around the femoral component, and he occasionally had right knee pain. However, he had been followed up conservatively because he had been doing well even while doing heavy agricultural work. At 8 years after surgery, because breakage of the femoral component was found, revision surgery was performed using bicruciate-retaining total knee arthroplasty. The removed fractured femoral component revealed a thick cement mantle detached from the bone surface. The postoperative course of the patient after the revision surgery was excellent. We suggest that the causes of femoral component breakage include early implant loosening caused by uneven cement crimping of the femoral component to the bone and excessive loading stress as a result of heavy labour.
format Online
Article
Text
id pubmed-6914986
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Hindawi
record_format MEDLINE/PubMed
spelling pubmed-69149862019-12-29 Fracture of the Femoral Component Caused by Insufficient Crimping after Modern Unicompartmental Knee Arthroplasty Ishibashi, Yoichi Mashiba, Tasuku Mori, Masaki Yamamoto, Tetsuji Case Rep Orthop Case Report Fracture of a femoral component after modern unicompartmental knee arthroplasty is very rare. Although this is not the first case on this subject, no study has reported insufficient crimping as the cause of femoral component loosening that led to breakage of a metallic component. A 69-year-old man underwent medial unicompartmental knee arthroplasty for right medial knee osteoarthritis. His early postoperative course was good; however, the 1-year postoperative radiograph showed an apparent radiolucent line around the femoral component, and he occasionally had right knee pain. However, he had been followed up conservatively because he had been doing well even while doing heavy agricultural work. At 8 years after surgery, because breakage of the femoral component was found, revision surgery was performed using bicruciate-retaining total knee arthroplasty. The removed fractured femoral component revealed a thick cement mantle detached from the bone surface. The postoperative course of the patient after the revision surgery was excellent. We suggest that the causes of femoral component breakage include early implant loosening caused by uneven cement crimping of the femoral component to the bone and excessive loading stress as a result of heavy labour. Hindawi 2019-12-03 /pmc/articles/PMC6914986/ /pubmed/31885982 http://dx.doi.org/10.1155/2019/5938598 Text en Copyright © 2019 Yoichi Ishibashi et al. http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Case Report
Ishibashi, Yoichi
Mashiba, Tasuku
Mori, Masaki
Yamamoto, Tetsuji
Fracture of the Femoral Component Caused by Insufficient Crimping after Modern Unicompartmental Knee Arthroplasty
title Fracture of the Femoral Component Caused by Insufficient Crimping after Modern Unicompartmental Knee Arthroplasty
title_full Fracture of the Femoral Component Caused by Insufficient Crimping after Modern Unicompartmental Knee Arthroplasty
title_fullStr Fracture of the Femoral Component Caused by Insufficient Crimping after Modern Unicompartmental Knee Arthroplasty
title_full_unstemmed Fracture of the Femoral Component Caused by Insufficient Crimping after Modern Unicompartmental Knee Arthroplasty
title_short Fracture of the Femoral Component Caused by Insufficient Crimping after Modern Unicompartmental Knee Arthroplasty
title_sort fracture of the femoral component caused by insufficient crimping after modern unicompartmental knee arthroplasty
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6914986/
https://www.ncbi.nlm.nih.gov/pubmed/31885982
http://dx.doi.org/10.1155/2019/5938598
work_keys_str_mv AT ishibashiyoichi fractureofthefemoralcomponentcausedbyinsufficientcrimpingaftermodernunicompartmentalkneearthroplasty
AT mashibatasuku fractureofthefemoralcomponentcausedbyinsufficientcrimpingaftermodernunicompartmentalkneearthroplasty
AT morimasaki fractureofthefemoralcomponentcausedbyinsufficientcrimpingaftermodernunicompartmentalkneearthroplasty
AT yamamototetsuji fractureofthefemoralcomponentcausedbyinsufficientcrimpingaftermodernunicompartmentalkneearthroplasty