Cargando…
Fracture of the Femoral Component Caused by Insufficient Crimping after Modern Unicompartmental Knee Arthroplasty
Fracture of a femoral component after modern unicompartmental knee arthroplasty is very rare. Although this is not the first case on this subject, no study has reported insufficient crimping as the cause of femoral component loosening that led to breakage of a metallic component. A 69-year-old man u...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Hindawi
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6914986/ https://www.ncbi.nlm.nih.gov/pubmed/31885982 http://dx.doi.org/10.1155/2019/5938598 |
_version_ | 1783479928531976192 |
---|---|
author | Ishibashi, Yoichi Mashiba, Tasuku Mori, Masaki Yamamoto, Tetsuji |
author_facet | Ishibashi, Yoichi Mashiba, Tasuku Mori, Masaki Yamamoto, Tetsuji |
author_sort | Ishibashi, Yoichi |
collection | PubMed |
description | Fracture of a femoral component after modern unicompartmental knee arthroplasty is very rare. Although this is not the first case on this subject, no study has reported insufficient crimping as the cause of femoral component loosening that led to breakage of a metallic component. A 69-year-old man underwent medial unicompartmental knee arthroplasty for right medial knee osteoarthritis. His early postoperative course was good; however, the 1-year postoperative radiograph showed an apparent radiolucent line around the femoral component, and he occasionally had right knee pain. However, he had been followed up conservatively because he had been doing well even while doing heavy agricultural work. At 8 years after surgery, because breakage of the femoral component was found, revision surgery was performed using bicruciate-retaining total knee arthroplasty. The removed fractured femoral component revealed a thick cement mantle detached from the bone surface. The postoperative course of the patient after the revision surgery was excellent. We suggest that the causes of femoral component breakage include early implant loosening caused by uneven cement crimping of the femoral component to the bone and excessive loading stress as a result of heavy labour. |
format | Online Article Text |
id | pubmed-6914986 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | Hindawi |
record_format | MEDLINE/PubMed |
spelling | pubmed-69149862019-12-29 Fracture of the Femoral Component Caused by Insufficient Crimping after Modern Unicompartmental Knee Arthroplasty Ishibashi, Yoichi Mashiba, Tasuku Mori, Masaki Yamamoto, Tetsuji Case Rep Orthop Case Report Fracture of a femoral component after modern unicompartmental knee arthroplasty is very rare. Although this is not the first case on this subject, no study has reported insufficient crimping as the cause of femoral component loosening that led to breakage of a metallic component. A 69-year-old man underwent medial unicompartmental knee arthroplasty for right medial knee osteoarthritis. His early postoperative course was good; however, the 1-year postoperative radiograph showed an apparent radiolucent line around the femoral component, and he occasionally had right knee pain. However, he had been followed up conservatively because he had been doing well even while doing heavy agricultural work. At 8 years after surgery, because breakage of the femoral component was found, revision surgery was performed using bicruciate-retaining total knee arthroplasty. The removed fractured femoral component revealed a thick cement mantle detached from the bone surface. The postoperative course of the patient after the revision surgery was excellent. We suggest that the causes of femoral component breakage include early implant loosening caused by uneven cement crimping of the femoral component to the bone and excessive loading stress as a result of heavy labour. Hindawi 2019-12-03 /pmc/articles/PMC6914986/ /pubmed/31885982 http://dx.doi.org/10.1155/2019/5938598 Text en Copyright © 2019 Yoichi Ishibashi et al. http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Case Report Ishibashi, Yoichi Mashiba, Tasuku Mori, Masaki Yamamoto, Tetsuji Fracture of the Femoral Component Caused by Insufficient Crimping after Modern Unicompartmental Knee Arthroplasty |
title | Fracture of the Femoral Component Caused by Insufficient Crimping after Modern Unicompartmental Knee Arthroplasty |
title_full | Fracture of the Femoral Component Caused by Insufficient Crimping after Modern Unicompartmental Knee Arthroplasty |
title_fullStr | Fracture of the Femoral Component Caused by Insufficient Crimping after Modern Unicompartmental Knee Arthroplasty |
title_full_unstemmed | Fracture of the Femoral Component Caused by Insufficient Crimping after Modern Unicompartmental Knee Arthroplasty |
title_short | Fracture of the Femoral Component Caused by Insufficient Crimping after Modern Unicompartmental Knee Arthroplasty |
title_sort | fracture of the femoral component caused by insufficient crimping after modern unicompartmental knee arthroplasty |
topic | Case Report |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6914986/ https://www.ncbi.nlm.nih.gov/pubmed/31885982 http://dx.doi.org/10.1155/2019/5938598 |
work_keys_str_mv | AT ishibashiyoichi fractureofthefemoralcomponentcausedbyinsufficientcrimpingaftermodernunicompartmentalkneearthroplasty AT mashibatasuku fractureofthefemoralcomponentcausedbyinsufficientcrimpingaftermodernunicompartmentalkneearthroplasty AT morimasaki fractureofthefemoralcomponentcausedbyinsufficientcrimpingaftermodernunicompartmentalkneearthroplasty AT yamamototetsuji fractureofthefemoralcomponentcausedbyinsufficientcrimpingaftermodernunicompartmentalkneearthroplasty |