A comparison of IROC and ACDS on‐site audits of reference and non‐reference dosimetry

PURPOSE: Consistency between different international quality assurance groups is important in the progress toward similar standards and expectations in radiotherapy dosimetry around the world, and in the context of consistent clinical trial data from international trial participants. This study comp...

Descripción completa

Detalles Bibliográficos
Autores principales: Lye, Jessica, Kry, Stephen, Shaw, Maddison, Gibbons, Francis, Keehan, Stephanie, Lehmann, Joerg, Kron, Tomas, Followill, David, Williams, Ivan
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6916618/
https://www.ncbi.nlm.nih.gov/pubmed/31494941
http://dx.doi.org/10.1002/mp.13800
_version_ 1783480269391527936
author Lye, Jessica
Kry, Stephen
Shaw, Maddison
Gibbons, Francis
Keehan, Stephanie
Lehmann, Joerg
Kron, Tomas
Followill, David
Williams, Ivan
author_facet Lye, Jessica
Kry, Stephen
Shaw, Maddison
Gibbons, Francis
Keehan, Stephanie
Lehmann, Joerg
Kron, Tomas
Followill, David
Williams, Ivan
author_sort Lye, Jessica
collection PubMed
description PURPOSE: Consistency between different international quality assurance groups is important in the progress toward similar standards and expectations in radiotherapy dosimetry around the world, and in the context of consistent clinical trial data from international trial participants. This study compares the dosimetry audit methodology and results of two international quality assurance groups performing a side‐by‐side comparison at the same radiotherapy department, and interrogates the ability of the audits to detect deliberately introduced errors. METHODS: A comparison of the core dosimetry components of reference and non‐reference audits was conducted by the Imaging and Radiation Oncology Core (IROC, Houston, USA) and the Australian Clinical Dosimetry Service (ACDS, Melbourne, Australia). A set of measurements were conducted over 2 days at an Australian radiation therapy facility in Melbourne. Each group evaluated the reference dosimetry, output factors, small field output factors, percentage depth dose (PDD), wedge, and off‐axis factors according to their standard protocols. IROC additionally investigated the Electron PDD and the ACDS investigated the effect of heterogeneities. In order to evaluate and compare the performance of these audits under suboptimal conditions, artificial errors in percentage depth dose (PDD), EDW, and small field output factors were introduced into the 6 MV beam model to simulate potential commissioning/modeling errors and both audits were tested for their sensitivity in detecting these errors. RESULTS: With the plans from the clinical beam model, almost all results were within tolerance and at an optimal pass level. Good consistency was found between the two audits as almost all findings were consistent between them. Only two results were different between the results of IROC and the ACDS. The measurements of reference FFF photons showed a discrepancy of 0.7% between ACDS and IROC due to the inclusion of a 0.5% nonuniformity correction by the ACDS. The second difference between IROC and the ACDS was seen with the lung phantom. The asymmetric field behind lung measured by the ACDS was slightly (0.3%) above the ACDS's pass (optimal) level of 3.3%. IROC did not detect this issue because their measurements were all assessed in a homogeneous phantom. When errors were deliberately introduced neither audit was sensitive enough to pick up a 2% change to the small field output factors. The introduced PDD change was flagged by both audits. Similarly, the introduced error of using 25° wedge instead of 30° wedge was detectible in both audits as out of tolerance. CONCLUSIONS: Despite different equipment, approach, and scope of measurements in on‐site audits, there were clear similarities between the results from the two groups. This finding is encouraging in the context of a global harmonized approach to radiotherapy quality assurance and dosimetry audit.
format Online
Article
Text
id pubmed-6916618
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-69166182019-12-23 A comparison of IROC and ACDS on‐site audits of reference and non‐reference dosimetry Lye, Jessica Kry, Stephen Shaw, Maddison Gibbons, Francis Keehan, Stephanie Lehmann, Joerg Kron, Tomas Followill, David Williams, Ivan Med Phys Medical Dataset Article PURPOSE: Consistency between different international quality assurance groups is important in the progress toward similar standards and expectations in radiotherapy dosimetry around the world, and in the context of consistent clinical trial data from international trial participants. This study compares the dosimetry audit methodology and results of two international quality assurance groups performing a side‐by‐side comparison at the same radiotherapy department, and interrogates the ability of the audits to detect deliberately introduced errors. METHODS: A comparison of the core dosimetry components of reference and non‐reference audits was conducted by the Imaging and Radiation Oncology Core (IROC, Houston, USA) and the Australian Clinical Dosimetry Service (ACDS, Melbourne, Australia). A set of measurements were conducted over 2 days at an Australian radiation therapy facility in Melbourne. Each group evaluated the reference dosimetry, output factors, small field output factors, percentage depth dose (PDD), wedge, and off‐axis factors according to their standard protocols. IROC additionally investigated the Electron PDD and the ACDS investigated the effect of heterogeneities. In order to evaluate and compare the performance of these audits under suboptimal conditions, artificial errors in percentage depth dose (PDD), EDW, and small field output factors were introduced into the 6 MV beam model to simulate potential commissioning/modeling errors and both audits were tested for their sensitivity in detecting these errors. RESULTS: With the plans from the clinical beam model, almost all results were within tolerance and at an optimal pass level. Good consistency was found between the two audits as almost all findings were consistent between them. Only two results were different between the results of IROC and the ACDS. The measurements of reference FFF photons showed a discrepancy of 0.7% between ACDS and IROC due to the inclusion of a 0.5% nonuniformity correction by the ACDS. The second difference between IROC and the ACDS was seen with the lung phantom. The asymmetric field behind lung measured by the ACDS was slightly (0.3%) above the ACDS's pass (optimal) level of 3.3%. IROC did not detect this issue because their measurements were all assessed in a homogeneous phantom. When errors were deliberately introduced neither audit was sensitive enough to pick up a 2% change to the small field output factors. The introduced PDD change was flagged by both audits. Similarly, the introduced error of using 25° wedge instead of 30° wedge was detectible in both audits as out of tolerance. CONCLUSIONS: Despite different equipment, approach, and scope of measurements in on‐site audits, there were clear similarities between the results from the two groups. This finding is encouraging in the context of a global harmonized approach to radiotherapy quality assurance and dosimetry audit. John Wiley and Sons Inc. 2019-10-25 2019-12 /pmc/articles/PMC6916618/ /pubmed/31494941 http://dx.doi.org/10.1002/mp.13800 Text en © 2019 The Authors. Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Medical Dataset Article
Lye, Jessica
Kry, Stephen
Shaw, Maddison
Gibbons, Francis
Keehan, Stephanie
Lehmann, Joerg
Kron, Tomas
Followill, David
Williams, Ivan
A comparison of IROC and ACDS on‐site audits of reference and non‐reference dosimetry
title A comparison of IROC and ACDS on‐site audits of reference and non‐reference dosimetry
title_full A comparison of IROC and ACDS on‐site audits of reference and non‐reference dosimetry
title_fullStr A comparison of IROC and ACDS on‐site audits of reference and non‐reference dosimetry
title_full_unstemmed A comparison of IROC and ACDS on‐site audits of reference and non‐reference dosimetry
title_short A comparison of IROC and ACDS on‐site audits of reference and non‐reference dosimetry
title_sort comparison of iroc and acds on‐site audits of reference and non‐reference dosimetry
topic Medical Dataset Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6916618/
https://www.ncbi.nlm.nih.gov/pubmed/31494941
http://dx.doi.org/10.1002/mp.13800
work_keys_str_mv AT lyejessica acomparisonofirocandacdsonsiteauditsofreferenceandnonreferencedosimetry
AT krystephen acomparisonofirocandacdsonsiteauditsofreferenceandnonreferencedosimetry
AT shawmaddison acomparisonofirocandacdsonsiteauditsofreferenceandnonreferencedosimetry
AT gibbonsfrancis acomparisonofirocandacdsonsiteauditsofreferenceandnonreferencedosimetry
AT keehanstephanie acomparisonofirocandacdsonsiteauditsofreferenceandnonreferencedosimetry
AT lehmannjoerg acomparisonofirocandacdsonsiteauditsofreferenceandnonreferencedosimetry
AT krontomas acomparisonofirocandacdsonsiteauditsofreferenceandnonreferencedosimetry
AT followilldavid acomparisonofirocandacdsonsiteauditsofreferenceandnonreferencedosimetry
AT williamsivan acomparisonofirocandacdsonsiteauditsofreferenceandnonreferencedosimetry
AT lyejessica comparisonofirocandacdsonsiteauditsofreferenceandnonreferencedosimetry
AT krystephen comparisonofirocandacdsonsiteauditsofreferenceandnonreferencedosimetry
AT shawmaddison comparisonofirocandacdsonsiteauditsofreferenceandnonreferencedosimetry
AT gibbonsfrancis comparisonofirocandacdsonsiteauditsofreferenceandnonreferencedosimetry
AT keehanstephanie comparisonofirocandacdsonsiteauditsofreferenceandnonreferencedosimetry
AT lehmannjoerg comparisonofirocandacdsonsiteauditsofreferenceandnonreferencedosimetry
AT krontomas comparisonofirocandacdsonsiteauditsofreferenceandnonreferencedosimetry
AT followilldavid comparisonofirocandacdsonsiteauditsofreferenceandnonreferencedosimetry
AT williamsivan comparisonofirocandacdsonsiteauditsofreferenceandnonreferencedosimetry