Validation of a Low-Cost Electromyography (EMG) System via a Commercial and Accurate EMG Device: Pilot Study
Electromyography (EMG) devices are well-suited for measuring the behaviour of muscles during an exercise or a task, and are widely used in many different research areas. Their disadvantage is that commercial systems are expensive. We designed a low-cost EMG system with enough accuracy and reliabilit...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6928739/ https://www.ncbi.nlm.nih.gov/pubmed/31795083 http://dx.doi.org/10.3390/s19235214 |
Sumario: | Electromyography (EMG) devices are well-suited for measuring the behaviour of muscles during an exercise or a task, and are widely used in many different research areas. Their disadvantage is that commercial systems are expensive. We designed a low-cost EMG system with enough accuracy and reliability to be used in a wide range of possible ways. The present article focuses on the validation of the low-cost system we designed, which is compared with a commercially available, accurate device. The evaluation was done by means of a set of experiments, in which volunteers performed isometric and dynamic exercises while EMG signals from the rectus femoris muscle were registered by both the proposed low-cost system and a commercial system simultaneously. Analysis and assessment of three indicators to estimate the similarity between both signals were developed. These indicated a very good result, with spearman’s correlation averaging above 0.60, the energy ratio close to the 80% and the linear correlation coefficient approximating 100%. The agreement between both systems (custom and commercial) is excellent, although there are also some limitations, such as the delay of the signal (<1 s) and noise due to the hardware and assembly in the proposed system. |
---|