A Sensor for Spirometric Feedback in Ventilation Maneuvers during Cardiopulmonary Resuscitation Training

This work proposes adapting an existing sensor and embedding it on mannequins used in cardiopulmonary resuscitation (CPR) training to accurately measure the amount of air supplied to the lungs during ventilation. Mathematical modeling, calibration, and validation of the sensor along with metrology,...

Descripción completa

Detalles Bibliográficos
Autores principales: Leocádio, Rodolfo Rocha Vieira, Segundo, Alan Kardek Rêgo, Louzada, Cibelle Ferreira
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6929026/
https://www.ncbi.nlm.nih.gov/pubmed/31766452
http://dx.doi.org/10.3390/s19235095
_version_ 1783482609169334272
author Leocádio, Rodolfo Rocha Vieira
Segundo, Alan Kardek Rêgo
Louzada, Cibelle Ferreira
author_facet Leocádio, Rodolfo Rocha Vieira
Segundo, Alan Kardek Rêgo
Louzada, Cibelle Ferreira
author_sort Leocádio, Rodolfo Rocha Vieira
collection PubMed
description This work proposes adapting an existing sensor and embedding it on mannequins used in cardiopulmonary resuscitation (CPR) training to accurately measure the amount of air supplied to the lungs during ventilation. Mathematical modeling, calibration, and validation of the sensor along with metrology, statistical inference, and spirometry techniques were used as a base for aquiring scientific knowledge of the system. The system directly measures the variable of interest (air volume) and refers to spirometric techniques in the elaboration of its model. This improves the realism of the dummies during the CPR training, because it estimates, in real-time, not only the volume of air entering in the lungs but also the Forced Vital Capacity (FVC), Forced Expiratory Volume (FEVt) and Medium Forced Expiratory Flow (FEF(20–75%)). The validation of the sensor achieved results that address the requirements for this application, that is, the error below 3.4% of full scale. During the spirometric tests, the system presented the measurement results of (305 ± 22, 450 ± 23, 603 ± 24, 751 ± 26, 922 ± 27, 1021 ± 30, 1182 ± 33, 1326 ± 36, 1476 ± 37, 1618 ± 45 and 1786 ± 56) × 10(−6) m(3) for reference values of (300, 450, 600, 750, 900, 1050, 1200, 1350, 1500, 1650 and 1800) × 10(−6) m(3), respectively. Therefore, considering the spirometry and pressure boundary conditions of the manikin lungs, the system achieves the objective of simulating valid spirometric data for debriefings, that is, there is an agreement between the measurement results when compared to the signal generated by a commercial spirometer (Koko brand). The main advantages that this work presents in relation to the sensors commonly used for this purpose are: (i) the reduced cost, which makes it possible, for the first time, to use a respiratory volume sensor in medical simulators or training dummies; (ii) the direct measurement of air entering the lung using a noninvasive method, which makes it possible to use spirometry parameters to characterize simulated human respiration during the CPR training; and (iii) the measurement of spirometric parameters (FVC, FEVt, and FEF(20–75%)), in real-time, during the CPR training, to achieve optimal ventilation performance. Therefore, the system developed in this work addresses the minimum requirements for the practice of ventilation in the CPR maneuvers and has great potential in several future applications.
format Online
Article
Text
id pubmed-6929026
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-69290262019-12-26 A Sensor for Spirometric Feedback in Ventilation Maneuvers during Cardiopulmonary Resuscitation Training Leocádio, Rodolfo Rocha Vieira Segundo, Alan Kardek Rêgo Louzada, Cibelle Ferreira Sensors (Basel) Article This work proposes adapting an existing sensor and embedding it on mannequins used in cardiopulmonary resuscitation (CPR) training to accurately measure the amount of air supplied to the lungs during ventilation. Mathematical modeling, calibration, and validation of the sensor along with metrology, statistical inference, and spirometry techniques were used as a base for aquiring scientific knowledge of the system. The system directly measures the variable of interest (air volume) and refers to spirometric techniques in the elaboration of its model. This improves the realism of the dummies during the CPR training, because it estimates, in real-time, not only the volume of air entering in the lungs but also the Forced Vital Capacity (FVC), Forced Expiratory Volume (FEVt) and Medium Forced Expiratory Flow (FEF(20–75%)). The validation of the sensor achieved results that address the requirements for this application, that is, the error below 3.4% of full scale. During the spirometric tests, the system presented the measurement results of (305 ± 22, 450 ± 23, 603 ± 24, 751 ± 26, 922 ± 27, 1021 ± 30, 1182 ± 33, 1326 ± 36, 1476 ± 37, 1618 ± 45 and 1786 ± 56) × 10(−6) m(3) for reference values of (300, 450, 600, 750, 900, 1050, 1200, 1350, 1500, 1650 and 1800) × 10(−6) m(3), respectively. Therefore, considering the spirometry and pressure boundary conditions of the manikin lungs, the system achieves the objective of simulating valid spirometric data for debriefings, that is, there is an agreement between the measurement results when compared to the signal generated by a commercial spirometer (Koko brand). The main advantages that this work presents in relation to the sensors commonly used for this purpose are: (i) the reduced cost, which makes it possible, for the first time, to use a respiratory volume sensor in medical simulators or training dummies; (ii) the direct measurement of air entering the lung using a noninvasive method, which makes it possible to use spirometry parameters to characterize simulated human respiration during the CPR training; and (iii) the measurement of spirometric parameters (FVC, FEVt, and FEF(20–75%)), in real-time, during the CPR training, to achieve optimal ventilation performance. Therefore, the system developed in this work addresses the minimum requirements for the practice of ventilation in the CPR maneuvers and has great potential in several future applications. MDPI 2019-11-21 /pmc/articles/PMC6929026/ /pubmed/31766452 http://dx.doi.org/10.3390/s19235095 Text en © 2019 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (http://creativecommons.org/licenses/by/4.0/).
spellingShingle Article
Leocádio, Rodolfo Rocha Vieira
Segundo, Alan Kardek Rêgo
Louzada, Cibelle Ferreira
A Sensor for Spirometric Feedback in Ventilation Maneuvers during Cardiopulmonary Resuscitation Training
title A Sensor for Spirometric Feedback in Ventilation Maneuvers during Cardiopulmonary Resuscitation Training
title_full A Sensor for Spirometric Feedback in Ventilation Maneuvers during Cardiopulmonary Resuscitation Training
title_fullStr A Sensor for Spirometric Feedback in Ventilation Maneuvers during Cardiopulmonary Resuscitation Training
title_full_unstemmed A Sensor for Spirometric Feedback in Ventilation Maneuvers during Cardiopulmonary Resuscitation Training
title_short A Sensor for Spirometric Feedback in Ventilation Maneuvers during Cardiopulmonary Resuscitation Training
title_sort sensor for spirometric feedback in ventilation maneuvers during cardiopulmonary resuscitation training
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6929026/
https://www.ncbi.nlm.nih.gov/pubmed/31766452
http://dx.doi.org/10.3390/s19235095
work_keys_str_mv AT leocadiorodolforochavieira asensorforspirometricfeedbackinventilationmaneuversduringcardiopulmonaryresuscitationtraining
AT segundoalankardekrego asensorforspirometricfeedbackinventilationmaneuversduringcardiopulmonaryresuscitationtraining
AT louzadacibelleferreira asensorforspirometricfeedbackinventilationmaneuversduringcardiopulmonaryresuscitationtraining
AT leocadiorodolforochavieira sensorforspirometricfeedbackinventilationmaneuversduringcardiopulmonaryresuscitationtraining
AT segundoalankardekrego sensorforspirometricfeedbackinventilationmaneuversduringcardiopulmonaryresuscitationtraining
AT louzadacibelleferreira sensorforspirometricfeedbackinventilationmaneuversduringcardiopulmonaryresuscitationtraining