Quantitative implications of the updated EARL 2019 PET–CT performance standards

PURPOSE: Recently, updated EARL specifications (EARL2) have been developed and announced. This study aims at investigating the impact of the EARL2 specifications on the quantitative reads of clinical PET–CT studies and testing a method to enable the use of the EARL2 standards whilst still generating...

Descripción completa

Detalles Bibliográficos
Autores principales: Kaalep, Andres, Burggraaff, Coreline N., Pieplenbosch, Simone, Verwer, Eline E., Sera, Terez, Zijlstra, Josee, Hoekstra, Otto S., Oprea-Lager, Daniela E., Boellaard, Ronald
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6933045/
https://www.ncbi.nlm.nih.gov/pubmed/31879795
http://dx.doi.org/10.1186/s40658-019-0257-8
_version_ 1783483134050828288
author Kaalep, Andres
Burggraaff, Coreline N.
Pieplenbosch, Simone
Verwer, Eline E.
Sera, Terez
Zijlstra, Josee
Hoekstra, Otto S.
Oprea-Lager, Daniela E.
Boellaard, Ronald
author_facet Kaalep, Andres
Burggraaff, Coreline N.
Pieplenbosch, Simone
Verwer, Eline E.
Sera, Terez
Zijlstra, Josee
Hoekstra, Otto S.
Oprea-Lager, Daniela E.
Boellaard, Ronald
author_sort Kaalep, Andres
collection PubMed
description PURPOSE: Recently, updated EARL specifications (EARL2) have been developed and announced. This study aims at investigating the impact of the EARL2 specifications on the quantitative reads of clinical PET–CT studies and testing a method to enable the use of the EARL2 standards whilst still generating quantitative reads compliant with current EARL standards (EARL1). METHODS: Thirteen non-small cell lung cancer (NSCLC) and seventeen lymphoma PET–CT studies were used to derive four image datasets—the first dataset complying with EARL1 specifications and the second reconstructed using parameters as described in EARL2. For the third (EARL2F6) and fourth (EARL2F7) dataset in EARL2, respectively, 6 mm and 7 mm Gaussian post-filtering was applied. We compared the results of quantitative metrics (MATV, SUVmax, SUVpeak, SUVmean, TLG, and tumor-to-liver and tumor-to-blood pool ratios) obtained with these 4 datasets in 55 suspected malignant lesions using three commonly used segmentation/volume of interest (VOI) methods (MAX41, A50P, SUV4). RESULTS: We found that with EARL2 MAX41 VOI method, MATV decreases by 22%, TLG remains unchanged and SUV values increase by 23–30% depending on the specific metric used. The EARL2F7 dataset produced quantitative metrics best aligning with EARL1, with no significant differences between most of the datasets (p>0.05). Different VOI methods performed similarly with regard to SUV metrics but differences in MATV as well as TLG were observed. No significant difference between NSCLC and lymphoma cancer types was observed. CONCLUSIONS: Application of EARL2 standards can result in higher SUVs, reduced MATV and slightly changed TLG values relative to EARL1. Applying a Gaussian filter to PET images reconstructed using EARL2 parameters successfully yielded EARL1 compliant data.
format Online
Article
Text
id pubmed-6933045
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Springer International Publishing
record_format MEDLINE/PubMed
spelling pubmed-69330452020-01-09 Quantitative implications of the updated EARL 2019 PET–CT performance standards Kaalep, Andres Burggraaff, Coreline N. Pieplenbosch, Simone Verwer, Eline E. Sera, Terez Zijlstra, Josee Hoekstra, Otto S. Oprea-Lager, Daniela E. Boellaard, Ronald EJNMMI Phys Original Research PURPOSE: Recently, updated EARL specifications (EARL2) have been developed and announced. This study aims at investigating the impact of the EARL2 specifications on the quantitative reads of clinical PET–CT studies and testing a method to enable the use of the EARL2 standards whilst still generating quantitative reads compliant with current EARL standards (EARL1). METHODS: Thirteen non-small cell lung cancer (NSCLC) and seventeen lymphoma PET–CT studies were used to derive four image datasets—the first dataset complying with EARL1 specifications and the second reconstructed using parameters as described in EARL2. For the third (EARL2F6) and fourth (EARL2F7) dataset in EARL2, respectively, 6 mm and 7 mm Gaussian post-filtering was applied. We compared the results of quantitative metrics (MATV, SUVmax, SUVpeak, SUVmean, TLG, and tumor-to-liver and tumor-to-blood pool ratios) obtained with these 4 datasets in 55 suspected malignant lesions using three commonly used segmentation/volume of interest (VOI) methods (MAX41, A50P, SUV4). RESULTS: We found that with EARL2 MAX41 VOI method, MATV decreases by 22%, TLG remains unchanged and SUV values increase by 23–30% depending on the specific metric used. The EARL2F7 dataset produced quantitative metrics best aligning with EARL1, with no significant differences between most of the datasets (p>0.05). Different VOI methods performed similarly with regard to SUV metrics but differences in MATV as well as TLG were observed. No significant difference between NSCLC and lymphoma cancer types was observed. CONCLUSIONS: Application of EARL2 standards can result in higher SUVs, reduced MATV and slightly changed TLG values relative to EARL1. Applying a Gaussian filter to PET images reconstructed using EARL2 parameters successfully yielded EARL1 compliant data. Springer International Publishing 2019-12-26 /pmc/articles/PMC6933045/ /pubmed/31879795 http://dx.doi.org/10.1186/s40658-019-0257-8 Text en © The Author(s). 2019 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
spellingShingle Original Research
Kaalep, Andres
Burggraaff, Coreline N.
Pieplenbosch, Simone
Verwer, Eline E.
Sera, Terez
Zijlstra, Josee
Hoekstra, Otto S.
Oprea-Lager, Daniela E.
Boellaard, Ronald
Quantitative implications of the updated EARL 2019 PET–CT performance standards
title Quantitative implications of the updated EARL 2019 PET–CT performance standards
title_full Quantitative implications of the updated EARL 2019 PET–CT performance standards
title_fullStr Quantitative implications of the updated EARL 2019 PET–CT performance standards
title_full_unstemmed Quantitative implications of the updated EARL 2019 PET–CT performance standards
title_short Quantitative implications of the updated EARL 2019 PET–CT performance standards
title_sort quantitative implications of the updated earl 2019 pet–ct performance standards
topic Original Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6933045/
https://www.ncbi.nlm.nih.gov/pubmed/31879795
http://dx.doi.org/10.1186/s40658-019-0257-8
work_keys_str_mv AT kaalepandres quantitativeimplicationsoftheupdatedearl2019petctperformancestandards
AT burggraaffcorelinen quantitativeimplicationsoftheupdatedearl2019petctperformancestandards
AT pieplenboschsimone quantitativeimplicationsoftheupdatedearl2019petctperformancestandards
AT verwerelinee quantitativeimplicationsoftheupdatedearl2019petctperformancestandards
AT seraterez quantitativeimplicationsoftheupdatedearl2019petctperformancestandards
AT zijlstrajosee quantitativeimplicationsoftheupdatedearl2019petctperformancestandards
AT hoekstraottos quantitativeimplicationsoftheupdatedearl2019petctperformancestandards
AT oprealagerdanielae quantitativeimplicationsoftheupdatedearl2019petctperformancestandards
AT boellaardronald quantitativeimplicationsoftheupdatedearl2019petctperformancestandards