Comparison of CAD/CAM abutment and prefabricated abutment in Morse taper internal type implant after cyclic loading: Axial displacement, removal torque, and tensile removal force
PURPOSE: The purpose of this study was to compare computer-aided design/computer-aided manufacturing (CAD/CAM) abutment and prefabricated abutment in Morse taper internal connection type implants after cyclic loading. MATERIALS AND METHODS: The study was conducted with internal type implants of two...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Korean Academy of Prosthodontics
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6933047/ https://www.ncbi.nlm.nih.gov/pubmed/31897269 http://dx.doi.org/10.4047/jap.2019.11.6.305 |
_version_ | 1783483134289903616 |
---|---|
author | Yi, Yuseung Heo, Seong-Joo Koak, Jai-Young Kim, Seong-Kyun |
author_facet | Yi, Yuseung Heo, Seong-Joo Koak, Jai-Young Kim, Seong-Kyun |
author_sort | Yi, Yuseung |
collection | PubMed |
description | PURPOSE: The purpose of this study was to compare computer-aided design/computer-aided manufacturing (CAD/CAM) abutment and prefabricated abutment in Morse taper internal connection type implants after cyclic loading. MATERIALS AND METHODS: The study was conducted with internal type implants of two different manufacturers (Group Os, De). Fourteen assemblies were prepared for each manufacturer group and divided into 2 groups (n=7): prefabricated abutments (Os-P, De-P) and CAD/CAM abutments (Os-C, De-C). The amount of axial displacement and the removal torque values (RTVs) were measured before and after cyclic loading (10(6) cycles, 3 Hz with 150 N), and the tensile removal force to dislodge the abutments was measured after cyclic loading. A repeated measures ANOVA and a pattern analysis based on the logarithmic regression model were conducted to evaluate the effect of cyclic loading on the axial displacement. The Wilcoxon signed-rank test and the Mann-Whitney test was conducted for comparison of RTV reduction% and tensile removal forces. RESULTS: There was no significant difference between CAD/CAM abutments and prefabricated abutments in axial displacement and tensile removal force; however, significantly greater RTV reduction% after cyclic loading was observed in CAD/CAM abutments. The correlation among the axial displacement, the RTV, and the tensile removal force was not significant. CONCLUSION: The use of CAD/CAM abutment did not significantly affect the amount of axial displacement and tensile removal force, but presented a significantly greater removal torque reduction% than prefabricated abutments. The connection stability due to the friction at the abutment-implant interface of CAD/CAM abutments may not be different from prefabricated abutment. |
format | Online Article Text |
id | pubmed-6933047 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | The Korean Academy of Prosthodontics |
record_format | MEDLINE/PubMed |
spelling | pubmed-69330472020-01-02 Comparison of CAD/CAM abutment and prefabricated abutment in Morse taper internal type implant after cyclic loading: Axial displacement, removal torque, and tensile removal force Yi, Yuseung Heo, Seong-Joo Koak, Jai-Young Kim, Seong-Kyun J Adv Prosthodont Original Article PURPOSE: The purpose of this study was to compare computer-aided design/computer-aided manufacturing (CAD/CAM) abutment and prefabricated abutment in Morse taper internal connection type implants after cyclic loading. MATERIALS AND METHODS: The study was conducted with internal type implants of two different manufacturers (Group Os, De). Fourteen assemblies were prepared for each manufacturer group and divided into 2 groups (n=7): prefabricated abutments (Os-P, De-P) and CAD/CAM abutments (Os-C, De-C). The amount of axial displacement and the removal torque values (RTVs) were measured before and after cyclic loading (10(6) cycles, 3 Hz with 150 N), and the tensile removal force to dislodge the abutments was measured after cyclic loading. A repeated measures ANOVA and a pattern analysis based on the logarithmic regression model were conducted to evaluate the effect of cyclic loading on the axial displacement. The Wilcoxon signed-rank test and the Mann-Whitney test was conducted for comparison of RTV reduction% and tensile removal forces. RESULTS: There was no significant difference between CAD/CAM abutments and prefabricated abutments in axial displacement and tensile removal force; however, significantly greater RTV reduction% after cyclic loading was observed in CAD/CAM abutments. The correlation among the axial displacement, the RTV, and the tensile removal force was not significant. CONCLUSION: The use of CAD/CAM abutment did not significantly affect the amount of axial displacement and tensile removal force, but presented a significantly greater removal torque reduction% than prefabricated abutments. The connection stability due to the friction at the abutment-implant interface of CAD/CAM abutments may not be different from prefabricated abutment. The Korean Academy of Prosthodontics 2019-12 2019-12-18 /pmc/articles/PMC6933047/ /pubmed/31897269 http://dx.doi.org/10.4047/jap.2019.11.6.305 Text en © 2019 The Korean Academy of Prosthodontics http://creativecommons.org/licenses/by-nc/4.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Yi, Yuseung Heo, Seong-Joo Koak, Jai-Young Kim, Seong-Kyun Comparison of CAD/CAM abutment and prefabricated abutment in Morse taper internal type implant after cyclic loading: Axial displacement, removal torque, and tensile removal force |
title | Comparison of CAD/CAM abutment and prefabricated abutment in Morse taper internal type implant after cyclic loading: Axial displacement, removal torque, and tensile removal force |
title_full | Comparison of CAD/CAM abutment and prefabricated abutment in Morse taper internal type implant after cyclic loading: Axial displacement, removal torque, and tensile removal force |
title_fullStr | Comparison of CAD/CAM abutment and prefabricated abutment in Morse taper internal type implant after cyclic loading: Axial displacement, removal torque, and tensile removal force |
title_full_unstemmed | Comparison of CAD/CAM abutment and prefabricated abutment in Morse taper internal type implant after cyclic loading: Axial displacement, removal torque, and tensile removal force |
title_short | Comparison of CAD/CAM abutment and prefabricated abutment in Morse taper internal type implant after cyclic loading: Axial displacement, removal torque, and tensile removal force |
title_sort | comparison of cad/cam abutment and prefabricated abutment in morse taper internal type implant after cyclic loading: axial displacement, removal torque, and tensile removal force |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6933047/ https://www.ncbi.nlm.nih.gov/pubmed/31897269 http://dx.doi.org/10.4047/jap.2019.11.6.305 |
work_keys_str_mv | AT yiyuseung comparisonofcadcamabutmentandprefabricatedabutmentinmorsetaperinternaltypeimplantaftercyclicloadingaxialdisplacementremovaltorqueandtensileremovalforce AT heoseongjoo comparisonofcadcamabutmentandprefabricatedabutmentinmorsetaperinternaltypeimplantaftercyclicloadingaxialdisplacementremovaltorqueandtensileremovalforce AT koakjaiyoung comparisonofcadcamabutmentandprefabricatedabutmentinmorsetaperinternaltypeimplantaftercyclicloadingaxialdisplacementremovaltorqueandtensileremovalforce AT kimseongkyun comparisonofcadcamabutmentandprefabricatedabutmentinmorsetaperinternaltypeimplantaftercyclicloadingaxialdisplacementremovaltorqueandtensileremovalforce |