4π Radiotherapy Using a Linear Accelerator: A Misnomer in Violation of the Solid Geometric Boundary Conditions in Three-Dimensional Euclidean Space

PURPOSE: The concept of 4π(c) radiotherapy is a radiotherapy planning technique receiving much attention in recent times. The aim of this article is to disprove the feasibility of the 4π radiotherapy using a cantilever-type linear accelerator or any other external-beam delivery machines. MATERIALS A...

Descripción completa

Detalles Bibliográficos
Autores principales: Sarkar, Biplab, Ganesh, Tharmarnadar, Munshi, Anusheel, Manikandan, Arjunan, Mohanti, Bidhu Kalyan
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer - Medknow 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6936196/
https://www.ncbi.nlm.nih.gov/pubmed/31908388
http://dx.doi.org/10.4103/jmp.JMP_2_19
_version_ 1783483698496143360
author Sarkar, Biplab
Ganesh, Tharmarnadar
Munshi, Anusheel
Manikandan, Arjunan
Mohanti, Bidhu Kalyan
author_facet Sarkar, Biplab
Ganesh, Tharmarnadar
Munshi, Anusheel
Manikandan, Arjunan
Mohanti, Bidhu Kalyan
author_sort Sarkar, Biplab
collection PubMed
description PURPOSE: The concept of 4π(c) radiotherapy is a radiotherapy planning technique receiving much attention in recent times. The aim of this article is to disprove the feasibility of the 4π radiotherapy using a cantilever-type linear accelerator or any other external-beam delivery machines. MATERIALS AND METHODS: A surface integral-based mathematical derivation for the maximum achievable solid angle for a linear accelerator was carried out respecting the rotational boundary conditions for gantry and couch in three-dimensional Euclidean space. The allowed movements include a gantry rotation of 0–2π(c) and a table rotation of [Image: see text]. RESULTS: Total achievable solid angle by cantilever-type linear accelerator (or any teletherapy machine employing a cantilever design) is [Image: see text] , which is applicable only for the foot and brain radiotherapy where the allowed table rotation is 90°–0°–270°. For other sites such as pelvis, thorax, or abdomen, achievable solid angle as the couch rotation comes down significantly. Practically, only suitable couch angle is 0° by avoiding gantry–couch–patient collision. CONCLUSIONS: Present cantilever design of linear accelerator prevents achieving a 4π radian solid angle at any point in the patient. Even the most modern therapy machines like CyberKnife which has a robotic arm also cannot achieve 4π geometry. Maximum achievable solid angle under the highest allowable boundary condition(s) cannot exceed 2π(c), which is restricted for only extremities such as foot and brain radiotherapy. For other parts of the body such as pelvis, thorax, and abdomen, the solid angle is reduced to 1/5(th) (maximum value) of the 4π(c). To obtain a 4π(c) solid angle in a three-dimensional Euclidean space, the patient has to be a zero-dimensional point and X-ray head of the linear accelerator has a freedom to rotate in every point of a hypothetical sphere of radius 1 m. This article establishes geometrically why it is not possible to achieve a 4π(c) solid angle.
format Online
Article
Text
id pubmed-6936196
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Wolters Kluwer - Medknow
record_format MEDLINE/PubMed
spelling pubmed-69361962020-01-06 4π Radiotherapy Using a Linear Accelerator: A Misnomer in Violation of the Solid Geometric Boundary Conditions in Three-Dimensional Euclidean Space Sarkar, Biplab Ganesh, Tharmarnadar Munshi, Anusheel Manikandan, Arjunan Mohanti, Bidhu Kalyan J Med Phys Technical Note PURPOSE: The concept of 4π(c) radiotherapy is a radiotherapy planning technique receiving much attention in recent times. The aim of this article is to disprove the feasibility of the 4π radiotherapy using a cantilever-type linear accelerator or any other external-beam delivery machines. MATERIALS AND METHODS: A surface integral-based mathematical derivation for the maximum achievable solid angle for a linear accelerator was carried out respecting the rotational boundary conditions for gantry and couch in three-dimensional Euclidean space. The allowed movements include a gantry rotation of 0–2π(c) and a table rotation of [Image: see text]. RESULTS: Total achievable solid angle by cantilever-type linear accelerator (or any teletherapy machine employing a cantilever design) is [Image: see text] , which is applicable only for the foot and brain radiotherapy where the allowed table rotation is 90°–0°–270°. For other sites such as pelvis, thorax, or abdomen, achievable solid angle as the couch rotation comes down significantly. Practically, only suitable couch angle is 0° by avoiding gantry–couch–patient collision. CONCLUSIONS: Present cantilever design of linear accelerator prevents achieving a 4π radian solid angle at any point in the patient. Even the most modern therapy machines like CyberKnife which has a robotic arm also cannot achieve 4π geometry. Maximum achievable solid angle under the highest allowable boundary condition(s) cannot exceed 2π(c), which is restricted for only extremities such as foot and brain radiotherapy. For other parts of the body such as pelvis, thorax, and abdomen, the solid angle is reduced to 1/5(th) (maximum value) of the 4π(c). To obtain a 4π(c) solid angle in a three-dimensional Euclidean space, the patient has to be a zero-dimensional point and X-ray head of the linear accelerator has a freedom to rotate in every point of a hypothetical sphere of radius 1 m. This article establishes geometrically why it is not possible to achieve a 4π(c) solid angle. Wolters Kluwer - Medknow 2019 2019-12-11 /pmc/articles/PMC6936196/ /pubmed/31908388 http://dx.doi.org/10.4103/jmp.JMP_2_19 Text en Copyright: © 2019 Journal of Medical Physics http://creativecommons.org/licenses/by-nc-sa/4.0 This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
spellingShingle Technical Note
Sarkar, Biplab
Ganesh, Tharmarnadar
Munshi, Anusheel
Manikandan, Arjunan
Mohanti, Bidhu Kalyan
4π Radiotherapy Using a Linear Accelerator: A Misnomer in Violation of the Solid Geometric Boundary Conditions in Three-Dimensional Euclidean Space
title 4π Radiotherapy Using a Linear Accelerator: A Misnomer in Violation of the Solid Geometric Boundary Conditions in Three-Dimensional Euclidean Space
title_full 4π Radiotherapy Using a Linear Accelerator: A Misnomer in Violation of the Solid Geometric Boundary Conditions in Three-Dimensional Euclidean Space
title_fullStr 4π Radiotherapy Using a Linear Accelerator: A Misnomer in Violation of the Solid Geometric Boundary Conditions in Three-Dimensional Euclidean Space
title_full_unstemmed 4π Radiotherapy Using a Linear Accelerator: A Misnomer in Violation of the Solid Geometric Boundary Conditions in Three-Dimensional Euclidean Space
title_short 4π Radiotherapy Using a Linear Accelerator: A Misnomer in Violation of the Solid Geometric Boundary Conditions in Three-Dimensional Euclidean Space
title_sort 4π radiotherapy using a linear accelerator: a misnomer in violation of the solid geometric boundary conditions in three-dimensional euclidean space
topic Technical Note
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6936196/
https://www.ncbi.nlm.nih.gov/pubmed/31908388
http://dx.doi.org/10.4103/jmp.JMP_2_19
work_keys_str_mv AT sarkarbiplab 4pradiotherapyusingalinearacceleratoramisnomerinviolationofthesolidgeometricboundaryconditionsinthreedimensionaleuclideanspace
AT ganeshtharmarnadar 4pradiotherapyusingalinearacceleratoramisnomerinviolationofthesolidgeometricboundaryconditionsinthreedimensionaleuclideanspace
AT munshianusheel 4pradiotherapyusingalinearacceleratoramisnomerinviolationofthesolidgeometricboundaryconditionsinthreedimensionaleuclideanspace
AT manikandanarjunan 4pradiotherapyusingalinearacceleratoramisnomerinviolationofthesolidgeometricboundaryconditionsinthreedimensionaleuclideanspace
AT mohantibidhukalyan 4pradiotherapyusingalinearacceleratoramisnomerinviolationofthesolidgeometricboundaryconditionsinthreedimensionaleuclideanspace