Estimation of Backscatter from Internal Shielding in Electron Beam Radiotherapy Using Monte Carlo Simulations (EGSnrc) and Gafchromic Film Measurements
PURPOSE: The purpose of the study was to estimate the backscatter electron dose in internal shielding during electron beam therapy using Monte Carlo (MC) simulations and Gafchromic film measurements. MATERIALS AND METHODS: About 6 and 9 MeV electron beams from a Varian 2100C linac were simulated usi...
Autores principales: | Singh, Sukhvir, Semwal, Manoj Kumar, Bhatt, C. P. |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Wolters Kluwer - Medknow
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6936201/ https://www.ncbi.nlm.nih.gov/pubmed/31908382 http://dx.doi.org/10.4103/jmp.JMP_21_19 |
Ejemplares similares
-
Evaluation of backscatter dose from internal lead shielding in clinical electron beams using EGSnrc Monte Carlo simulations
por: de Vries, Rowen J., et al.
Publicado: (2015) -
Commissioning Validation of a Brachytherapy Treatment Planning System with EGSnrc Monte Carlo Code and EBT3 GAFChromic Film
por: Lahlabou, Mohammed, et al.
Publicado: (2022) -
Design and Evaluation of Structural Shielding of a Typical Radiotherapy Facility Using EGSnrc Monte Carlo Code
por: Ibitoye, Zaccheaus Ayo, et al.
Publicado: (2022) -
Assessment of Surface and Build-up Doses for a 6 MV Photon Beam using Parallel Plate Chamber, EBT3 Gafchromic Films, and PRIMO Monte Carlo Simulation Code
por: Mahur, Mamta, et al.
Publicado: (2022) -
Optimization of Variance Reduction Techniques used in EGSnrc Monte Carlo Codes
por: Shanmugasundaram, Sangeetha, et al.
Publicado: (2018)