Cardiac magnetic resonance imaging in a patient with temporary external pacemaker: a case report

BACKGROUND : Magnetic resonance imaging (MRI) is increasingly becoming the imaging modality of choice for many clinical disorders due to superior image quality and absence of radiation. However, access to MRI remains limited for most patients with cardiac implantable electronic devices due to potent...

Descripción completa

Detalles Bibliográficos
Autores principales: Chaudhry, Uzma, Svensson, Jonas, Mosén, Henrik, Mörtsell, David
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Oxford University Press 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6939813/
https://www.ncbi.nlm.nih.gov/pubmed/31911995
http://dx.doi.org/10.1093/ehjcr/ytz228
Descripción
Sumario:BACKGROUND : Magnetic resonance imaging (MRI) is increasingly becoming the imaging modality of choice for many clinical disorders due to superior image quality and absence of radiation. However, access to MRI remains limited for most patients with cardiac implantable electronic devices due to potential safety concerns. In line with guidelines, there is no absolute contraindication to perform MRI, but warrants careful risk-benefit assessment. CASE SUMMARY : A 59-year-old man was admitted with a 5-day history of central chest pain and few week’s history of general malaise, dry cough, and breathlessness. Electrocardiogram confirmed complete atrioventricular block (CAVB). A slight increase in cardiac enzyme was noted. Coronary angiogram revealed atheromatous changes, but no obstructive coronary lesion. A temporary transvenous pacemaker was inserted. Transthoracic echocardiogram confirmed a dilated left ventricle with severely reduced left ventricular function. To facilitate diagnosis (hence prognosis), management and mobilization, investigation with cardiovascular magnetic resonance (CMR) was warranted but contraindicated by the temporary transvenous pacemaker. An active fixation pacemaker lead was therefore placed in the right ventricle via percutaneous puncture of the right subclavian vein and connected to a pulse generator, both secured to the skin with sutures and adhesive medical dressing. Appropriate device programming and close patient monitoring ensured that CMR could be performed without any adverse effects. A diagnosis of acute myocarditis was confirmed. Regular device interrogation during an extended 3-week period with temporary pacing ruled out any device failure. As there was no resolution of CAVB, the patient received a dual-chamber pacemaker. DISCUSSION : Cardiovascular magnetic resonance was feasible and safely performed on a patient with a temporary permanent external pacemaker system using a standard screw-in pacing lead and a regular pulse generator fixed to the skin. Although more studies are needed for generalizability, CMR may be used in highly selected patients with a temporary pacemaker.