Study of Dosimetric Properties of Flattening Filter Free Photon Beam Passing through Cadmium Free Compensator Alloy

BACKGROUND: This study aims to investigate radiation beam geometry of Cyberknife beam and change in dosimetric characteristics of six megavoltage (6MV) flattening filter free (FFF) beam after passing through high density cadmium free compensator alloy MATERIAL AND METHODS: In this experimental study...

Descripción completa

Detalles Bibliográficos
Autores principales: Kaushik, S., Punia, R., Tyagi, A.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Shiraz University of Medical Sciences 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6943851/
https://www.ncbi.nlm.nih.gov/pubmed/32039095
http://dx.doi.org/10.31661/jbpe.v0i0.745
_version_ 1783484961690484736
author Kaushik, S.
Punia, R.
Tyagi, A.
author_facet Kaushik, S.
Punia, R.
Tyagi, A.
author_sort Kaushik, S.
collection PubMed
description BACKGROUND: This study aims to investigate radiation beam geometry of Cyberknife beam and change in dosimetric characteristics of six megavoltage (6MV) flattening filter free (FFF) beam after passing through high density cadmium free compensator alloy MATERIAL AND METHODS: In this experimental study, changes in FFF beam dosimetric characteristics after passing through compensator alloy was measured. Transmitted intensity of FFF beam was measured in air by an ion chamber at a source to detector distance (SDD) of 800mm. Extended SDD measurement also has been performed at a distance of 1270mm to analyze scattering due to compensator. Linear attenuation coefficient (µ(eff)) was measured for cadmium free compensator alloy using simple exponential attenuation model. Percentage depth doses (PDDs) have been measured by a radiation field analyzer with compensator material to observe the beam hardening and change in surface doses and depth doses. RESULTS: Linear attenuation coefficient of compensator alloy was measured 0.042 (Standard Deviation ±0.00099) mm(-1) and it was found that there is no change with increase in collimator size. Even after increasing distance source from detector, µ(eff) has no change. PDDs were found to increase with thickness of compensator. PDD from a 60mm collimator size increased by 5% and 6% at a depth of 100mm and 200mm, respectively in water. PDD also increased with collimator size less significantly. Surface dose was found to decrease with increase in compensator thickness CONCLUSION: Cyberknife beam has been found to be narrow beam geometry. FFF beam contains lesser scattered photons. Presence of high density compensator filters out the soft x-ray photon causes significant dosimetric changes
format Online
Article
Text
id pubmed-6943851
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Shiraz University of Medical Sciences
record_format MEDLINE/PubMed
spelling pubmed-69438512020-02-07 Study of Dosimetric Properties of Flattening Filter Free Photon Beam Passing through Cadmium Free Compensator Alloy Kaushik, S. Punia, R. Tyagi, A. J Biomed Phys Eng Original Article BACKGROUND: This study aims to investigate radiation beam geometry of Cyberknife beam and change in dosimetric characteristics of six megavoltage (6MV) flattening filter free (FFF) beam after passing through high density cadmium free compensator alloy MATERIAL AND METHODS: In this experimental study, changes in FFF beam dosimetric characteristics after passing through compensator alloy was measured. Transmitted intensity of FFF beam was measured in air by an ion chamber at a source to detector distance (SDD) of 800mm. Extended SDD measurement also has been performed at a distance of 1270mm to analyze scattering due to compensator. Linear attenuation coefficient (µ(eff)) was measured for cadmium free compensator alloy using simple exponential attenuation model. Percentage depth doses (PDDs) have been measured by a radiation field analyzer with compensator material to observe the beam hardening and change in surface doses and depth doses. RESULTS: Linear attenuation coefficient of compensator alloy was measured 0.042 (Standard Deviation ±0.00099) mm(-1) and it was found that there is no change with increase in collimator size. Even after increasing distance source from detector, µ(eff) has no change. PDDs were found to increase with thickness of compensator. PDD from a 60mm collimator size increased by 5% and 6% at a depth of 100mm and 200mm, respectively in water. PDD also increased with collimator size less significantly. Surface dose was found to decrease with increase in compensator thickness CONCLUSION: Cyberknife beam has been found to be narrow beam geometry. FFF beam contains lesser scattered photons. Presence of high density compensator filters out the soft x-ray photon causes significant dosimetric changes Shiraz University of Medical Sciences 2019-12-01 /pmc/articles/PMC6943851/ /pubmed/32039095 http://dx.doi.org/10.31661/jbpe.v0i0.745 Text en Copyright: © Shiraz University of Medical Sciences http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open-access article distributed under the terms of the Creative Commons Attribution-Noncommercial-Share Alike 3.0 Unported, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Kaushik, S.
Punia, R.
Tyagi, A.
Study of Dosimetric Properties of Flattening Filter Free Photon Beam Passing through Cadmium Free Compensator Alloy
title Study of Dosimetric Properties of Flattening Filter Free Photon Beam Passing through Cadmium Free Compensator Alloy
title_full Study of Dosimetric Properties of Flattening Filter Free Photon Beam Passing through Cadmium Free Compensator Alloy
title_fullStr Study of Dosimetric Properties of Flattening Filter Free Photon Beam Passing through Cadmium Free Compensator Alloy
title_full_unstemmed Study of Dosimetric Properties of Flattening Filter Free Photon Beam Passing through Cadmium Free Compensator Alloy
title_short Study of Dosimetric Properties of Flattening Filter Free Photon Beam Passing through Cadmium Free Compensator Alloy
title_sort study of dosimetric properties of flattening filter free photon beam passing through cadmium free compensator alloy
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6943851/
https://www.ncbi.nlm.nih.gov/pubmed/32039095
http://dx.doi.org/10.31661/jbpe.v0i0.745
work_keys_str_mv AT kaushiks studyofdosimetricpropertiesofflatteningfilterfreephotonbeampassingthroughcadmiumfreecompensatoralloy
AT puniar studyofdosimetricpropertiesofflatteningfilterfreephotonbeampassingthroughcadmiumfreecompensatoralloy
AT tyagia studyofdosimetricpropertiesofflatteningfilterfreephotonbeampassingthroughcadmiumfreecompensatoralloy