Cargando…

Evaluation of screening effectiveness of hepatitis B surface antigen and anti-HCV rapid test kits in Pakistan

OBJECTIVE: The current study was conducted to evaluate the performance and screening effectiveness of commercially available rapid screening kits in comparison with chemiluminescence immunoassay (CLIA) and polymerase chain reaction (PCR). MATERIALS AND METHODS: This single-center, cross-sectional st...

Descripción completa

Detalles Bibliográficos
Autores principales: Waheed, Usman, Abdella, Yetmgeta Eyayou, Saba, Noor e, Arshad, Muhammad, Wazeer, Akhlaaq, Farooq, Ahmad, Usman, Javaid, Arshad, Abida, Zaheer, Hasan Abbas
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer - Medknow 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6943868/
https://www.ncbi.nlm.nih.gov/pubmed/31929706
http://dx.doi.org/10.4103/JLP.JLP_172_19
Descripción
Sumario:OBJECTIVE: The current study was conducted to evaluate the performance and screening effectiveness of commercially available rapid screening kits in comparison with chemiluminescence immunoassay (CLIA) and polymerase chain reaction (PCR). MATERIALS AND METHODS: This single-center, cross-sectional study was conducted at the Department of Pathology and Blood Transfusion Services, Shaheed Zulfiqar Ali Bhutto Medical University, PIMS, Islamabad, from January to April 2019. A total of 10 commercially available immunochromatographic test (ICT) devices and one CLIA kit (LIAISON XL) were tested for their sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), and accuracy using 100 positive and 100 negative samples each for HBV and HCV, respectively. RESULTS: The sensitivities and specificities of ICT kits for hepatitis B surface antigen were 65% and 70% (Hightop), 67% and 85% (RightSign), 62% and 73% (Wondfo), 70% and 80% (Accu-Chek), 68% and 77% (Fastep), 73% and 85% (Abon), 77% and 83% (ImmuMed), 80% and 90% (Insta-Answer), 67% and 81% (BioCheck), and 72% and 83% CTK Biotech, respectively. Similarly, the sensitivities and specificities of different ICT kits for HCV were 69% and 80% (Hightop), 76% and 83% (RightSign), 69% and 81% (Wondfo), 78% and 79% (Accu-Check), 68% and 68% (Fastep), 63% and 73% (Abon), 71% and 70% (ImmuMed), 79% and 68% (Insta-Answer), 62% and 66% (BioChek), and 69% and 78% CTK Biotech, respectively. The sensitivity and specificity of Diasorin Liaison Murex assay for both HBV and HCV were found to be 100% when compared with PCR. The PPV, NPV and Accuracy were determined accordingly. CONCLUSION: Rapid testing ICT devices for both HBV and HCV available in Pakistan were found to have a variable degree of sensitivity and specificity when compared with CLIA and PCR. Comparatively expensive but quality methods are more reliable as compared to rapid devices.