Performance characteristic evaluation of a bismuth germanate-based high-sensitivity 5-ring discovery image quality positron emission tomography/computed tomography system as per National Electrical Manufacturers Association NU 2-2012

National Electrical Manufacturers Association (NEMA) provides guidelines to assess the performance of Positron Emission Tomography (PET). A PET/CT scanner, Discovery IQ, GE Medical systems, Milwaukee, USA was installed in our department which has high a sensitivity PET component. We have performed t...

Descripción completa

Detalles Bibliográficos
Autores principales: Jha, Ashish Kumar, Mithun, Sneha, Puranik, Ameya D., Purandare, Nilendu C., Shah, Sneha, Agrawal, Archi, Rangarajan, Venkatesh
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer - Medknow 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6945355/
https://www.ncbi.nlm.nih.gov/pubmed/31933550
http://dx.doi.org/10.4103/wjnm.WJNM_72_18
_version_ 1783485160378859520
author Jha, Ashish Kumar
Mithun, Sneha
Puranik, Ameya D.
Purandare, Nilendu C.
Shah, Sneha
Agrawal, Archi
Rangarajan, Venkatesh
author_facet Jha, Ashish Kumar
Mithun, Sneha
Puranik, Ameya D.
Purandare, Nilendu C.
Shah, Sneha
Agrawal, Archi
Rangarajan, Venkatesh
author_sort Jha, Ashish Kumar
collection PubMed
description National Electrical Manufacturers Association (NEMA) provides guidelines to assess the performance of Positron Emission Tomography (PET). A PET/CT scanner, Discovery IQ, GE Medical systems, Milwaukee, USA was installed in our department which has high a sensitivity PET component. We have performed the NEMA NU-2 2012 quality control tests to evaluate this system on site before clinical use. Performance measurements of the PET scanner were made using the NEMA NU2-2012 procedures for spatial resolution, scatter fraction, sensitivity, count rate loss and random coincidence estimation, Noise Equivalent Count Rate (NECR) and image quality. As per NU2 2012, spatial resolution was measured at 1 cm, 10 cm and 20 cm vertically from the centre and at each of these points resolution was measured at tangential, radial and axial directions. Sensitivity was measured at centre and 10 cm off center vertically from the center. The system sensitivity is reported as an average of the two measured values. Scatter fraction and NECR measurements, Image quality test was also performed. The tangential, radial and axial FWHM were 4.99 mm, 4.20 mm and 4.79 mm at 1 cm off centre, 5.49 mm, 4.69 mm and 4.81 mm at 10 cm off centre and 7.99 mm, 5.07 mm and 4.95 mm at 20 cm off centre respectively. The absolute sensitivity of this scanner was found to be 20.1 cps/kBq. The scatter fraction calculated from the decay method was 37.94% and NECR was 125 kcps. The peak NECR was achieved at activity concentration of 8.7 KBq/ml and the count loss below the peak NECR was found to be 0.68%. Image quality test for, contrast recovery, background variability and lung error residual mean met all specifications. Overall PET performance of Discovery IQ whole-body scanner was satisfactory and the scanner met all the performance specifications required by NEMA 2012.
format Online
Article
Text
id pubmed-6945355
institution National Center for Biotechnology Information
language English
publishDate 2019
publisher Wolters Kluwer - Medknow
record_format MEDLINE/PubMed
spelling pubmed-69453552020-01-13 Performance characteristic evaluation of a bismuth germanate-based high-sensitivity 5-ring discovery image quality positron emission tomography/computed tomography system as per National Electrical Manufacturers Association NU 2-2012 Jha, Ashish Kumar Mithun, Sneha Puranik, Ameya D. Purandare, Nilendu C. Shah, Sneha Agrawal, Archi Rangarajan, Venkatesh World J Nucl Med Original Article National Electrical Manufacturers Association (NEMA) provides guidelines to assess the performance of Positron Emission Tomography (PET). A PET/CT scanner, Discovery IQ, GE Medical systems, Milwaukee, USA was installed in our department which has high a sensitivity PET component. We have performed the NEMA NU-2 2012 quality control tests to evaluate this system on site before clinical use. Performance measurements of the PET scanner were made using the NEMA NU2-2012 procedures for spatial resolution, scatter fraction, sensitivity, count rate loss and random coincidence estimation, Noise Equivalent Count Rate (NECR) and image quality. As per NU2 2012, spatial resolution was measured at 1 cm, 10 cm and 20 cm vertically from the centre and at each of these points resolution was measured at tangential, radial and axial directions. Sensitivity was measured at centre and 10 cm off center vertically from the center. The system sensitivity is reported as an average of the two measured values. Scatter fraction and NECR measurements, Image quality test was also performed. The tangential, radial and axial FWHM were 4.99 mm, 4.20 mm and 4.79 mm at 1 cm off centre, 5.49 mm, 4.69 mm and 4.81 mm at 10 cm off centre and 7.99 mm, 5.07 mm and 4.95 mm at 20 cm off centre respectively. The absolute sensitivity of this scanner was found to be 20.1 cps/kBq. The scatter fraction calculated from the decay method was 37.94% and NECR was 125 kcps. The peak NECR was achieved at activity concentration of 8.7 KBq/ml and the count loss below the peak NECR was found to be 0.68%. Image quality test for, contrast recovery, background variability and lung error residual mean met all specifications. Overall PET performance of Discovery IQ whole-body scanner was satisfactory and the scanner met all the performance specifications required by NEMA 2012. Wolters Kluwer - Medknow 2019-12-18 /pmc/articles/PMC6945355/ /pubmed/31933550 http://dx.doi.org/10.4103/wjnm.WJNM_72_18 Text en Copyright: © 2019 World Journal of Nuclear Medicine http://creativecommons.org/licenses/by-nc-sa/4.0 This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
spellingShingle Original Article
Jha, Ashish Kumar
Mithun, Sneha
Puranik, Ameya D.
Purandare, Nilendu C.
Shah, Sneha
Agrawal, Archi
Rangarajan, Venkatesh
Performance characteristic evaluation of a bismuth germanate-based high-sensitivity 5-ring discovery image quality positron emission tomography/computed tomography system as per National Electrical Manufacturers Association NU 2-2012
title Performance characteristic evaluation of a bismuth germanate-based high-sensitivity 5-ring discovery image quality positron emission tomography/computed tomography system as per National Electrical Manufacturers Association NU 2-2012
title_full Performance characteristic evaluation of a bismuth germanate-based high-sensitivity 5-ring discovery image quality positron emission tomography/computed tomography system as per National Electrical Manufacturers Association NU 2-2012
title_fullStr Performance characteristic evaluation of a bismuth germanate-based high-sensitivity 5-ring discovery image quality positron emission tomography/computed tomography system as per National Electrical Manufacturers Association NU 2-2012
title_full_unstemmed Performance characteristic evaluation of a bismuth germanate-based high-sensitivity 5-ring discovery image quality positron emission tomography/computed tomography system as per National Electrical Manufacturers Association NU 2-2012
title_short Performance characteristic evaluation of a bismuth germanate-based high-sensitivity 5-ring discovery image quality positron emission tomography/computed tomography system as per National Electrical Manufacturers Association NU 2-2012
title_sort performance characteristic evaluation of a bismuth germanate-based high-sensitivity 5-ring discovery image quality positron emission tomography/computed tomography system as per national electrical manufacturers association nu 2-2012
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6945355/
https://www.ncbi.nlm.nih.gov/pubmed/31933550
http://dx.doi.org/10.4103/wjnm.WJNM_72_18
work_keys_str_mv AT jhaashishkumar performancecharacteristicevaluationofabismuthgermanatebasedhighsensitivity5ringdiscoveryimagequalitypositronemissiontomographycomputedtomographysystemaspernationalelectricalmanufacturersassociationnu22012
AT mithunsneha performancecharacteristicevaluationofabismuthgermanatebasedhighsensitivity5ringdiscoveryimagequalitypositronemissiontomographycomputedtomographysystemaspernationalelectricalmanufacturersassociationnu22012
AT puranikameyad performancecharacteristicevaluationofabismuthgermanatebasedhighsensitivity5ringdiscoveryimagequalitypositronemissiontomographycomputedtomographysystemaspernationalelectricalmanufacturersassociationnu22012
AT purandarenilenduc performancecharacteristicevaluationofabismuthgermanatebasedhighsensitivity5ringdiscoveryimagequalitypositronemissiontomographycomputedtomographysystemaspernationalelectricalmanufacturersassociationnu22012
AT shahsneha performancecharacteristicevaluationofabismuthgermanatebasedhighsensitivity5ringdiscoveryimagequalitypositronemissiontomographycomputedtomographysystemaspernationalelectricalmanufacturersassociationnu22012
AT agrawalarchi performancecharacteristicevaluationofabismuthgermanatebasedhighsensitivity5ringdiscoveryimagequalitypositronemissiontomographycomputedtomographysystemaspernationalelectricalmanufacturersassociationnu22012
AT rangarajanvenkatesh performancecharacteristicevaluationofabismuthgermanatebasedhighsensitivity5ringdiscoveryimagequalitypositronemissiontomographycomputedtomographysystemaspernationalelectricalmanufacturersassociationnu22012