Technical and clinical evaluation of a closed loop TIVA system with SEDLine(TM) spectral density monitoring: Multicentric prospective cohort study

INTRODUCTION: Closed loop total intravenous anesthesia is a technique in which the patient’s hemodynamic and anesthetic depth variables are monitored, and based on this information, a computer controls the infusion rate of drugs to keep them within pre-established clinical parameters. OBJECTIVE: To...

Descripción completa

Detalles Bibliográficos
Autores principales: Castellanos Peñaranda, Claudia, Casas Arroyave, Fabián D., Gómez, Francisco J., Pinzón Corredor, Paola A., Fernández, Juan M., Velez Botero, Marcela, Bohórquez Bedoya, Juan D., Marulanda Toro, Carlos
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6950910/
https://www.ncbi.nlm.nih.gov/pubmed/31921411
http://dx.doi.org/10.1186/s13741-019-0130-2
_version_ 1783486179547545600
author Castellanos Peñaranda, Claudia
Casas Arroyave, Fabián D.
Gómez, Francisco J.
Pinzón Corredor, Paola A.
Fernández, Juan M.
Velez Botero, Marcela
Bohórquez Bedoya, Juan D.
Marulanda Toro, Carlos
author_facet Castellanos Peñaranda, Claudia
Casas Arroyave, Fabián D.
Gómez, Francisco J.
Pinzón Corredor, Paola A.
Fernández, Juan M.
Velez Botero, Marcela
Bohórquez Bedoya, Juan D.
Marulanda Toro, Carlos
author_sort Castellanos Peñaranda, Claudia
collection PubMed
description INTRODUCTION: Closed loop total intravenous anesthesia is a technique in which the patient’s hemodynamic and anesthetic depth variables are monitored, and based on this information, a computer controls the infusion rate of drugs to keep them within pre-established clinical parameters. OBJECTIVE: To describe the technical and clinical performance of a closed loop system for total intravenous anesthesia with propofol and remifentanil, using the SEDLine(TM) monitor DESIGN: Multicentric prospective cohort study SETTING: Surgery room PATIENTS: ASA I-II undergoing elective surgery MEASUREMENTS: The authors designed a closed loop system that implements a control algorithm based on anesthetic depth monitoring and the Patient State Index (PSI(TM)) of the SEDLine monitor for propofol, and on hemodynamic variables for remifentanil. The measurement of clinical performance was made based on the percentage of PSI(TM) maintenance time in the range 20–50. Precision analysis was evaluated by measuring median performance error (MDPE) can be defined as the median difference between actual and desired values, which refers to the degree of precision in which the controller is able to maintain the control variable within the objective set by the anesthesiologist; it represents the direction (over-prediction or under-prediction) of performance error (PE) rather than size of errors, which is represented by MDAPE, median absolute percentage error, Wobble index, which is used for measuring the intrasubject variability in performance error. RESULTS: Data were obtained from 93 patients in three healthcare centers. The percentage of PSI(TM) maintenance time in the 20–50 range was 92% (80.7–97.0). MDPE was 10.7 (− 11.0–18.0), MDAPE 21.0 (14.2–26.8) and wobble 10.7 (7.0–16.9). No adverse surgical or anesthetic events were found. CONCLUSIONS: The closed loop total intravenous anesthesia system with SEDLine developed by the authors was used without major complication and appear to be feasible its use in clinical performance.
format Online
Article
Text
id pubmed-6950910
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-69509102020-01-09 Technical and clinical evaluation of a closed loop TIVA system with SEDLine(TM) spectral density monitoring: Multicentric prospective cohort study Castellanos Peñaranda, Claudia Casas Arroyave, Fabián D. Gómez, Francisco J. Pinzón Corredor, Paola A. Fernández, Juan M. Velez Botero, Marcela Bohórquez Bedoya, Juan D. Marulanda Toro, Carlos Perioper Med (Lond) Research INTRODUCTION: Closed loop total intravenous anesthesia is a technique in which the patient’s hemodynamic and anesthetic depth variables are monitored, and based on this information, a computer controls the infusion rate of drugs to keep them within pre-established clinical parameters. OBJECTIVE: To describe the technical and clinical performance of a closed loop system for total intravenous anesthesia with propofol and remifentanil, using the SEDLine(TM) monitor DESIGN: Multicentric prospective cohort study SETTING: Surgery room PATIENTS: ASA I-II undergoing elective surgery MEASUREMENTS: The authors designed a closed loop system that implements a control algorithm based on anesthetic depth monitoring and the Patient State Index (PSI(TM)) of the SEDLine monitor for propofol, and on hemodynamic variables for remifentanil. The measurement of clinical performance was made based on the percentage of PSI(TM) maintenance time in the range 20–50. Precision analysis was evaluated by measuring median performance error (MDPE) can be defined as the median difference between actual and desired values, which refers to the degree of precision in which the controller is able to maintain the control variable within the objective set by the anesthesiologist; it represents the direction (over-prediction or under-prediction) of performance error (PE) rather than size of errors, which is represented by MDAPE, median absolute percentage error, Wobble index, which is used for measuring the intrasubject variability in performance error. RESULTS: Data were obtained from 93 patients in three healthcare centers. The percentage of PSI(TM) maintenance time in the 20–50 range was 92% (80.7–97.0). MDPE was 10.7 (− 11.0–18.0), MDAPE 21.0 (14.2–26.8) and wobble 10.7 (7.0–16.9). No adverse surgical or anesthetic events were found. CONCLUSIONS: The closed loop total intravenous anesthesia system with SEDLine developed by the authors was used without major complication and appear to be feasible its use in clinical performance. BioMed Central 2020-01-09 /pmc/articles/PMC6950910/ /pubmed/31921411 http://dx.doi.org/10.1186/s13741-019-0130-2 Text en © The Author(s). 2020 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research
Castellanos Peñaranda, Claudia
Casas Arroyave, Fabián D.
Gómez, Francisco J.
Pinzón Corredor, Paola A.
Fernández, Juan M.
Velez Botero, Marcela
Bohórquez Bedoya, Juan D.
Marulanda Toro, Carlos
Technical and clinical evaluation of a closed loop TIVA system with SEDLine(TM) spectral density monitoring: Multicentric prospective cohort study
title Technical and clinical evaluation of a closed loop TIVA system with SEDLine(TM) spectral density monitoring: Multicentric prospective cohort study
title_full Technical and clinical evaluation of a closed loop TIVA system with SEDLine(TM) spectral density monitoring: Multicentric prospective cohort study
title_fullStr Technical and clinical evaluation of a closed loop TIVA system with SEDLine(TM) spectral density monitoring: Multicentric prospective cohort study
title_full_unstemmed Technical and clinical evaluation of a closed loop TIVA system with SEDLine(TM) spectral density monitoring: Multicentric prospective cohort study
title_short Technical and clinical evaluation of a closed loop TIVA system with SEDLine(TM) spectral density monitoring: Multicentric prospective cohort study
title_sort technical and clinical evaluation of a closed loop tiva system with sedline(tm) spectral density monitoring: multicentric prospective cohort study
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6950910/
https://www.ncbi.nlm.nih.gov/pubmed/31921411
http://dx.doi.org/10.1186/s13741-019-0130-2
work_keys_str_mv AT castellanospenarandaclaudia technicalandclinicalevaluationofaclosedlooptivasystemwithsedlinetmspectraldensitymonitoringmulticentricprospectivecohortstudy
AT casasarroyavefabiand technicalandclinicalevaluationofaclosedlooptivasystemwithsedlinetmspectraldensitymonitoringmulticentricprospectivecohortstudy
AT gomezfranciscoj technicalandclinicalevaluationofaclosedlooptivasystemwithsedlinetmspectraldensitymonitoringmulticentricprospectivecohortstudy
AT pinzoncorredorpaolaa technicalandclinicalevaluationofaclosedlooptivasystemwithsedlinetmspectraldensitymonitoringmulticentricprospectivecohortstudy
AT fernandezjuanm technicalandclinicalevaluationofaclosedlooptivasystemwithsedlinetmspectraldensitymonitoringmulticentricprospectivecohortstudy
AT velezboteromarcela technicalandclinicalevaluationofaclosedlooptivasystemwithsedlinetmspectraldensitymonitoringmulticentricprospectivecohortstudy
AT bohorquezbedoyajuand technicalandclinicalevaluationofaclosedlooptivasystemwithsedlinetmspectraldensitymonitoringmulticentricprospectivecohortstudy
AT marulandatorocarlos technicalandclinicalevaluationofaclosedlooptivasystemwithsedlinetmspectraldensitymonitoringmulticentricprospectivecohortstudy