Cargando…
Comparison of transesophageal and intracardiac echocardiography in guiding percutaneous left atrial appendage closure with an Amplatzer Amulet device
INTRODUCTION: Percutaneous occlusion of the left atrial appendage (LAAO) is becoming an extensively used method of stroke prevention in individuals with contraindications to oral anticoagulants. Transesophageal echocardiography (TOE) is the gold standard for LAAO guiding, but intracardiac echocardio...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Termedia Publishing House
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6956456/ https://www.ncbi.nlm.nih.gov/pubmed/31933661 http://dx.doi.org/10.5114/aic.2019.90219 |
_version_ | 1783487154972786688 |
---|---|
author | Streb, Witold Mitręga, Katarzyna Podolecki, Tomasz Morawski, Stanisław Szulik, Mariola Jędrzejczyk-Patej, Ewa Kukulski, Tomasz Kalarus, Zbigniew |
author_facet | Streb, Witold Mitręga, Katarzyna Podolecki, Tomasz Morawski, Stanisław Szulik, Mariola Jędrzejczyk-Patej, Ewa Kukulski, Tomasz Kalarus, Zbigniew |
author_sort | Streb, Witold |
collection | PubMed |
description | INTRODUCTION: Percutaneous occlusion of the left atrial appendage (LAAO) is becoming an extensively used method of stroke prevention in individuals with contraindications to oral anticoagulants. Transesophageal echocardiography (TOE) is the gold standard for LAAO guiding, but intracardiac echocardiography (ICE) appears to be a potential alternative. AIM: To compare the LAAO procedure guided by TOE or ICE with respect to procedural success and safety. MATERIAL AND METHODS: TOE-guided LAAO was performed in 12 patients and ICE-guided LAAO in 11 patients. ICE was performed using an 8F AcuNav probe and the ACUSON SC2000 system. For LAAO the Amplatzer Amulet was used. After 1 month TOE was performed. RESULTS: Procedural success was achieved in all patients in TOE and ICE groups. There was 1 complication (groin hematoma). The procedure time was significantly longer in the TOE group (43 to 80 min; median: 54 min) compared to the ICE group (28 to 67 min; median: 45 min), (p = 0.02) The time needed to puncture the interatrial septum and time needed to remove the sheath did not differ between groups. Fluoroscopic time was insignificantly longer in the ICE group (9.91 ±4.01s) compared to the TOE group (7.69 ±3.21s), and a significantly larger contrast media volume was used in the ICE group (30.00 ±6.67 ml vs. 40.45 ±23.18 ml, p = 0.03). There were no statistically significant differences in the results between TOE and ICE groups in follow-up assessments. CONCLUSIONS: LAAO using the Amplatzer Amulet may be successfully and safely guided by ICE. ICE offered shorter procedure time and similar results irrespectively of left atrial appendage anatomy compared to TOE guidance. |
format | Online Article Text |
id | pubmed-6956456 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | Termedia Publishing House |
record_format | MEDLINE/PubMed |
spelling | pubmed-69564562020-01-13 Comparison of transesophageal and intracardiac echocardiography in guiding percutaneous left atrial appendage closure with an Amplatzer Amulet device Streb, Witold Mitręga, Katarzyna Podolecki, Tomasz Morawski, Stanisław Szulik, Mariola Jędrzejczyk-Patej, Ewa Kukulski, Tomasz Kalarus, Zbigniew Postepy Kardiol Interwencyjnej Original Paper INTRODUCTION: Percutaneous occlusion of the left atrial appendage (LAAO) is becoming an extensively used method of stroke prevention in individuals with contraindications to oral anticoagulants. Transesophageal echocardiography (TOE) is the gold standard for LAAO guiding, but intracardiac echocardiography (ICE) appears to be a potential alternative. AIM: To compare the LAAO procedure guided by TOE or ICE with respect to procedural success and safety. MATERIAL AND METHODS: TOE-guided LAAO was performed in 12 patients and ICE-guided LAAO in 11 patients. ICE was performed using an 8F AcuNav probe and the ACUSON SC2000 system. For LAAO the Amplatzer Amulet was used. After 1 month TOE was performed. RESULTS: Procedural success was achieved in all patients in TOE and ICE groups. There was 1 complication (groin hematoma). The procedure time was significantly longer in the TOE group (43 to 80 min; median: 54 min) compared to the ICE group (28 to 67 min; median: 45 min), (p = 0.02) The time needed to puncture the interatrial septum and time needed to remove the sheath did not differ between groups. Fluoroscopic time was insignificantly longer in the ICE group (9.91 ±4.01s) compared to the TOE group (7.69 ±3.21s), and a significantly larger contrast media volume was used in the ICE group (30.00 ±6.67 ml vs. 40.45 ±23.18 ml, p = 0.03). There were no statistically significant differences in the results between TOE and ICE groups in follow-up assessments. CONCLUSIONS: LAAO using the Amplatzer Amulet may be successfully and safely guided by ICE. ICE offered shorter procedure time and similar results irrespectively of left atrial appendage anatomy compared to TOE guidance. Termedia Publishing House 2019-12-08 2019 /pmc/articles/PMC6956456/ /pubmed/31933661 http://dx.doi.org/10.5114/aic.2019.90219 Text en Copyright: © 2019 Termedia Sp. z o. o. http://creativecommons.org/licenses/by-nc-sa/4.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0) License, allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material, provided the original work is properly cited and states its license. |
spellingShingle | Original Paper Streb, Witold Mitręga, Katarzyna Podolecki, Tomasz Morawski, Stanisław Szulik, Mariola Jędrzejczyk-Patej, Ewa Kukulski, Tomasz Kalarus, Zbigniew Comparison of transesophageal and intracardiac echocardiography in guiding percutaneous left atrial appendage closure with an Amplatzer Amulet device |
title | Comparison of transesophageal and intracardiac echocardiography in guiding percutaneous left atrial appendage closure with an Amplatzer Amulet device |
title_full | Comparison of transesophageal and intracardiac echocardiography in guiding percutaneous left atrial appendage closure with an Amplatzer Amulet device |
title_fullStr | Comparison of transesophageal and intracardiac echocardiography in guiding percutaneous left atrial appendage closure with an Amplatzer Amulet device |
title_full_unstemmed | Comparison of transesophageal and intracardiac echocardiography in guiding percutaneous left atrial appendage closure with an Amplatzer Amulet device |
title_short | Comparison of transesophageal and intracardiac echocardiography in guiding percutaneous left atrial appendage closure with an Amplatzer Amulet device |
title_sort | comparison of transesophageal and intracardiac echocardiography in guiding percutaneous left atrial appendage closure with an amplatzer amulet device |
topic | Original Paper |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6956456/ https://www.ncbi.nlm.nih.gov/pubmed/31933661 http://dx.doi.org/10.5114/aic.2019.90219 |
work_keys_str_mv | AT strebwitold comparisonoftransesophagealandintracardiacechocardiographyinguidingpercutaneousleftatrialappendageclosurewithanamplatzeramuletdevice AT mitregakatarzyna comparisonoftransesophagealandintracardiacechocardiographyinguidingpercutaneousleftatrialappendageclosurewithanamplatzeramuletdevice AT podoleckitomasz comparisonoftransesophagealandintracardiacechocardiographyinguidingpercutaneousleftatrialappendageclosurewithanamplatzeramuletdevice AT morawskistanisław comparisonoftransesophagealandintracardiacechocardiographyinguidingpercutaneousleftatrialappendageclosurewithanamplatzeramuletdevice AT szulikmariola comparisonoftransesophagealandintracardiacechocardiographyinguidingpercutaneousleftatrialappendageclosurewithanamplatzeramuletdevice AT jedrzejczykpatejewa comparisonoftransesophagealandintracardiacechocardiographyinguidingpercutaneousleftatrialappendageclosurewithanamplatzeramuletdevice AT kukulskitomasz comparisonoftransesophagealandintracardiacechocardiographyinguidingpercutaneousleftatrialappendageclosurewithanamplatzeramuletdevice AT kalaruszbigniew comparisonoftransesophagealandintracardiacechocardiographyinguidingpercutaneousleftatrialappendageclosurewithanamplatzeramuletdevice |