Cargando…

Technical Note: Investigation of the dosimetric impact of stray radiation on the Common Control Unit of the IBA Blue Phantom(2)

PURPOSE: This technical note aims to investigate the dosimetric impact of stray radiation on the Common Control Unit (CCU) of the IBA Blue Phantom(2) and the measured beam data. METHODS: Three CCUs of the same model were used for the study. The primary test CCU was placed at five distances from the...

Descripción completa

Detalles Bibliográficos
Autores principales: Cui, Guoqiang, Duan, Jun, Yang, Yun, Yin, Fang‐Fang
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6964763/
https://www.ncbi.nlm.nih.gov/pubmed/31729812
http://dx.doi.org/10.1002/acm2.12769
Descripción
Sumario:PURPOSE: This technical note aims to investigate the dosimetric impact of stray radiation on the Common Control Unit (CCU) of the IBA Blue Phantom(2) and the measured beam data. METHODS: Three CCUs of the same model were used for the study. The primary test CCU was placed at five distances from the radiation beam central axis. At each distance, a set of depth dose and beam profiles for two open and two wedge fields were measured. The field sizes were 10 × 10 cm(2) and 30 × 30 cm(2) for the open fields, and 30 × 30 cm(2) and 15 × 15 cm(2) for the 30° and 60° wedges, respectively. The other two CCUs were used to cross check the data of the primary CCU. Assuming the effect of stray radiation on the data measured at the farthest reachable distance 4.5 m is negligible, the dosimetric impact of stray radiation on the CCU and consequently on the measured data can be extracted for analysis by comparing it with those measured at shorter distances. RESULTS: The results of three CCUs were consistent. The dosimetric impact of stray radiation was greater for lower energies at larger field sizes. For open fields, the data variation was up to 4.5% for depth dose curves and 7.1% for beam profiles. For wedge fields, the data variation was up to 9.3% for depth dose curves and 10.6% for beam profiles. Moreover, for wedge field profiles in the wedge direction, they became flatter as the CCU was placed closer to the primary radiation beam, manifesting smaller wedge angles. CONCLUSION: The stray radiation added a uniform background noise on all measured data. The magnitude of the noise is inversely proportional to the square of the distance of the CCU to the primary radiation beam, approximately following the inverse square law.