Cargando…
Monitoring linear accelerators electron beam energy constancy with a 2D ionization chamber array and double‐wedge phantom
Validate that a two‐dimensional (2D) ionization chamber array (ICA) combined with a double‐wedge plate (DWP) can track changes in electron beam energy well within 2.0 mms as recommended by TG‐142 for monthly quality assurance (QA). Electron beam profiles of 4–22 MeV were measured for a 25 × 25 cm(2)...
Autores principales: | Gao, Song, Chetvertkov, Mikhail A., Simon, William E., Sadeghi, Amir, Balter, Peter A. |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6964766/ https://www.ncbi.nlm.nih.gov/pubmed/31633877 http://dx.doi.org/10.1002/acm2.12751 |
Ejemplares similares
-
Electron energy constancy verification using a double‐wedge phantom
por: Wells, Derek M., et al.
Publicado: (2003) -
Quantification of beam steering with an ionization chamber array
por: Gao, Song, et al.
Publicado: (2018) -
Beam energy metrics for the acceptance and quality assurance of Halcyon linear accelerator
por: Gao, Song, et al.
Publicado: (2021) -
A comparison of methods for monitoring photon beam energy constancy
por: Gao, Song, et al.
Publicado: (2016) -
Acceptance and verification of the Halcyon‐Eclipse linear accelerator‐treatment planning system without 3D water scanning system
por: Gao, Song, et al.
Publicado: (2019)