Cargando…
Effects of image distortion and Hounsfield unit variations on radiation treatment plans: An extended field-of-view reconstruction in a large bore CT scanner
This study aimed to evaluate the effect of image distortion and Hounsfield unit (HU) variation due to the extended field-of-view (eFOV) of the large-bore (LB) computed tomography (CT) on dose distribution. Both home-made inhomogeneity and breast phantoms were scanned at the geometric center position...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Nature Publishing Group UK
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6965617/ https://www.ncbi.nlm.nih.gov/pubmed/31949301 http://dx.doi.org/10.1038/s41598-020-57422-y |
_version_ | 1783488652140085248 |
---|---|
author | Bae, Yong-Ki Lee, Jeong-Woo Hong, Semie |
author_facet | Bae, Yong-Ki Lee, Jeong-Woo Hong, Semie |
author_sort | Bae, Yong-Ki |
collection | PubMed |
description | This study aimed to evaluate the effect of image distortion and Hounsfield unit (HU) variation due to the extended field-of-view (eFOV) of the large-bore (LB) computed tomography (CT) on dose distribution. Both home-made inhomogeneity and breast phantoms were scanned at the geometric center position and four different offset positions. We also performed dose optimizations based on different breast phantom CT sets for evaluating the effects of image artifacts on the intensity-modulated radiation techniques. The volume changes were 0.0% to 0.5% in the air, −0.5% to 3.0% in the water, and 4.0% to 5.0% in the high-density material of the inhomogeneity phantom. Both phantoms scanning results indicate that more distortions occurred in the eFOV area due to the biased scanning center. The gamma index differences ranged from 0.87% to 4.87% for the FIF plan and from 0.52% to 6.26% for the VMAT plan. This resulted in decrease of the minimum (7.3–13.1%), maximum (−0.8–2.2%), and mean doses (−0.2–4.4%). We recommend that it should be evaluated whether the applied CT would have an appropriate eFOV range for clinical radiation treatment planning for patients. |
format | Online Article Text |
id | pubmed-6965617 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | Nature Publishing Group UK |
record_format | MEDLINE/PubMed |
spelling | pubmed-69656172020-01-23 Effects of image distortion and Hounsfield unit variations on radiation treatment plans: An extended field-of-view reconstruction in a large bore CT scanner Bae, Yong-Ki Lee, Jeong-Woo Hong, Semie Sci Rep Article This study aimed to evaluate the effect of image distortion and Hounsfield unit (HU) variation due to the extended field-of-view (eFOV) of the large-bore (LB) computed tomography (CT) on dose distribution. Both home-made inhomogeneity and breast phantoms were scanned at the geometric center position and four different offset positions. We also performed dose optimizations based on different breast phantom CT sets for evaluating the effects of image artifacts on the intensity-modulated radiation techniques. The volume changes were 0.0% to 0.5% in the air, −0.5% to 3.0% in the water, and 4.0% to 5.0% in the high-density material of the inhomogeneity phantom. Both phantoms scanning results indicate that more distortions occurred in the eFOV area due to the biased scanning center. The gamma index differences ranged from 0.87% to 4.87% for the FIF plan and from 0.52% to 6.26% for the VMAT plan. This resulted in decrease of the minimum (7.3–13.1%), maximum (−0.8–2.2%), and mean doses (−0.2–4.4%). We recommend that it should be evaluated whether the applied CT would have an appropriate eFOV range for clinical radiation treatment planning for patients. Nature Publishing Group UK 2020-01-16 /pmc/articles/PMC6965617/ /pubmed/31949301 http://dx.doi.org/10.1038/s41598-020-57422-y Text en © The Author(s) 2020 Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/. |
spellingShingle | Article Bae, Yong-Ki Lee, Jeong-Woo Hong, Semie Effects of image distortion and Hounsfield unit variations on radiation treatment plans: An extended field-of-view reconstruction in a large bore CT scanner |
title | Effects of image distortion and Hounsfield unit variations on radiation treatment plans: An extended field-of-view reconstruction in a large bore CT scanner |
title_full | Effects of image distortion and Hounsfield unit variations on radiation treatment plans: An extended field-of-view reconstruction in a large bore CT scanner |
title_fullStr | Effects of image distortion and Hounsfield unit variations on radiation treatment plans: An extended field-of-view reconstruction in a large bore CT scanner |
title_full_unstemmed | Effects of image distortion and Hounsfield unit variations on radiation treatment plans: An extended field-of-view reconstruction in a large bore CT scanner |
title_short | Effects of image distortion and Hounsfield unit variations on radiation treatment plans: An extended field-of-view reconstruction in a large bore CT scanner |
title_sort | effects of image distortion and hounsfield unit variations on radiation treatment plans: an extended field-of-view reconstruction in a large bore ct scanner |
topic | Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6965617/ https://www.ncbi.nlm.nih.gov/pubmed/31949301 http://dx.doi.org/10.1038/s41598-020-57422-y |
work_keys_str_mv | AT baeyongki effectsofimagedistortionandhounsfieldunitvariationsonradiationtreatmentplansanextendedfieldofviewreconstructioninalargeborectscanner AT leejeongwoo effectsofimagedistortionandhounsfieldunitvariationsonradiationtreatmentplansanextendedfieldofviewreconstructioninalargeborectscanner AT hongsemie effectsofimagedistortionandhounsfieldunitvariationsonradiationtreatmentplansanextendedfieldofviewreconstructioninalargeborectscanner |