Preservation of residual hearing after cochlear implant surgery with slim modiolar electrode
PURPOSE: To evaluate the insertion results and hearing preservation of a novel slim modiolar electrode (SME) in patients with residual hearing. METHODS: We retrospectively collected the data from the medical files of 17 patients (18 ears) implanted with a SME. All patients had functional low frequen...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer Berlin Heidelberg
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6981311/ https://www.ncbi.nlm.nih.gov/pubmed/31673779 http://dx.doi.org/10.1007/s00405-019-05708-x |
Sumario: | PURPOSE: To evaluate the insertion results and hearing preservation of a novel slim modiolar electrode (SME) in patients with residual hearing. METHODS: We retrospectively collected the data from the medical files of 17 patients (18 ears) implanted with a SME. All patients had functional low frequency hearing (PTA ((0.125–0.5 kHz)) ≤ 80 dB HL). The insertion results were re-examined from the postoperative cone-beam computed tomography scans. Postoperative thresholds were obtained at the time of switch-on of the sound processors (mean 43 days) and at latest follow-up (mean 582 days). The speech recognition in noise was measured with the Finnish matrix sentence test preoperatively and at follow-up. RESULTS: The mean insertion depth angle (IDA) was 395°. Neither scala dislocations nor tip fold over were detected. There were no total hearing losses. Functional low-frequency hearing was preserved in 15/18 (83%) ears at switch-on and in 14/17 (82%) ears at follow-up. According to HEARRING classification, 55% (10/18) had complete HP at switch-on and 41% (7/17) still at follow-up. Thirteen patients (14 ears) were initially fitted with electric–acoustic stimulation and seven patients (8 ears) continued to use it after follow-up. CONCLUSIONS: The preliminary hearing preservation results with the SME were more favorable than reported for other perimodiolar electrodes. The results show that the array may also be feasible for electro-acoustic stimulation; it is beneficial in that it provides adequate cochlear coverage for pure electrical stimulation in the event of postoperative or progressive hearing loss. |
---|