Accuracy and Precision of Three Consumer-Grade Motion Sensors During Overground and Treadmill Walking in People With Parkinson Disease: Cross-Sectional Comparative Study
BACKGROUND: Wearable motion sensors are gaining popularity for monitoring free-living physical activity among people with Parkinson disease (PD), but more evidence supporting the accuracy and precision of motion sensors for capturing step counts is required in people with PD. OBJECTIVE: This study a...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
JMIR Publications
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6996761/ https://www.ncbi.nlm.nih.gov/pubmed/31944182 http://dx.doi.org/10.2196/14059 |
_version_ | 1783493570154463232 |
---|---|
author | Lai, Byron Sasaki, Jeffer E Jeng, Brenda Cederberg, Katie L Bamman, Marcas M Motl, Robert W |
author_facet | Lai, Byron Sasaki, Jeffer E Jeng, Brenda Cederberg, Katie L Bamman, Marcas M Motl, Robert W |
author_sort | Lai, Byron |
collection | PubMed |
description | BACKGROUND: Wearable motion sensors are gaining popularity for monitoring free-living physical activity among people with Parkinson disease (PD), but more evidence supporting the accuracy and precision of motion sensors for capturing step counts is required in people with PD. OBJECTIVE: This study aimed to examine the accuracy and precision of 3 common consumer-grade motion sensors for measuring actual steps taken during prolonged periods of overground and treadmill walking in people with PD. METHODS: A total of 31 ambulatory participants with PD underwent 6-min bouts of overground and treadmill walking at a comfortable speed. Participants wore 3 devices (Garmin Vivosmart 3, Fitbit One, and Fitbit Charge 2 HR), and a single researcher manually counted the actual steps taken. Accuracy and precision were based on absolute and relative metrics, including intraclass correlation coefficients (ICCs) and Bland-Altman plots. RESULTS: Participants walked 628 steps over ground based on manual counting, and Garmin Vivosmart, Fitbit One, and Fitbit Charge 2 HR devices had absolute (relative) error values of 6 (6/628, 1.0%), 8 (8/628, 1.3%), and 30 (30/628, 4.8%) steps, respectively. ICC values demonstrated excellent agreement between manually counted steps and steps counted by both Garmin Vivosmart (0.97) and Fitbit One (0.98) but poor agreement for Fitbit Charge 2 HR (0.47). The absolute (relative) precision values for Garmin Vivosmart, Fitbit One, and Fitbit Charge 2 HR were 11.1 (11.1/625, 1.8%), 14.7 (14.7/620, 2.4%), and 74.4 (74.4/598, 12.4%) steps, respectively. ICC confidence intervals demonstrated low variability for Garmin Vivosmart (0.96 to 0.99) and Fitbit One (0.93 to 0.99) but high variability for Fitbit Charge 2 HR (–0.57 to 0.74). The Fitbit One device maintained high accuracy and precision values for treadmill walking, but both Garmin Vivosmart and Fitbit Charge 2 HR (the wrist-worn devices) had worse accuracy and precision for treadmill walking. CONCLUSIONS: The waist-worn sensor (Fitbit One) was accurate and precise in measuring steps with overground and treadmill walking. The wrist-worn sensors were accurate and precise only during overground walking. Similar research should inform the application of these devices in clinical research and practice involving patients with PD. |
format | Online Article Text |
id | pubmed-6996761 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | JMIR Publications |
record_format | MEDLINE/PubMed |
spelling | pubmed-69967612020-02-20 Accuracy and Precision of Three Consumer-Grade Motion Sensors During Overground and Treadmill Walking in People With Parkinson Disease: Cross-Sectional Comparative Study Lai, Byron Sasaki, Jeffer E Jeng, Brenda Cederberg, Katie L Bamman, Marcas M Motl, Robert W JMIR Rehabil Assist Technol Original Paper BACKGROUND: Wearable motion sensors are gaining popularity for monitoring free-living physical activity among people with Parkinson disease (PD), but more evidence supporting the accuracy and precision of motion sensors for capturing step counts is required in people with PD. OBJECTIVE: This study aimed to examine the accuracy and precision of 3 common consumer-grade motion sensors for measuring actual steps taken during prolonged periods of overground and treadmill walking in people with PD. METHODS: A total of 31 ambulatory participants with PD underwent 6-min bouts of overground and treadmill walking at a comfortable speed. Participants wore 3 devices (Garmin Vivosmart 3, Fitbit One, and Fitbit Charge 2 HR), and a single researcher manually counted the actual steps taken. Accuracy and precision were based on absolute and relative metrics, including intraclass correlation coefficients (ICCs) and Bland-Altman plots. RESULTS: Participants walked 628 steps over ground based on manual counting, and Garmin Vivosmart, Fitbit One, and Fitbit Charge 2 HR devices had absolute (relative) error values of 6 (6/628, 1.0%), 8 (8/628, 1.3%), and 30 (30/628, 4.8%) steps, respectively. ICC values demonstrated excellent agreement between manually counted steps and steps counted by both Garmin Vivosmart (0.97) and Fitbit One (0.98) but poor agreement for Fitbit Charge 2 HR (0.47). The absolute (relative) precision values for Garmin Vivosmart, Fitbit One, and Fitbit Charge 2 HR were 11.1 (11.1/625, 1.8%), 14.7 (14.7/620, 2.4%), and 74.4 (74.4/598, 12.4%) steps, respectively. ICC confidence intervals demonstrated low variability for Garmin Vivosmart (0.96 to 0.99) and Fitbit One (0.93 to 0.99) but high variability for Fitbit Charge 2 HR (–0.57 to 0.74). The Fitbit One device maintained high accuracy and precision values for treadmill walking, but both Garmin Vivosmart and Fitbit Charge 2 HR (the wrist-worn devices) had worse accuracy and precision for treadmill walking. CONCLUSIONS: The waist-worn sensor (Fitbit One) was accurate and precise in measuring steps with overground and treadmill walking. The wrist-worn sensors were accurate and precise only during overground walking. Similar research should inform the application of these devices in clinical research and practice involving patients with PD. JMIR Publications 2020-01-16 /pmc/articles/PMC6996761/ /pubmed/31944182 http://dx.doi.org/10.2196/14059 Text en ©Byron Lai, Jeffer E Sasaki, Brenda Jeng, Katie L Cederberg, Marcas M Bamman, Robert W Motl. Originally published in JMIR Rehabilitation and Assistive Technology (http://rehab.jmir.org), 16.01.2020. https://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR Rehabilitation and Assistive Technology, is properly cited. The complete bibliographic information, a link to the original publication on http://rehab.jmir.org/, as well as this copyright and license information must be included. |
spellingShingle | Original Paper Lai, Byron Sasaki, Jeffer E Jeng, Brenda Cederberg, Katie L Bamman, Marcas M Motl, Robert W Accuracy and Precision of Three Consumer-Grade Motion Sensors During Overground and Treadmill Walking in People With Parkinson Disease: Cross-Sectional Comparative Study |
title | Accuracy and Precision of Three Consumer-Grade Motion Sensors During Overground and Treadmill Walking in People With Parkinson Disease: Cross-Sectional Comparative Study |
title_full | Accuracy and Precision of Three Consumer-Grade Motion Sensors During Overground and Treadmill Walking in People With Parkinson Disease: Cross-Sectional Comparative Study |
title_fullStr | Accuracy and Precision of Three Consumer-Grade Motion Sensors During Overground and Treadmill Walking in People With Parkinson Disease: Cross-Sectional Comparative Study |
title_full_unstemmed | Accuracy and Precision of Three Consumer-Grade Motion Sensors During Overground and Treadmill Walking in People With Parkinson Disease: Cross-Sectional Comparative Study |
title_short | Accuracy and Precision of Three Consumer-Grade Motion Sensors During Overground and Treadmill Walking in People With Parkinson Disease: Cross-Sectional Comparative Study |
title_sort | accuracy and precision of three consumer-grade motion sensors during overground and treadmill walking in people with parkinson disease: cross-sectional comparative study |
topic | Original Paper |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6996761/ https://www.ncbi.nlm.nih.gov/pubmed/31944182 http://dx.doi.org/10.2196/14059 |
work_keys_str_mv | AT laibyron accuracyandprecisionofthreeconsumergrademotionsensorsduringovergroundandtreadmillwalkinginpeoplewithparkinsondiseasecrosssectionalcomparativestudy AT sasakijeffere accuracyandprecisionofthreeconsumergrademotionsensorsduringovergroundandtreadmillwalkinginpeoplewithparkinsondiseasecrosssectionalcomparativestudy AT jengbrenda accuracyandprecisionofthreeconsumergrademotionsensorsduringovergroundandtreadmillwalkinginpeoplewithparkinsondiseasecrosssectionalcomparativestudy AT cederbergkatiel accuracyandprecisionofthreeconsumergrademotionsensorsduringovergroundandtreadmillwalkinginpeoplewithparkinsondiseasecrosssectionalcomparativestudy AT bammanmarcasm accuracyandprecisionofthreeconsumergrademotionsensorsduringovergroundandtreadmillwalkinginpeoplewithparkinsondiseasecrosssectionalcomparativestudy AT motlrobertw accuracyandprecisionofthreeconsumergrademotionsensorsduringovergroundandtreadmillwalkinginpeoplewithparkinsondiseasecrosssectionalcomparativestudy |