In vitro Assessment of the Impacts of Leaflet Design on the Hemodynamic Characteristics of ePTFE Pulmonary Prosthetic Valves

Prosthetic pulmonary valves are widely used in the management procedures of various congenital heart diseases, including the surgical pulmonary valve replacement (PVR) and right ventricular outflow tract reconstruction (RVOT). The discouraging long-term outcomes of standard prostheses, including hom...

Descripción completa

Detalles Bibliográficos
Autores principales: Zhu, Guangyu, Wei, Yuan, Yuan, Qi, Cai, Li, Nakao, Masakazu, Yeo, Joon Hock
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Frontiers Media S.A. 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7006451/
https://www.ncbi.nlm.nih.gov/pubmed/32076599
http://dx.doi.org/10.3389/fbioe.2019.00477
_version_ 1783495146660167680
author Zhu, Guangyu
Wei, Yuan
Yuan, Qi
Cai, Li
Nakao, Masakazu
Yeo, Joon Hock
author_facet Zhu, Guangyu
Wei, Yuan
Yuan, Qi
Cai, Li
Nakao, Masakazu
Yeo, Joon Hock
author_sort Zhu, Guangyu
collection PubMed
description Prosthetic pulmonary valves are widely used in the management procedures of various congenital heart diseases, including the surgical pulmonary valve replacement (PVR) and right ventricular outflow tract reconstruction (RVOT). The discouraging long-term outcomes of standard prostheses, including homografts and bioprosthetic, constrained their indications. Recent developments in the expanded-polytetrafluoroethylene (ePTFE) pulmonary prosthetic valves provide promising alternatives. In this study, the hemodynamic characteristics of bileaflet and trileaflet ePTFE valve designs were experimentally evaluated. The in vitro tests were performed under the right ventricle (RV) flow conditions by using an in vitro RV circulatory system and particle image velocimetry (PIV). The leaflet kinetics, trans-valvular pressure gradients, effective orifice areas, regurgitant fractions, energy losses, velocity fields, and Reynolds shear stress (RSS) in both prostheses were evaluated. The opening of the bileaflet and trileaflet valve takes 0.060 and 0.088 s, respectively. The closing of the former takes 0.140 s, in contrast to 0.176 s of the latter. The trans-valvular pressure is 6.8 mmHg in the bileaflet valve vs. 7.9 mmHg in the trileaflet valve. The effective orifice area is 1.83 cm(2) in the bileaflet valve and 1.72 cm(2) in the trileaflet valve. The regurgitant fraction and energy loss of bileaflet are 7.13% and 82 mJ, which are 7.84% and 101.64 mJ in its bileaflet counterpart. The maximum RSS of 48.0 and 49.2 Pa occur at the systole peak in the bileaflet and trileaflet valve, respectively. A higher average RSS level is found in the bileaflet valve. The results from this preliminary study indicate that the current bileaflet prosthetic valve design is capable of providing a better overall hemodynamic performance than the trileaflet design.
format Online
Article
Text
id pubmed-7006451
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Frontiers Media S.A.
record_format MEDLINE/PubMed
spelling pubmed-70064512020-02-19 In vitro Assessment of the Impacts of Leaflet Design on the Hemodynamic Characteristics of ePTFE Pulmonary Prosthetic Valves Zhu, Guangyu Wei, Yuan Yuan, Qi Cai, Li Nakao, Masakazu Yeo, Joon Hock Front Bioeng Biotechnol Bioengineering and Biotechnology Prosthetic pulmonary valves are widely used in the management procedures of various congenital heart diseases, including the surgical pulmonary valve replacement (PVR) and right ventricular outflow tract reconstruction (RVOT). The discouraging long-term outcomes of standard prostheses, including homografts and bioprosthetic, constrained their indications. Recent developments in the expanded-polytetrafluoroethylene (ePTFE) pulmonary prosthetic valves provide promising alternatives. In this study, the hemodynamic characteristics of bileaflet and trileaflet ePTFE valve designs were experimentally evaluated. The in vitro tests were performed under the right ventricle (RV) flow conditions by using an in vitro RV circulatory system and particle image velocimetry (PIV). The leaflet kinetics, trans-valvular pressure gradients, effective orifice areas, regurgitant fractions, energy losses, velocity fields, and Reynolds shear stress (RSS) in both prostheses were evaluated. The opening of the bileaflet and trileaflet valve takes 0.060 and 0.088 s, respectively. The closing of the former takes 0.140 s, in contrast to 0.176 s of the latter. The trans-valvular pressure is 6.8 mmHg in the bileaflet valve vs. 7.9 mmHg in the trileaflet valve. The effective orifice area is 1.83 cm(2) in the bileaflet valve and 1.72 cm(2) in the trileaflet valve. The regurgitant fraction and energy loss of bileaflet are 7.13% and 82 mJ, which are 7.84% and 101.64 mJ in its bileaflet counterpart. The maximum RSS of 48.0 and 49.2 Pa occur at the systole peak in the bileaflet and trileaflet valve, respectively. A higher average RSS level is found in the bileaflet valve. The results from this preliminary study indicate that the current bileaflet prosthetic valve design is capable of providing a better overall hemodynamic performance than the trileaflet design. Frontiers Media S.A. 2020-01-31 /pmc/articles/PMC7006451/ /pubmed/32076599 http://dx.doi.org/10.3389/fbioe.2019.00477 Text en Copyright © 2020 Zhu, Wei, Yuan, Cai, Nakao and Yeo. http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
spellingShingle Bioengineering and Biotechnology
Zhu, Guangyu
Wei, Yuan
Yuan, Qi
Cai, Li
Nakao, Masakazu
Yeo, Joon Hock
In vitro Assessment of the Impacts of Leaflet Design on the Hemodynamic Characteristics of ePTFE Pulmonary Prosthetic Valves
title In vitro Assessment of the Impacts of Leaflet Design on the Hemodynamic Characteristics of ePTFE Pulmonary Prosthetic Valves
title_full In vitro Assessment of the Impacts of Leaflet Design on the Hemodynamic Characteristics of ePTFE Pulmonary Prosthetic Valves
title_fullStr In vitro Assessment of the Impacts of Leaflet Design on the Hemodynamic Characteristics of ePTFE Pulmonary Prosthetic Valves
title_full_unstemmed In vitro Assessment of the Impacts of Leaflet Design on the Hemodynamic Characteristics of ePTFE Pulmonary Prosthetic Valves
title_short In vitro Assessment of the Impacts of Leaflet Design on the Hemodynamic Characteristics of ePTFE Pulmonary Prosthetic Valves
title_sort in vitro assessment of the impacts of leaflet design on the hemodynamic characteristics of eptfe pulmonary prosthetic valves
topic Bioengineering and Biotechnology
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7006451/
https://www.ncbi.nlm.nih.gov/pubmed/32076599
http://dx.doi.org/10.3389/fbioe.2019.00477
work_keys_str_mv AT zhuguangyu invitroassessmentoftheimpactsofleafletdesignonthehemodynamiccharacteristicsofeptfepulmonaryprostheticvalves
AT weiyuan invitroassessmentoftheimpactsofleafletdesignonthehemodynamiccharacteristicsofeptfepulmonaryprostheticvalves
AT yuanqi invitroassessmentoftheimpactsofleafletdesignonthehemodynamiccharacteristicsofeptfepulmonaryprostheticvalves
AT caili invitroassessmentoftheimpactsofleafletdesignonthehemodynamiccharacteristicsofeptfepulmonaryprostheticvalves
AT nakaomasakazu invitroassessmentoftheimpactsofleafletdesignonthehemodynamiccharacteristicsofeptfepulmonaryprostheticvalves
AT yeojoonhock invitroassessmentoftheimpactsofleafletdesignonthehemodynamiccharacteristicsofeptfepulmonaryprostheticvalves