In vitro Assessment of the Impacts of Leaflet Design on the Hemodynamic Characteristics of ePTFE Pulmonary Prosthetic Valves
Prosthetic pulmonary valves are widely used in the management procedures of various congenital heart diseases, including the surgical pulmonary valve replacement (PVR) and right ventricular outflow tract reconstruction (RVOT). The discouraging long-term outcomes of standard prostheses, including hom...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Frontiers Media S.A.
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7006451/ https://www.ncbi.nlm.nih.gov/pubmed/32076599 http://dx.doi.org/10.3389/fbioe.2019.00477 |
_version_ | 1783495146660167680 |
---|---|
author | Zhu, Guangyu Wei, Yuan Yuan, Qi Cai, Li Nakao, Masakazu Yeo, Joon Hock |
author_facet | Zhu, Guangyu Wei, Yuan Yuan, Qi Cai, Li Nakao, Masakazu Yeo, Joon Hock |
author_sort | Zhu, Guangyu |
collection | PubMed |
description | Prosthetic pulmonary valves are widely used in the management procedures of various congenital heart diseases, including the surgical pulmonary valve replacement (PVR) and right ventricular outflow tract reconstruction (RVOT). The discouraging long-term outcomes of standard prostheses, including homografts and bioprosthetic, constrained their indications. Recent developments in the expanded-polytetrafluoroethylene (ePTFE) pulmonary prosthetic valves provide promising alternatives. In this study, the hemodynamic characteristics of bileaflet and trileaflet ePTFE valve designs were experimentally evaluated. The in vitro tests were performed under the right ventricle (RV) flow conditions by using an in vitro RV circulatory system and particle image velocimetry (PIV). The leaflet kinetics, trans-valvular pressure gradients, effective orifice areas, regurgitant fractions, energy losses, velocity fields, and Reynolds shear stress (RSS) in both prostheses were evaluated. The opening of the bileaflet and trileaflet valve takes 0.060 and 0.088 s, respectively. The closing of the former takes 0.140 s, in contrast to 0.176 s of the latter. The trans-valvular pressure is 6.8 mmHg in the bileaflet valve vs. 7.9 mmHg in the trileaflet valve. The effective orifice area is 1.83 cm(2) in the bileaflet valve and 1.72 cm(2) in the trileaflet valve. The regurgitant fraction and energy loss of bileaflet are 7.13% and 82 mJ, which are 7.84% and 101.64 mJ in its bileaflet counterpart. The maximum RSS of 48.0 and 49.2 Pa occur at the systole peak in the bileaflet and trileaflet valve, respectively. A higher average RSS level is found in the bileaflet valve. The results from this preliminary study indicate that the current bileaflet prosthetic valve design is capable of providing a better overall hemodynamic performance than the trileaflet design. |
format | Online Article Text |
id | pubmed-7006451 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | Frontiers Media S.A. |
record_format | MEDLINE/PubMed |
spelling | pubmed-70064512020-02-19 In vitro Assessment of the Impacts of Leaflet Design on the Hemodynamic Characteristics of ePTFE Pulmonary Prosthetic Valves Zhu, Guangyu Wei, Yuan Yuan, Qi Cai, Li Nakao, Masakazu Yeo, Joon Hock Front Bioeng Biotechnol Bioengineering and Biotechnology Prosthetic pulmonary valves are widely used in the management procedures of various congenital heart diseases, including the surgical pulmonary valve replacement (PVR) and right ventricular outflow tract reconstruction (RVOT). The discouraging long-term outcomes of standard prostheses, including homografts and bioprosthetic, constrained their indications. Recent developments in the expanded-polytetrafluoroethylene (ePTFE) pulmonary prosthetic valves provide promising alternatives. In this study, the hemodynamic characteristics of bileaflet and trileaflet ePTFE valve designs were experimentally evaluated. The in vitro tests were performed under the right ventricle (RV) flow conditions by using an in vitro RV circulatory system and particle image velocimetry (PIV). The leaflet kinetics, trans-valvular pressure gradients, effective orifice areas, regurgitant fractions, energy losses, velocity fields, and Reynolds shear stress (RSS) in both prostheses were evaluated. The opening of the bileaflet and trileaflet valve takes 0.060 and 0.088 s, respectively. The closing of the former takes 0.140 s, in contrast to 0.176 s of the latter. The trans-valvular pressure is 6.8 mmHg in the bileaflet valve vs. 7.9 mmHg in the trileaflet valve. The effective orifice area is 1.83 cm(2) in the bileaflet valve and 1.72 cm(2) in the trileaflet valve. The regurgitant fraction and energy loss of bileaflet are 7.13% and 82 mJ, which are 7.84% and 101.64 mJ in its bileaflet counterpart. The maximum RSS of 48.0 and 49.2 Pa occur at the systole peak in the bileaflet and trileaflet valve, respectively. A higher average RSS level is found in the bileaflet valve. The results from this preliminary study indicate that the current bileaflet prosthetic valve design is capable of providing a better overall hemodynamic performance than the trileaflet design. Frontiers Media S.A. 2020-01-31 /pmc/articles/PMC7006451/ /pubmed/32076599 http://dx.doi.org/10.3389/fbioe.2019.00477 Text en Copyright © 2020 Zhu, Wei, Yuan, Cai, Nakao and Yeo. http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms. |
spellingShingle | Bioengineering and Biotechnology Zhu, Guangyu Wei, Yuan Yuan, Qi Cai, Li Nakao, Masakazu Yeo, Joon Hock In vitro Assessment of the Impacts of Leaflet Design on the Hemodynamic Characteristics of ePTFE Pulmonary Prosthetic Valves |
title | In vitro Assessment of the Impacts of Leaflet Design on the Hemodynamic Characteristics of ePTFE Pulmonary Prosthetic Valves |
title_full | In vitro Assessment of the Impacts of Leaflet Design on the Hemodynamic Characteristics of ePTFE Pulmonary Prosthetic Valves |
title_fullStr | In vitro Assessment of the Impacts of Leaflet Design on the Hemodynamic Characteristics of ePTFE Pulmonary Prosthetic Valves |
title_full_unstemmed | In vitro Assessment of the Impacts of Leaflet Design on the Hemodynamic Characteristics of ePTFE Pulmonary Prosthetic Valves |
title_short | In vitro Assessment of the Impacts of Leaflet Design on the Hemodynamic Characteristics of ePTFE Pulmonary Prosthetic Valves |
title_sort | in vitro assessment of the impacts of leaflet design on the hemodynamic characteristics of eptfe pulmonary prosthetic valves |
topic | Bioengineering and Biotechnology |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7006451/ https://www.ncbi.nlm.nih.gov/pubmed/32076599 http://dx.doi.org/10.3389/fbioe.2019.00477 |
work_keys_str_mv | AT zhuguangyu invitroassessmentoftheimpactsofleafletdesignonthehemodynamiccharacteristicsofeptfepulmonaryprostheticvalves AT weiyuan invitroassessmentoftheimpactsofleafletdesignonthehemodynamiccharacteristicsofeptfepulmonaryprostheticvalves AT yuanqi invitroassessmentoftheimpactsofleafletdesignonthehemodynamiccharacteristicsofeptfepulmonaryprostheticvalves AT caili invitroassessmentoftheimpactsofleafletdesignonthehemodynamiccharacteristicsofeptfepulmonaryprostheticvalves AT nakaomasakazu invitroassessmentoftheimpactsofleafletdesignonthehemodynamiccharacteristicsofeptfepulmonaryprostheticvalves AT yeojoonhock invitroassessmentoftheimpactsofleafletdesignonthehemodynamiccharacteristicsofeptfepulmonaryprostheticvalves |