A High-Throughput Microfluidic Magnetic Separation (µFMS) Platform for Water Quality Monitoring
The long-term aim of this work is to develop a biosensing system that rapidly detects bacterial targets of interest, such as Escherichia coli, in drinking and recreational water quality monitoring. For these applications, a standard sample size is 100 mL, which is quite large for magnetic separation...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
MDPI
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7019623/ https://www.ncbi.nlm.nih.gov/pubmed/31877902 http://dx.doi.org/10.3390/mi11010016 |
Sumario: | The long-term aim of this work is to develop a biosensing system that rapidly detects bacterial targets of interest, such as Escherichia coli, in drinking and recreational water quality monitoring. For these applications, a standard sample size is 100 mL, which is quite large for magnetic separation microfluidic analysis platforms that typically function with <20 µL/s throughput. Here, we report the use of 1.5-µm-diameter magnetic microdisc to selectively tag target bacteria, and a high-throughput microfluidic device that can potentially isolate the magnetically tagged bacteria from 100 mL water samples in less than 15 min. Simulations and experiments show ~90% capture efficiencies of magnetic particles at flow rates up to 120 µL/s. Also, the platform enables the magnetic microdiscs/bacteria conjugates to be directly imaged, providing a path for quantitative assay. |
---|