Wedged Sensor in Distress? Lessons Learned from Troubleshooting Dampened Transmitted PA Waveforms of CardioMEMS Device
Introduction. Cardiovascular disease is a leading cause of morbidity and mortality with heart failure constituting a large portion of this spectrum. Heart failure patients have 90-day readmission rates of nearly 41% associated with a high expense. Numerous strategies to reduce readmissions have been...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Hindawi
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7036126/ https://www.ncbi.nlm.nih.gov/pubmed/32095291 http://dx.doi.org/10.1155/2020/3856940 |
_version_ | 1783500161320747008 |
---|---|
author | Singh, Robby Scarfone, Santo Zughaib, Marcel |
author_facet | Singh, Robby Scarfone, Santo Zughaib, Marcel |
author_sort | Singh, Robby |
collection | PubMed |
description | Introduction. Cardiovascular disease is a leading cause of morbidity and mortality with heart failure constituting a large portion of this spectrum. Heart failure patients have 90-day readmission rates of nearly 41% associated with a high expense. Numerous strategies to reduce readmissions have been attempted with the CardioMEMS pulmonary artery pressure monitoring system as one of the more successful ones. As this device becomes used more frequently, it is important to recognize procedural complications. We present of a rare complication where a patient underwent successful device placement and was subsequently found to have dampened waveforms which were due to device migration. Case Report. A 79-year-old male underwent successful CardioMEMS placement due to recurrent heart failure hospitalizations. 6 months later, the transmitted waveforms appeared dampened, and repeat angiography revealed a device that had migrated. Rather than abandoning the device, it was recalibrated and continued to transmit data and helped manage the patient's heart failure. Conclusion. CardioMEMS is a cost-effective tool to help reduce heart failure hospitalizations. Device migration is a rare complication and can lead to inaccurate data. However, as seen in this case, the device can be successfully recalibrated and can continue to be utilized to help reduce heart failure admissions. |
format | Online Article Text |
id | pubmed-7036126 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | Hindawi |
record_format | MEDLINE/PubMed |
spelling | pubmed-70361262020-02-24 Wedged Sensor in Distress? Lessons Learned from Troubleshooting Dampened Transmitted PA Waveforms of CardioMEMS Device Singh, Robby Scarfone, Santo Zughaib, Marcel Case Rep Cardiol Case Report Introduction. Cardiovascular disease is a leading cause of morbidity and mortality with heart failure constituting a large portion of this spectrum. Heart failure patients have 90-day readmission rates of nearly 41% associated with a high expense. Numerous strategies to reduce readmissions have been attempted with the CardioMEMS pulmonary artery pressure monitoring system as one of the more successful ones. As this device becomes used more frequently, it is important to recognize procedural complications. We present of a rare complication where a patient underwent successful device placement and was subsequently found to have dampened waveforms which were due to device migration. Case Report. A 79-year-old male underwent successful CardioMEMS placement due to recurrent heart failure hospitalizations. 6 months later, the transmitted waveforms appeared dampened, and repeat angiography revealed a device that had migrated. Rather than abandoning the device, it was recalibrated and continued to transmit data and helped manage the patient's heart failure. Conclusion. CardioMEMS is a cost-effective tool to help reduce heart failure hospitalizations. Device migration is a rare complication and can lead to inaccurate data. However, as seen in this case, the device can be successfully recalibrated and can continue to be utilized to help reduce heart failure admissions. Hindawi 2020-02-11 /pmc/articles/PMC7036126/ /pubmed/32095291 http://dx.doi.org/10.1155/2020/3856940 Text en Copyright © 2020 Robby Singh et al. http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Case Report Singh, Robby Scarfone, Santo Zughaib, Marcel Wedged Sensor in Distress? Lessons Learned from Troubleshooting Dampened Transmitted PA Waveforms of CardioMEMS Device |
title | Wedged Sensor in Distress? Lessons Learned from Troubleshooting Dampened Transmitted PA Waveforms of CardioMEMS Device |
title_full | Wedged Sensor in Distress? Lessons Learned from Troubleshooting Dampened Transmitted PA Waveforms of CardioMEMS Device |
title_fullStr | Wedged Sensor in Distress? Lessons Learned from Troubleshooting Dampened Transmitted PA Waveforms of CardioMEMS Device |
title_full_unstemmed | Wedged Sensor in Distress? Lessons Learned from Troubleshooting Dampened Transmitted PA Waveforms of CardioMEMS Device |
title_short | Wedged Sensor in Distress? Lessons Learned from Troubleshooting Dampened Transmitted PA Waveforms of CardioMEMS Device |
title_sort | wedged sensor in distress? lessons learned from troubleshooting dampened transmitted pa waveforms of cardiomems device |
topic | Case Report |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7036126/ https://www.ncbi.nlm.nih.gov/pubmed/32095291 http://dx.doi.org/10.1155/2020/3856940 |
work_keys_str_mv | AT singhrobby wedgedsensorindistresslessonslearnedfromtroubleshootingdampenedtransmittedpawaveformsofcardiomemsdevice AT scarfonesanto wedgedsensorindistresslessonslearnedfromtroubleshootingdampenedtransmittedpawaveformsofcardiomemsdevice AT zughaibmarcel wedgedsensorindistresslessonslearnedfromtroubleshootingdampenedtransmittedpawaveformsofcardiomemsdevice |