Introduction. Cardiovascular disease is a leading cause of morbidity and mortality with heart failure constituting a large portion of this spectrum. Heart failure patients have 90-day readmission rates of nearly 41% associated with a high expense. Numerous strategies to reduce readmissions have been...

Descripción completa

Detalles Bibliográficos
Autores principales: Singh, Robby, Scarfone, Santo, Zughaib, Marcel
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7036126/
https://www.ncbi.nlm.nih.gov/pubmed/32095291
http://dx.doi.org/10.1155/2020/3856940
_version_ 1783500161320747008
author Singh, Robby
Scarfone, Santo
Zughaib, Marcel
author_facet Singh, Robby
Scarfone, Santo
Zughaib, Marcel
author_sort Singh, Robby
collection PubMed
description Introduction. Cardiovascular disease is a leading cause of morbidity and mortality with heart failure constituting a large portion of this spectrum. Heart failure patients have 90-day readmission rates of nearly 41% associated with a high expense. Numerous strategies to reduce readmissions have been attempted with the CardioMEMS pulmonary artery pressure monitoring system as one of the more successful ones. As this device becomes used more frequently, it is important to recognize procedural complications. We present of a rare complication where a patient underwent successful device placement and was subsequently found to have dampened waveforms which were due to device migration. Case Report. A 79-year-old male underwent successful CardioMEMS placement due to recurrent heart failure hospitalizations. 6 months later, the transmitted waveforms appeared dampened, and repeat angiography revealed a device that had migrated. Rather than abandoning the device, it was recalibrated and continued to transmit data and helped manage the patient's heart failure. Conclusion. CardioMEMS is a cost-effective tool to help reduce heart failure hospitalizations. Device migration is a rare complication and can lead to inaccurate data. However, as seen in this case, the device can be successfully recalibrated and can continue to be utilized to help reduce heart failure admissions.
format Online
Article
Text
id pubmed-7036126
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Hindawi
record_format MEDLINE/PubMed
spelling pubmed-70361262020-02-24 Wedged Sensor in Distress? Lessons Learned from Troubleshooting Dampened Transmitted PA Waveforms of CardioMEMS Device Singh, Robby Scarfone, Santo Zughaib, Marcel Case Rep Cardiol Case Report Introduction. Cardiovascular disease is a leading cause of morbidity and mortality with heart failure constituting a large portion of this spectrum. Heart failure patients have 90-day readmission rates of nearly 41% associated with a high expense. Numerous strategies to reduce readmissions have been attempted with the CardioMEMS pulmonary artery pressure monitoring system as one of the more successful ones. As this device becomes used more frequently, it is important to recognize procedural complications. We present of a rare complication where a patient underwent successful device placement and was subsequently found to have dampened waveforms which were due to device migration. Case Report. A 79-year-old male underwent successful CardioMEMS placement due to recurrent heart failure hospitalizations. 6 months later, the transmitted waveforms appeared dampened, and repeat angiography revealed a device that had migrated. Rather than abandoning the device, it was recalibrated and continued to transmit data and helped manage the patient's heart failure. Conclusion. CardioMEMS is a cost-effective tool to help reduce heart failure hospitalizations. Device migration is a rare complication and can lead to inaccurate data. However, as seen in this case, the device can be successfully recalibrated and can continue to be utilized to help reduce heart failure admissions. Hindawi 2020-02-11 /pmc/articles/PMC7036126/ /pubmed/32095291 http://dx.doi.org/10.1155/2020/3856940 Text en Copyright © 2020 Robby Singh et al. http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Case Report
Singh, Robby
Scarfone, Santo
Zughaib, Marcel
Wedged Sensor in Distress? Lessons Learned from Troubleshooting Dampened Transmitted PA Waveforms of CardioMEMS Device
title Wedged Sensor in Distress? Lessons Learned from Troubleshooting Dampened Transmitted PA Waveforms of CardioMEMS Device
title_full Wedged Sensor in Distress? Lessons Learned from Troubleshooting Dampened Transmitted PA Waveforms of CardioMEMS Device
title_fullStr Wedged Sensor in Distress? Lessons Learned from Troubleshooting Dampened Transmitted PA Waveforms of CardioMEMS Device
title_full_unstemmed Wedged Sensor in Distress? Lessons Learned from Troubleshooting Dampened Transmitted PA Waveforms of CardioMEMS Device
title_short Wedged Sensor in Distress? Lessons Learned from Troubleshooting Dampened Transmitted PA Waveforms of CardioMEMS Device
title_sort wedged sensor in distress? lessons learned from troubleshooting dampened transmitted pa waveforms of cardiomems device
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7036126/
https://www.ncbi.nlm.nih.gov/pubmed/32095291
http://dx.doi.org/10.1155/2020/3856940
work_keys_str_mv AT singhrobby wedgedsensorindistresslessonslearnedfromtroubleshootingdampenedtransmittedpawaveformsofcardiomemsdevice
AT scarfonesanto wedgedsensorindistresslessonslearnedfromtroubleshootingdampenedtransmittedpawaveformsofcardiomemsdevice
AT zughaibmarcel wedgedsensorindistresslessonslearnedfromtroubleshootingdampenedtransmittedpawaveformsofcardiomemsdevice