Cargando…

Microstructure and Properties of Nanostructured Coating on Ti6Al4V

Implant surface properties of Ti6Al4V alloy that is currently used as a biocompatible material because of a variety of unique properties can be improved by a self-organized TiO(2) layer. The TiO(2) nanotubes forming on the titanium-based materials is a relatively recent technology for the surface pr...

Descripción completa

Detalles Bibliográficos
Autores principales: Jordanovová, Veronika, Losertová, Monika, Štencek, Michal, Lukášová, Tereza, Simha Martynková, Gražyna, Peikertová, Pavlína
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7040900/
https://www.ncbi.nlm.nih.gov/pubmed/32033268
http://dx.doi.org/10.3390/ma13030708
Descripción
Sumario:Implant surface properties of Ti6Al4V alloy that is currently used as a biocompatible material because of a variety of unique properties can be improved by a self-organized TiO(2) layer. The TiO(2) nanotubes forming on the titanium-based materials is a relatively recent technology for the surface properties modification and represents pronounced potential in promoting cell adhesion, proliferation, and differentiation that facilitate an implant osseointegration. This work focuses on the influence of surface treatment quality and anodic oxidation parameters on the structure features and properties of TiO(2) nanotube coatings. The nanotubes were formed on Ti6Al4V alloy substrates by simultaneous surface oxidation and controlled dissolving of an oxide film in the presence of fluorine ions. The anodization process on ground or polished samples was performed at experimental condition of 30 V for 1 h. The selected anodized samples were heat treated for 2 h at 500 °C under flowing argon. All samples were characterized by scanning electron microscopy, X-ray diffraction analysis, and Raman spectroscopy. The corrosion rate in physiological solution reached 0.0043, 0.0182, and 0.0998 mm per year for the samples in polished and not-anodized, as-anodized, and anodized-heat treated conditions, respectively.