BACKGROUND AND AIMS: The problem of difficult and failed intubation led to increased development of equipment for airway management. A number of supraglottic airways have now been developed to facilitate the passage of tracheal tubes. Conventional PVC tracheal tubes are recommended for intubation th...

Descripción completa

Detalles Bibliográficos
Autores principales: Lal, Jatin, Bansal, Teena, Dhawan, Gaurav, Taxak, Susheela, Smriti, Manu, Sharma, Jyoti, Thaper, Deepali
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer - Medknow 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7047682/
https://www.ncbi.nlm.nih.gov/pubmed/32174656
http://dx.doi.org/10.4103/joacp.JOACP_227_18
_version_ 1783502163472809984
author Lal, Jatin
Bansal, Teena
Dhawan, Gaurav
Taxak, Susheela
Smriti, Manu
Sharma, Jyoti
Thaper, Deepali
author_facet Lal, Jatin
Bansal, Teena
Dhawan, Gaurav
Taxak, Susheela
Smriti, Manu
Sharma, Jyoti
Thaper, Deepali
author_sort Lal, Jatin
collection PubMed
description BACKGROUND AND AIMS: The problem of difficult and failed intubation led to increased development of equipment for airway management. A number of supraglottic airways have now been developed to facilitate the passage of tracheal tubes. Conventional PVC tracheal tubes are recommended for intubation through the air-Q ILA. No study has compared different PVC tubes for blind intubation through air-Q ILA. Thus, we undertook this prospective, randomised, single blind study to compare two PVC tracheal tubes with different designs viz. conventional PVC tracheal tube (TT) and Parker flex-tip TT with regards to success rate, ease of intubation and total time required for successful intubation through air-Q ILA. MATERIAL AND METHODS: One hundred patients of either sex, aged 18–60 years, belonging to American Society of Anesthesiologists (ASA) physical status class I and II scheduled for elective surgery under general anesthesia requiring endotracheal intubation were included in the study. Blind intubation using conventional PVC TT and Parker flex-tip tube was done in group A (n = 50) and group B (n = 50), respectively. RESULTS: The first attempt success rate in Parker flex-tip TT was significantly more as compared to conventional PVC TT (P = 0.002). Success rate of intubation was significantly more in Parker flex-tip TT as compared to conventional PVC TT (P = 0.004). The intubation was significantly easy in Parker flex-tip tube as compared to conventional PVC TT (P = 0.002). Total time of intubation was less in Parker flex-tip tube as compared to PVC TT (P = 0.043). CONCLUSION: Unique design of the Parker Flex-tip TT resulted in increase in success rate, first attempt success rate and ease of intubation in group B in present study.
format Online
Article
Text
id pubmed-7047682
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Wolters Kluwer - Medknow
record_format MEDLINE/PubMed
spelling pubmed-70476822020-03-13 Comparison of conventional with Parker flex-tip tracheal tube for intubation through air-Q intubating laryngeal airway Lal, Jatin Bansal, Teena Dhawan, Gaurav Taxak, Susheela Smriti, Manu Sharma, Jyoti Thaper, Deepali J Anaesthesiol Clin Pharmacol Original Article BACKGROUND AND AIMS: The problem of difficult and failed intubation led to increased development of equipment for airway management. A number of supraglottic airways have now been developed to facilitate the passage of tracheal tubes. Conventional PVC tracheal tubes are recommended for intubation through the air-Q ILA. No study has compared different PVC tubes for blind intubation through air-Q ILA. Thus, we undertook this prospective, randomised, single blind study to compare two PVC tracheal tubes with different designs viz. conventional PVC tracheal tube (TT) and Parker flex-tip TT with regards to success rate, ease of intubation and total time required for successful intubation through air-Q ILA. MATERIAL AND METHODS: One hundred patients of either sex, aged 18–60 years, belonging to American Society of Anesthesiologists (ASA) physical status class I and II scheduled for elective surgery under general anesthesia requiring endotracheal intubation were included in the study. Blind intubation using conventional PVC TT and Parker flex-tip tube was done in group A (n = 50) and group B (n = 50), respectively. RESULTS: The first attempt success rate in Parker flex-tip TT was significantly more as compared to conventional PVC TT (P = 0.002). Success rate of intubation was significantly more in Parker flex-tip TT as compared to conventional PVC TT (P = 0.004). The intubation was significantly easy in Parker flex-tip tube as compared to conventional PVC TT (P = 0.002). Total time of intubation was less in Parker flex-tip tube as compared to PVC TT (P = 0.043). CONCLUSION: Unique design of the Parker Flex-tip TT resulted in increase in success rate, first attempt success rate and ease of intubation in group B in present study. Wolters Kluwer - Medknow 2020 2020-02-18 /pmc/articles/PMC7047682/ /pubmed/32174656 http://dx.doi.org/10.4103/joacp.JOACP_227_18 Text en Copyright: © 2020 Journal of Anaesthesiology Clinical Pharmacology http://creativecommons.org/licenses/by-nc-sa/4.0 This is an open access journal, and articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 License, which allows others to remix, tweak, and build upon the work non-commercially, as long as appropriate credit is given and the new creations are licensed under the identical terms.
spellingShingle Original Article
Lal, Jatin
Bansal, Teena
Dhawan, Gaurav
Taxak, Susheela
Smriti, Manu
Sharma, Jyoti
Thaper, Deepali
Comparison of conventional with Parker flex-tip tracheal tube for intubation through air-Q intubating laryngeal airway
title Comparison of conventional with Parker flex-tip tracheal tube for intubation through air-Q intubating laryngeal airway
title_full Comparison of conventional with Parker flex-tip tracheal tube for intubation through air-Q intubating laryngeal airway
title_fullStr Comparison of conventional with Parker flex-tip tracheal tube for intubation through air-Q intubating laryngeal airway
title_full_unstemmed Comparison of conventional with Parker flex-tip tracheal tube for intubation through air-Q intubating laryngeal airway
title_short Comparison of conventional with Parker flex-tip tracheal tube for intubation through air-Q intubating laryngeal airway
title_sort comparison of conventional with parker flex-tip tracheal tube for intubation through air-q intubating laryngeal airway
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7047682/
https://www.ncbi.nlm.nih.gov/pubmed/32174656
http://dx.doi.org/10.4103/joacp.JOACP_227_18
work_keys_str_mv AT laljatin comparisonofconventionalwithparkerflextiptrachealtubeforintubationthroughairqintubatinglaryngealairway
AT bansalteena comparisonofconventionalwithparkerflextiptrachealtubeforintubationthroughairqintubatinglaryngealairway
AT dhawangaurav comparisonofconventionalwithparkerflextiptrachealtubeforintubationthroughairqintubatinglaryngealairway
AT taxaksusheela comparisonofconventionalwithparkerflextiptrachealtubeforintubationthroughairqintubatinglaryngealairway
AT smritimanu comparisonofconventionalwithparkerflextiptrachealtubeforintubationthroughairqintubatinglaryngealairway
AT sharmajyoti comparisonofconventionalwithparkerflextiptrachealtubeforintubationthroughairqintubatinglaryngealairway
AT thaperdeepali comparisonofconventionalwithparkerflextiptrachealtubeforintubationthroughairqintubatinglaryngealairway