Effect of patient size on image quality in radiotherapy kV planar verification imaging: a phantom study
INTRODUCTION: This study aimed to determine a maximal pelvic separation and waist circumference in pelvic patients to guide radiation therapists in acquiring kilovoltage (kV) planar images of acceptable quality for treatment verification. METHODS: A pelvic anthropomorphic phantom modified with diffe...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
John Wiley and Sons Inc.
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7063253/ https://www.ncbi.nlm.nih.gov/pubmed/31621206 http://dx.doi.org/10.1002/jmrs.355 |
_version_ | 1783504680479883264 |
---|---|
author | Chan, Sara Giles, Eileen Newmarch, Lyndal Short, Michala |
author_facet | Chan, Sara Giles, Eileen Newmarch, Lyndal Short, Michala |
author_sort | Chan, Sara |
collection | PubMed |
description | INTRODUCTION: This study aimed to determine a maximal pelvic separation and waist circumference in pelvic patients to guide radiation therapists in acquiring kilovoltage (kV) planar images of acceptable quality for treatment verification. METHODS: A pelvic anthropomorphic phantom modified with different bolus thicknesses was imaged at various default kV exposure settings. Radiation therapists rated image quality and acceptance/rejection of these images for treatment verification. RESULTS: Sixteen radiation therapists participated in the study. Image quality was inversely proportional to phantom size. AP and lateral kV images were acceptable for treatment verification up to a waist circumference of 143 cm. CONCLUSIONS: Exposure settings for kV image verification of large patients should be individualised to avoid unnecessary patient radiation dose through repeated imaging. |
format | Online Article Text |
id | pubmed-7063253 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | John Wiley and Sons Inc. |
record_format | MEDLINE/PubMed |
spelling | pubmed-70632532020-03-16 Effect of patient size on image quality in radiotherapy kV planar verification imaging: a phantom study Chan, Sara Giles, Eileen Newmarch, Lyndal Short, Michala J Med Radiat Sci Original Articles INTRODUCTION: This study aimed to determine a maximal pelvic separation and waist circumference in pelvic patients to guide radiation therapists in acquiring kilovoltage (kV) planar images of acceptable quality for treatment verification. METHODS: A pelvic anthropomorphic phantom modified with different bolus thicknesses was imaged at various default kV exposure settings. Radiation therapists rated image quality and acceptance/rejection of these images for treatment verification. RESULTS: Sixteen radiation therapists participated in the study. Image quality was inversely proportional to phantom size. AP and lateral kV images were acceptable for treatment verification up to a waist circumference of 143 cm. CONCLUSIONS: Exposure settings for kV image verification of large patients should be individualised to avoid unnecessary patient radiation dose through repeated imaging. John Wiley and Sons Inc. 2019-10-16 2020-03 /pmc/articles/PMC7063253/ /pubmed/31621206 http://dx.doi.org/10.1002/jmrs.355 Text en © 2019 The Authors. Journal of Medical Radiation Sciences published by John Wiley & Sons Australia, Ltd on behalf of Australian Society of Medical Imaging and Radiation Therapy and New Zealand Institute of Medical Radiation Technology This is an open access article under the terms of the http://creativecommons.org/licenses/by-nc-nd/4.0/ License, which permits use and distribution in any medium, provided the original work is properly cited, the use is non‐commercial and no modifications or adaptations are made. |
spellingShingle | Original Articles Chan, Sara Giles, Eileen Newmarch, Lyndal Short, Michala Effect of patient size on image quality in radiotherapy kV planar verification imaging: a phantom study |
title | Effect of patient size on image quality in radiotherapy kV planar verification imaging: a phantom study |
title_full | Effect of patient size on image quality in radiotherapy kV planar verification imaging: a phantom study |
title_fullStr | Effect of patient size on image quality in radiotherapy kV planar verification imaging: a phantom study |
title_full_unstemmed | Effect of patient size on image quality in radiotherapy kV planar verification imaging: a phantom study |
title_short | Effect of patient size on image quality in radiotherapy kV planar verification imaging: a phantom study |
title_sort | effect of patient size on image quality in radiotherapy kv planar verification imaging: a phantom study |
topic | Original Articles |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7063253/ https://www.ncbi.nlm.nih.gov/pubmed/31621206 http://dx.doi.org/10.1002/jmrs.355 |
work_keys_str_mv | AT chansara effectofpatientsizeonimagequalityinradiotherapykvplanarverificationimagingaphantomstudy AT gileseileen effectofpatientsizeonimagequalityinradiotherapykvplanarverificationimagingaphantomstudy AT newmarchlyndal effectofpatientsizeonimagequalityinradiotherapykvplanarverificationimagingaphantomstudy AT shortmichala effectofpatientsizeonimagequalityinradiotherapykvplanarverificationimagingaphantomstudy |