Selection and application of coils in temporomandibular joint MRI

OBJECTIVE: To compare and evaluate the signal-to-noise ratio (SNR) and the contrast-to-noise ratio (CNR) values between a 15-channel phased array head coil and 6-channel dS Flex M surface coil in the MRI of temporomandibular joint. METHODS: 300 patients were randomly assigned to two groups: 150 pati...

Descripción completa

Detalles Bibliográficos
Autores principales: Sun, Qi, Dong, Min-jun, Tao, Xiao-feng, Jiang, Meng-da, Yang, Chi
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The British Institute of Radiology. 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7068082/
https://www.ncbi.nlm.nih.gov/pubmed/31559845
http://dx.doi.org/10.1259/dmfr.20190002
Descripción
Sumario:OBJECTIVE: To compare and evaluate the signal-to-noise ratio (SNR) and the contrast-to-noise ratio (CNR) values between a 15-channel phased array head coil and 6-channel dS Flex M surface coil in the MRI of temporomandibular joint. METHODS: 300 patients were randomly assigned to two groups: 150 patients were examined by using a 15-channel phased array head coil and the other 150 patients were scanned by using a 6-channel dS Flex M surface coil. All of the data were set in the same 6 regions of interest including the temporal lobe, condyle neck, lateral pterygoid muscle, parotid gland, the adipose area and an area of the background noise). SNR and CNR values were measured respectively. RESULTS: The numerical variation law of SNR and CNR values measured in regionsof interest of each group was similar, although different coils were used. There were statistically significant differences of SNR values in all of the oblique sagittal (OSag) proton density-weighted imaging, the part of OSag T (2) weighted image (T (2)WI) except for SNR(4) and SNR(5). and oblique coronal (OCor) T (2)WI sequence except for SNR(2). On the contrary, SNR(4) and SNR(5) values in the OCor T (2)WI and SNR(5) values in OSag T (2)WI sequences by using the surface coil were higher than those by using the head coil. There were no statistically significant intergroup differences of CNR values in OSag proton density-weighted imaging sequence except CNR(1) and in OSag T (2)WI sequence except CNR(5). But, statistically significant differences of all the values in the OCor T (2)WI sequence except for CNR(1) were observed. CONCLUSION: Both the phased array head coil and dS Flex M surface coil can be used for temporomandibular joint MRI.