Dosimetric characterization of a body‐conforming radiochromic sheet

PURPOSE: A novel radiochromic PRESAGE sheet (Heuris Inc.) with 3 mm thickness has been developed as a measurement tool for 2D dosimetry. Its inherent ability to conform to irregular surfaces makes this dosimeter advantageous for patient surface dosimetry. This study is a comprehensive investigation...

Descripción completa

Detalles Bibliográficos
Autores principales: Wang, Yi‐Fang, Dona, Olga, Liu, Kevin, Adamovics, John, Wuu, Cheng‐Shie
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7075375/
https://www.ncbi.nlm.nih.gov/pubmed/32100948
http://dx.doi.org/10.1002/acm2.12838
_version_ 1783507029186314240
author Wang, Yi‐Fang
Dona, Olga
Liu, Kevin
Adamovics, John
Wuu, Cheng‐Shie
author_facet Wang, Yi‐Fang
Dona, Olga
Liu, Kevin
Adamovics, John
Wuu, Cheng‐Shie
author_sort Wang, Yi‐Fang
collection PubMed
description PURPOSE: A novel radiochromic PRESAGE sheet (Heuris Inc.) with 3 mm thickness has been developed as a measurement tool for 2D dosimetry. Its inherent ability to conform to irregular surfaces makes this dosimeter advantageous for patient surface dosimetry. This study is a comprehensive investigation into the PRESAGE sheet’s dosimetric characteristic, accuracy and its potential use as a dosimeter for clinical applications. METHODS: The characterization of the dosimeter included evaluation of the temporal stability of the dose linearity, reproducibility, measurement uncertainties, dose rate, energy, temperature and angular dependence, lateral response artifacts, percent depth dose curve, and 2D dose measurement. Dose distribution measurements were acquired for regular square fields on a flat and irregular surface and an irregular modulated field on the smooth surface. All measurements were performed using an Epson 11000XL high‐resolution scanner. RESULTS: The examined dosimeters exhibit stable linear response, standard error of repeated measurements within 2%, negligible dose rate, energy, and angular dependence. The same linear dose response was measured while the dosimeter was in contact with a heated water surface. Gamma test and histogram analysis of the dose difference between PRESAGE and EBT3 film, PRESAGE and the treatment planning system (TPS) were used to evaluate the measured dose distributions. The PRESAGE sheet dose distributions showed good agreement with EBT3 film and TPS. A discrepancy smaller than the statistical error of the two dosimeters was reported. CONCLUSIONS: This study established a full dosimetric characterization of the PRESAGE sheets with the purpose of laying the foundation for future clinical uses. The results presented here for the comparison of this novel dosimeter with those currently in use reinforce the possibility of using this dosimeter as an alternative for irregular surface dose measurements.
format Online
Article
Text
id pubmed-7075375
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-70753752020-03-17 Dosimetric characterization of a body‐conforming radiochromic sheet Wang, Yi‐Fang Dona, Olga Liu, Kevin Adamovics, John Wuu, Cheng‐Shie J Appl Clin Med Phys Radiation Measurements PURPOSE: A novel radiochromic PRESAGE sheet (Heuris Inc.) with 3 mm thickness has been developed as a measurement tool for 2D dosimetry. Its inherent ability to conform to irregular surfaces makes this dosimeter advantageous for patient surface dosimetry. This study is a comprehensive investigation into the PRESAGE sheet’s dosimetric characteristic, accuracy and its potential use as a dosimeter for clinical applications. METHODS: The characterization of the dosimeter included evaluation of the temporal stability of the dose linearity, reproducibility, measurement uncertainties, dose rate, energy, temperature and angular dependence, lateral response artifacts, percent depth dose curve, and 2D dose measurement. Dose distribution measurements were acquired for regular square fields on a flat and irregular surface and an irregular modulated field on the smooth surface. All measurements were performed using an Epson 11000XL high‐resolution scanner. RESULTS: The examined dosimeters exhibit stable linear response, standard error of repeated measurements within 2%, negligible dose rate, energy, and angular dependence. The same linear dose response was measured while the dosimeter was in contact with a heated water surface. Gamma test and histogram analysis of the dose difference between PRESAGE and EBT3 film, PRESAGE and the treatment planning system (TPS) were used to evaluate the measured dose distributions. The PRESAGE sheet dose distributions showed good agreement with EBT3 film and TPS. A discrepancy smaller than the statistical error of the two dosimeters was reported. CONCLUSIONS: This study established a full dosimetric characterization of the PRESAGE sheets with the purpose of laying the foundation for future clinical uses. The results presented here for the comparison of this novel dosimeter with those currently in use reinforce the possibility of using this dosimeter as an alternative for irregular surface dose measurements. John Wiley and Sons Inc. 2020-02-26 /pmc/articles/PMC7075375/ /pubmed/32100948 http://dx.doi.org/10.1002/acm2.12838 Text en © 2020 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Measurements
Wang, Yi‐Fang
Dona, Olga
Liu, Kevin
Adamovics, John
Wuu, Cheng‐Shie
Dosimetric characterization of a body‐conforming radiochromic sheet
title Dosimetric characterization of a body‐conforming radiochromic sheet
title_full Dosimetric characterization of a body‐conforming radiochromic sheet
title_fullStr Dosimetric characterization of a body‐conforming radiochromic sheet
title_full_unstemmed Dosimetric characterization of a body‐conforming radiochromic sheet
title_short Dosimetric characterization of a body‐conforming radiochromic sheet
title_sort dosimetric characterization of a body‐conforming radiochromic sheet
topic Radiation Measurements
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7075375/
https://www.ncbi.nlm.nih.gov/pubmed/32100948
http://dx.doi.org/10.1002/acm2.12838
work_keys_str_mv AT wangyifang dosimetriccharacterizationofabodyconformingradiochromicsheet
AT donaolga dosimetriccharacterizationofabodyconformingradiochromicsheet
AT liukevin dosimetriccharacterizationofabodyconformingradiochromicsheet
AT adamovicsjohn dosimetriccharacterizationofabodyconformingradiochromicsheet
AT wuuchengshie dosimetriccharacterizationofabodyconformingradiochromicsheet