Photon skyshine from medical linear accelerators

A widely used formula for the prediction of photon skyshine has been shown to be very inaccurate by comparison with numerous measurements. Discrepancies of up to an order of magnitude have been observed. In addition to this, the formula does not predict the observed dependence on field size, nor the...

Descripción completa

Detalles Bibliográficos
Autor principal: McDermott, Patrick N.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: John Wiley and Sons Inc. 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7075386/
https://www.ncbi.nlm.nih.gov/pubmed/32115894
http://dx.doi.org/10.1002/acm2.12833
_version_ 1783507031590699008
author McDermott, Patrick N.
author_facet McDermott, Patrick N.
author_sort McDermott, Patrick N.
collection PubMed
description A widely used formula for the prediction of photon skyshine has been shown to be very inaccurate by comparison with numerous measurements. Discrepancies of up to an order of magnitude have been observed. In addition to this, the formula does not predict the observed dependence on field size, nor the fact that skyshine dose rates exhibit a local maximum. A scaling formula is derived here, with a single fitting parameter, which properly accounts for these properties, provides physical insight into the skyshine phenomenon, and is more accurate. The location of the maximum dose rate depends on the ratio of the roof height above isocenter to the distance from the isocenter to the outer surface of the sidewall. For nominal linac room dimensions, the maximum dose occurs at a distance from the outer wall of approximately two times the height of the roof above the isocenter. The skyshine dose rate is proportional to the field area and not Ω(1.3), as predicted by the standard formula, where Ω is the solid angle subtended by the beam. For lightly shielded roofs (concrete thickness less than about 0.5 m), the photon skyshine for 6 MV exceeds that for 18 MV. Evidence is presented that at intermediate distances the skyshine declines as one over the distance and not one over the distance squared. Predictions of skyshine dose rates depend critically on accurate knowledge of the roof transmission factor. If a roof is shielded so as to avoid designation as a “high radiation area,” photon skyshine will be negligible.
format Online
Article
Text
id pubmed-7075386
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher John Wiley and Sons Inc.
record_format MEDLINE/PubMed
spelling pubmed-70753862020-03-17 Photon skyshine from medical linear accelerators McDermott, Patrick N. J Appl Clin Med Phys Radiation Oncology Physics A widely used formula for the prediction of photon skyshine has been shown to be very inaccurate by comparison with numerous measurements. Discrepancies of up to an order of magnitude have been observed. In addition to this, the formula does not predict the observed dependence on field size, nor the fact that skyshine dose rates exhibit a local maximum. A scaling formula is derived here, with a single fitting parameter, which properly accounts for these properties, provides physical insight into the skyshine phenomenon, and is more accurate. The location of the maximum dose rate depends on the ratio of the roof height above isocenter to the distance from the isocenter to the outer surface of the sidewall. For nominal linac room dimensions, the maximum dose occurs at a distance from the outer wall of approximately two times the height of the roof above the isocenter. The skyshine dose rate is proportional to the field area and not Ω(1.3), as predicted by the standard formula, where Ω is the solid angle subtended by the beam. For lightly shielded roofs (concrete thickness less than about 0.5 m), the photon skyshine for 6 MV exceeds that for 18 MV. Evidence is presented that at intermediate distances the skyshine declines as one over the distance and not one over the distance squared. Predictions of skyshine dose rates depend critically on accurate knowledge of the roof transmission factor. If a roof is shielded so as to avoid designation as a “high radiation area,” photon skyshine will be negligible. John Wiley and Sons Inc. 2020-03-01 /pmc/articles/PMC7075386/ /pubmed/32115894 http://dx.doi.org/10.1002/acm2.12833 Text en © 2020 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine. This is an open access article under the terms of the http://creativecommons.org/licenses/by/4.0/ License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Radiation Oncology Physics
McDermott, Patrick N.
Photon skyshine from medical linear accelerators
title Photon skyshine from medical linear accelerators
title_full Photon skyshine from medical linear accelerators
title_fullStr Photon skyshine from medical linear accelerators
title_full_unstemmed Photon skyshine from medical linear accelerators
title_short Photon skyshine from medical linear accelerators
title_sort photon skyshine from medical linear accelerators
topic Radiation Oncology Physics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7075386/
https://www.ncbi.nlm.nih.gov/pubmed/32115894
http://dx.doi.org/10.1002/acm2.12833
work_keys_str_mv AT mcdermottpatrickn photonskyshinefrommedicallinearaccelerators