Magnitude of beam-hardening artifacts produced by gutta-percha and metal posts on cone-beam computed tomography with varying tube current
PURPOSE: This study was performed to evaluate the magnitude of artifacts produced by gutta-percha and metal posts on cone-beam computed tomography (CBCT) scans obtained with different tube currents and with or without metal artifact reduction (MAR). MATERIALS AND METHODS: A tooth was inserted in a d...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Korean Academy of Oral and Maxillofacial Radiology
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7078413/ https://www.ncbi.nlm.nih.gov/pubmed/32206614 http://dx.doi.org/10.5624/isd.2020.50.1.1 |
_version_ | 1783507616158187520 |
---|---|
author | Gaêta-Araujo, Hugo Nascimento, Eduarda Helena Leandro Fontenele, Rocharles Cavalcante Mancini, Arthur Xavier Maseti Freitas, Deborah Queiroz Oliveira-Santos, Christiano |
author_facet | Gaêta-Araujo, Hugo Nascimento, Eduarda Helena Leandro Fontenele, Rocharles Cavalcante Mancini, Arthur Xavier Maseti Freitas, Deborah Queiroz Oliveira-Santos, Christiano |
author_sort | Gaêta-Araujo, Hugo |
collection | PubMed |
description | PURPOSE: This study was performed to evaluate the magnitude of artifacts produced by gutta-percha and metal posts on cone-beam computed tomography (CBCT) scans obtained with different tube currents and with or without metal artifact reduction (MAR). MATERIALS AND METHODS: A tooth was inserted in a dry human mandible socket, and CBCT scans were acquired after root canal instrumentation, root canal filling, and metal post placement with various tube currents with and without MAR activation. The artifact magnitude was assessed by the standard deviation (SD) of gray values and the contrast-to-noise ratio (CNR) at the various distances from the tooth. Data were compared using multi-way analysis of variance. RESULTS: At all distances, a current of 4 mA was associated with a higher SD and a lower CNR than 8 mA or 10 mA (P<0.05). For the metal posts without MAR, the artifact magnitude as assessed by SD was greatest at 1.5 cm or less (P<0.05). When MAR was applied, SD values for distances 1.5 cm or closer to the tooth were reduced (P<0.05). MAR usage did not influence the magnitude of artifacts in the control and gutta-percha groups (P>0.05). CONCLUSION: Increasing the tube current from 4 mA to 8 mA may reduce the magnitude of artifacts from metal posts. The magnitude of artifacts arising from metal posts was significantly higher at distances of 1.5 cm or less than at greater distances. MAR usage improved image quality near the metal post, but had no significant influence farther than 1.5 cm from the tooth. |
format | Online Article Text |
id | pubmed-7078413 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | Korean Academy of Oral and Maxillofacial Radiology |
record_format | MEDLINE/PubMed |
spelling | pubmed-70784132020-03-23 Magnitude of beam-hardening artifacts produced by gutta-percha and metal posts on cone-beam computed tomography with varying tube current Gaêta-Araujo, Hugo Nascimento, Eduarda Helena Leandro Fontenele, Rocharles Cavalcante Mancini, Arthur Xavier Maseti Freitas, Deborah Queiroz Oliveira-Santos, Christiano Imaging Sci Dent Original Article PURPOSE: This study was performed to evaluate the magnitude of artifacts produced by gutta-percha and metal posts on cone-beam computed tomography (CBCT) scans obtained with different tube currents and with or without metal artifact reduction (MAR). MATERIALS AND METHODS: A tooth was inserted in a dry human mandible socket, and CBCT scans were acquired after root canal instrumentation, root canal filling, and metal post placement with various tube currents with and without MAR activation. The artifact magnitude was assessed by the standard deviation (SD) of gray values and the contrast-to-noise ratio (CNR) at the various distances from the tooth. Data were compared using multi-way analysis of variance. RESULTS: At all distances, a current of 4 mA was associated with a higher SD and a lower CNR than 8 mA or 10 mA (P<0.05). For the metal posts without MAR, the artifact magnitude as assessed by SD was greatest at 1.5 cm or less (P<0.05). When MAR was applied, SD values for distances 1.5 cm or closer to the tooth were reduced (P<0.05). MAR usage did not influence the magnitude of artifacts in the control and gutta-percha groups (P>0.05). CONCLUSION: Increasing the tube current from 4 mA to 8 mA may reduce the magnitude of artifacts from metal posts. The magnitude of artifacts arising from metal posts was significantly higher at distances of 1.5 cm or less than at greater distances. MAR usage improved image quality near the metal post, but had no significant influence farther than 1.5 cm from the tooth. Korean Academy of Oral and Maxillofacial Radiology 2020-03 2020-03-17 /pmc/articles/PMC7078413/ /pubmed/32206614 http://dx.doi.org/10.5624/isd.2020.50.1.1 Text en Copyright © 2020 by Korean Academy of Oral and Maxillofacial Radiology http://creativecommons.org/licenses/by-nc/3.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Gaêta-Araujo, Hugo Nascimento, Eduarda Helena Leandro Fontenele, Rocharles Cavalcante Mancini, Arthur Xavier Maseti Freitas, Deborah Queiroz Oliveira-Santos, Christiano Magnitude of beam-hardening artifacts produced by gutta-percha and metal posts on cone-beam computed tomography with varying tube current |
title | Magnitude of beam-hardening artifacts produced by gutta-percha and metal posts on cone-beam computed tomography with varying tube current |
title_full | Magnitude of beam-hardening artifacts produced by gutta-percha and metal posts on cone-beam computed tomography with varying tube current |
title_fullStr | Magnitude of beam-hardening artifacts produced by gutta-percha and metal posts on cone-beam computed tomography with varying tube current |
title_full_unstemmed | Magnitude of beam-hardening artifacts produced by gutta-percha and metal posts on cone-beam computed tomography with varying tube current |
title_short | Magnitude of beam-hardening artifacts produced by gutta-percha and metal posts on cone-beam computed tomography with varying tube current |
title_sort | magnitude of beam-hardening artifacts produced by gutta-percha and metal posts on cone-beam computed tomography with varying tube current |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7078413/ https://www.ncbi.nlm.nih.gov/pubmed/32206614 http://dx.doi.org/10.5624/isd.2020.50.1.1 |
work_keys_str_mv | AT gaetaaraujohugo magnitudeofbeamhardeningartifactsproducedbyguttaperchaandmetalpostsonconebeamcomputedtomographywithvaryingtubecurrent AT nascimentoeduardahelenaleandro magnitudeofbeamhardeningartifactsproducedbyguttaperchaandmetalpostsonconebeamcomputedtomographywithvaryingtubecurrent AT fontenelerocharlescavalcante magnitudeofbeamhardeningartifactsproducedbyguttaperchaandmetalpostsonconebeamcomputedtomographywithvaryingtubecurrent AT manciniarthurxaviermaseti magnitudeofbeamhardeningartifactsproducedbyguttaperchaandmetalpostsonconebeamcomputedtomographywithvaryingtubecurrent AT freitasdeborahqueiroz magnitudeofbeamhardeningartifactsproducedbyguttaperchaandmetalpostsonconebeamcomputedtomographywithvaryingtubecurrent AT oliveirasantoschristiano magnitudeofbeamhardeningartifactsproducedbyguttaperchaandmetalpostsonconebeamcomputedtomographywithvaryingtubecurrent |