Cargando…
Efficacy and safety of kaolin-based hemostatic pad vs. standard mechanical compression following transradial and transulnar access for elective coronary angiography and PCI: RAUL trial substudy
Hemostatic devices used in the transradial approach (TRA) and transulnar approach (TUA) are limited. This study compared the efficacy and safety of hemostasis using the QuikClot Radial hemostatic pad (QC) vs. standard mechanical compression (SC) after coronary angiography (CAG). This prospective sin...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer Japan
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7085472/ https://www.ncbi.nlm.nih.gov/pubmed/31655876 http://dx.doi.org/10.1007/s00380-019-01520-z |
_version_ | 1783508941569785856 |
---|---|
author | Pawel, Lewandowski Dagmara, Gralak-Lachowska Pawel, Maciejewski Bogumil, Ramotowski Andrzej, Budaj Sebastian, Stec |
author_facet | Pawel, Lewandowski Dagmara, Gralak-Lachowska Pawel, Maciejewski Bogumil, Ramotowski Andrzej, Budaj Sebastian, Stec |
author_sort | Pawel, Lewandowski |
collection | PubMed |
description | Hemostatic devices used in the transradial approach (TRA) and transulnar approach (TUA) are limited. This study compared the efficacy and safety of hemostasis using the QuikClot Radial hemostatic pad (QC) vs. standard mechanical compression (SC) after coronary angiography (CAG). This prospective single-center randomized trial included CAG patients. The primary and secondary endpoints were efficacy (successful hemostasis) and safety (total artery occlusion [TAO], pseudoaneurysm, hematoma), respectively. A visual analog scale (VAS) evaluated patient pain during compression. In 2013–2017, 200 patients were randomized 2 × 2 into the: (1) TRA and TUA groups and (2) QC and SC groups. Successful hemostasis was achieved in 92 (92%) patients in the QC group and 100 (100%) patients in the SC group (p < 0.006). The TRA SC subgroup showed significantly better results than the TRA QC subgroup (100% vs. 90.0%; p < 0.03). Similar results were obtained in the TUA QC and TUA SC subgroups (95% vs. 100%; p = 0.5). The secondary endpoint was achieved in the QC and SC groups (8% vs. 9%; p = 0.8). Patients reported significantly less pain during QC application than during SC (VAS: 2.6 ± 2.6 vs. 3.4 ± 2.9; p < 0.03). In patients undergoing CAG with TRA or TUA, QC was associated with lower efficacy, less discomfort, and similar safety compared to SC. |
format | Online Article Text |
id | pubmed-7085472 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2019 |
publisher | Springer Japan |
record_format | MEDLINE/PubMed |
spelling | pubmed-70854722020-03-23 Efficacy and safety of kaolin-based hemostatic pad vs. standard mechanical compression following transradial and transulnar access for elective coronary angiography and PCI: RAUL trial substudy Pawel, Lewandowski Dagmara, Gralak-Lachowska Pawel, Maciejewski Bogumil, Ramotowski Andrzej, Budaj Sebastian, Stec Heart Vessels Original Article Hemostatic devices used in the transradial approach (TRA) and transulnar approach (TUA) are limited. This study compared the efficacy and safety of hemostasis using the QuikClot Radial hemostatic pad (QC) vs. standard mechanical compression (SC) after coronary angiography (CAG). This prospective single-center randomized trial included CAG patients. The primary and secondary endpoints were efficacy (successful hemostasis) and safety (total artery occlusion [TAO], pseudoaneurysm, hematoma), respectively. A visual analog scale (VAS) evaluated patient pain during compression. In 2013–2017, 200 patients were randomized 2 × 2 into the: (1) TRA and TUA groups and (2) QC and SC groups. Successful hemostasis was achieved in 92 (92%) patients in the QC group and 100 (100%) patients in the SC group (p < 0.006). The TRA SC subgroup showed significantly better results than the TRA QC subgroup (100% vs. 90.0%; p < 0.03). Similar results were obtained in the TUA QC and TUA SC subgroups (95% vs. 100%; p = 0.5). The secondary endpoint was achieved in the QC and SC groups (8% vs. 9%; p = 0.8). Patients reported significantly less pain during QC application than during SC (VAS: 2.6 ± 2.6 vs. 3.4 ± 2.9; p < 0.03). In patients undergoing CAG with TRA or TUA, QC was associated with lower efficacy, less discomfort, and similar safety compared to SC. Springer Japan 2019-10-26 2020 /pmc/articles/PMC7085472/ /pubmed/31655876 http://dx.doi.org/10.1007/s00380-019-01520-z Text en © The Author(s) 2019 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. |
spellingShingle | Original Article Pawel, Lewandowski Dagmara, Gralak-Lachowska Pawel, Maciejewski Bogumil, Ramotowski Andrzej, Budaj Sebastian, Stec Efficacy and safety of kaolin-based hemostatic pad vs. standard mechanical compression following transradial and transulnar access for elective coronary angiography and PCI: RAUL trial substudy |
title | Efficacy and safety of kaolin-based hemostatic pad vs. standard mechanical compression following transradial and transulnar access for elective coronary angiography and PCI: RAUL trial substudy |
title_full | Efficacy and safety of kaolin-based hemostatic pad vs. standard mechanical compression following transradial and transulnar access for elective coronary angiography and PCI: RAUL trial substudy |
title_fullStr | Efficacy and safety of kaolin-based hemostatic pad vs. standard mechanical compression following transradial and transulnar access for elective coronary angiography and PCI: RAUL trial substudy |
title_full_unstemmed | Efficacy and safety of kaolin-based hemostatic pad vs. standard mechanical compression following transradial and transulnar access for elective coronary angiography and PCI: RAUL trial substudy |
title_short | Efficacy and safety of kaolin-based hemostatic pad vs. standard mechanical compression following transradial and transulnar access for elective coronary angiography and PCI: RAUL trial substudy |
title_sort | efficacy and safety of kaolin-based hemostatic pad vs. standard mechanical compression following transradial and transulnar access for elective coronary angiography and pci: raul trial substudy |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7085472/ https://www.ncbi.nlm.nih.gov/pubmed/31655876 http://dx.doi.org/10.1007/s00380-019-01520-z |
work_keys_str_mv | AT pawellewandowski efficacyandsafetyofkaolinbasedhemostaticpadvsstandardmechanicalcompressionfollowingtransradialandtransulnaraccessforelectivecoronaryangiographyandpciraultrialsubstudy AT dagmaragralaklachowska efficacyandsafetyofkaolinbasedhemostaticpadvsstandardmechanicalcompressionfollowingtransradialandtransulnaraccessforelectivecoronaryangiographyandpciraultrialsubstudy AT pawelmaciejewski efficacyandsafetyofkaolinbasedhemostaticpadvsstandardmechanicalcompressionfollowingtransradialandtransulnaraccessforelectivecoronaryangiographyandpciraultrialsubstudy AT bogumilramotowski efficacyandsafetyofkaolinbasedhemostaticpadvsstandardmechanicalcompressionfollowingtransradialandtransulnaraccessforelectivecoronaryangiographyandpciraultrialsubstudy AT andrzejbudaj efficacyandsafetyofkaolinbasedhemostaticpadvsstandardmechanicalcompressionfollowingtransradialandtransulnaraccessforelectivecoronaryangiographyandpciraultrialsubstudy AT sebastianstec efficacyandsafetyofkaolinbasedhemostaticpadvsstandardmechanicalcompressionfollowingtransradialandtransulnaraccessforelectivecoronaryangiographyandpciraultrialsubstudy |