Quantitative Screening of Cervical Cancers for Low-Resource Settings: Pilot Study of Smartphone-Based Endoscopic Visual Inspection After Acetic Acid Using Machine Learning Techniques
BACKGROUND: Approximately 90% of global cervical cancer (CC) is mostly found in low- and middle-income countries. In most cases, CC can be detected early through routine screening programs, including a cytology-based test. However, it is logistically difficult to offer this program in low-resource s...
Autores principales: | , , , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
JMIR Publications
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7097827/ https://www.ncbi.nlm.nih.gov/pubmed/32159521 http://dx.doi.org/10.2196/16467 |
_version_ | 1783511060603469824 |
---|---|
author | Bae, Jung Kweon Roh, Hyun-Jin You, Joon S Kim, Kyungbin Ahn, Yujin Askaruly, Sanzhar Park, Kibeom Yang, Hyunmo Jang, Gil-Jin Moon, Kyung Hyun Jung, Woonggyu |
author_facet | Bae, Jung Kweon Roh, Hyun-Jin You, Joon S Kim, Kyungbin Ahn, Yujin Askaruly, Sanzhar Park, Kibeom Yang, Hyunmo Jang, Gil-Jin Moon, Kyung Hyun Jung, Woonggyu |
author_sort | Bae, Jung Kweon |
collection | PubMed |
description | BACKGROUND: Approximately 90% of global cervical cancer (CC) is mostly found in low- and middle-income countries. In most cases, CC can be detected early through routine screening programs, including a cytology-based test. However, it is logistically difficult to offer this program in low-resource settings due to limited resources and infrastructure, and few trained experts. A visual inspection following the application of acetic acid (VIA) has been widely promoted and is routinely recommended as a viable form of CC screening in resource-constrained countries. Digital images of the cervix have been acquired during VIA procedure with better quality assurance and visualization, leading to higher diagnostic accuracy and reduction of the variability of detection rate. However, a colposcope is bulky, expensive, electricity-dependent, and needs routine maintenance, and to confirm the grade of abnormality through its images, a specialist must be present. Recently, smartphone-based imaging systems have made a significant impact on the practice of medicine by offering a cost-effective, rapid, and noninvasive method of evaluation. Furthermore, computer-aided analyses, including image processing–based methods and machine learning techniques, have also shown great potential for a high impact on medicinal evaluations. OBJECTIVE: In this study, we demonstrate a new quantitative CC screening technique and implement a machine learning algorithm for smartphone-based endoscopic VIA. We also evaluated the diagnostic performance and practicability of the approach based on the results compared to the gold standard and from physicians’ interpretation. METHODS: A smartphone-based endoscope system was developed and applied to the VIA screening. A total of 20 patients were recruited for this study to evaluate the system. Overall, five were healthy, and 15 were patients who had shown a low to high grade of cervical intraepithelial neoplasia (CIN) from both colposcopy and cytology tests. Endoscopic VIA images were obtained before a loop electrosurgical excision procedure for patients with abnormal tissues, and their histology tissues were collected. Endoscopic VIA images were assessed by four expert physicians relative to the gold standard of histopathology. Also, VIA features were extracted from multiple steps of image processing techniques to find the differences between abnormal (CIN2+) and normal (≤CIN1). By using the extracted features, the performance of different machine learning classifiers, such as k-nearest neighbors (KNN), support vector machine, and decision tree (DT), were compared to find the best algorithm for VIA. After determining the best performing classifying model, it was used to evaluate the screening performance of VIA. RESULTS: An average accuracy of 78%, with a Cohen kappa of 0.571, was observed for the evaluation of the system by four physicians. Through image processing, 240 sliced images were obtained from the cervicogram at each clock position, and five features of VIA were extracted. Among the three models, KNN showed the best performance for finding VIA within holdout 10-fold cross-validation, with an accuracy of 78.3%, area under the curve of 0.807, a specificity of 80.3%, and a sensitivity of 75.0%, respectively. The trained model performed using an unprovided data set resulted in an accuracy of 80.8%, specificity of 84.1%, and sensitivity of 71.9%. Predictions were visualized with intuitive color labels, indicating the normal/abnormal tissue using a circular clock-type segmentation. Calculating the overlapped abnormal tissues between the gold standard and predicted value, the KNN model overperformed the average assessments of physicians for finding VIA. CONCLUSIONS: We explored the potential of the smartphone-based endoscopic VIA as an evaluation technique and used the cervicogram to evaluate normal/abnormal tissue using machine learning techniques. The results of this study demonstrate its potential as a screening tool in low-resource settings. |
format | Online Article Text |
id | pubmed-7097827 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | JMIR Publications |
record_format | MEDLINE/PubMed |
spelling | pubmed-70978272020-03-31 Quantitative Screening of Cervical Cancers for Low-Resource Settings: Pilot Study of Smartphone-Based Endoscopic Visual Inspection After Acetic Acid Using Machine Learning Techniques Bae, Jung Kweon Roh, Hyun-Jin You, Joon S Kim, Kyungbin Ahn, Yujin Askaruly, Sanzhar Park, Kibeom Yang, Hyunmo Jang, Gil-Jin Moon, Kyung Hyun Jung, Woonggyu JMIR Mhealth Uhealth Original Paper BACKGROUND: Approximately 90% of global cervical cancer (CC) is mostly found in low- and middle-income countries. In most cases, CC can be detected early through routine screening programs, including a cytology-based test. However, it is logistically difficult to offer this program in low-resource settings due to limited resources and infrastructure, and few trained experts. A visual inspection following the application of acetic acid (VIA) has been widely promoted and is routinely recommended as a viable form of CC screening in resource-constrained countries. Digital images of the cervix have been acquired during VIA procedure with better quality assurance and visualization, leading to higher diagnostic accuracy and reduction of the variability of detection rate. However, a colposcope is bulky, expensive, electricity-dependent, and needs routine maintenance, and to confirm the grade of abnormality through its images, a specialist must be present. Recently, smartphone-based imaging systems have made a significant impact on the practice of medicine by offering a cost-effective, rapid, and noninvasive method of evaluation. Furthermore, computer-aided analyses, including image processing–based methods and machine learning techniques, have also shown great potential for a high impact on medicinal evaluations. OBJECTIVE: In this study, we demonstrate a new quantitative CC screening technique and implement a machine learning algorithm for smartphone-based endoscopic VIA. We also evaluated the diagnostic performance and practicability of the approach based on the results compared to the gold standard and from physicians’ interpretation. METHODS: A smartphone-based endoscope system was developed and applied to the VIA screening. A total of 20 patients were recruited for this study to evaluate the system. Overall, five were healthy, and 15 were patients who had shown a low to high grade of cervical intraepithelial neoplasia (CIN) from both colposcopy and cytology tests. Endoscopic VIA images were obtained before a loop electrosurgical excision procedure for patients with abnormal tissues, and their histology tissues were collected. Endoscopic VIA images were assessed by four expert physicians relative to the gold standard of histopathology. Also, VIA features were extracted from multiple steps of image processing techniques to find the differences between abnormal (CIN2+) and normal (≤CIN1). By using the extracted features, the performance of different machine learning classifiers, such as k-nearest neighbors (KNN), support vector machine, and decision tree (DT), were compared to find the best algorithm for VIA. After determining the best performing classifying model, it was used to evaluate the screening performance of VIA. RESULTS: An average accuracy of 78%, with a Cohen kappa of 0.571, was observed for the evaluation of the system by four physicians. Through image processing, 240 sliced images were obtained from the cervicogram at each clock position, and five features of VIA were extracted. Among the three models, KNN showed the best performance for finding VIA within holdout 10-fold cross-validation, with an accuracy of 78.3%, area under the curve of 0.807, a specificity of 80.3%, and a sensitivity of 75.0%, respectively. The trained model performed using an unprovided data set resulted in an accuracy of 80.8%, specificity of 84.1%, and sensitivity of 71.9%. Predictions were visualized with intuitive color labels, indicating the normal/abnormal tissue using a circular clock-type segmentation. Calculating the overlapped abnormal tissues between the gold standard and predicted value, the KNN model overperformed the average assessments of physicians for finding VIA. CONCLUSIONS: We explored the potential of the smartphone-based endoscopic VIA as an evaluation technique and used the cervicogram to evaluate normal/abnormal tissue using machine learning techniques. The results of this study demonstrate its potential as a screening tool in low-resource settings. JMIR Publications 2020-03-11 /pmc/articles/PMC7097827/ /pubmed/32159521 http://dx.doi.org/10.2196/16467 Text en ©Jung Kweon Bae, Hyun-Jin Roh, Joon S You, Kyungbin Kim, Yujin Ahn, Sanzhar Askaruly, Kibeom Park, Hyunmo Yang, Gil-Jin Jang, Kyung Hyun Moon, Woonggyu Jung. Originally published in JMIR mHealth and uHealth (http://mhealth.jmir.org), 11.03.2020. https://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work, first published in JMIR mHealth and uHealth, is properly cited. The complete bibliographic information, a link to the original publication on http://mhealth.jmir.org/, as well as this copyright and license information must be included. |
spellingShingle | Original Paper Bae, Jung Kweon Roh, Hyun-Jin You, Joon S Kim, Kyungbin Ahn, Yujin Askaruly, Sanzhar Park, Kibeom Yang, Hyunmo Jang, Gil-Jin Moon, Kyung Hyun Jung, Woonggyu Quantitative Screening of Cervical Cancers for Low-Resource Settings: Pilot Study of Smartphone-Based Endoscopic Visual Inspection After Acetic Acid Using Machine Learning Techniques |
title | Quantitative Screening of Cervical Cancers for Low-Resource Settings: Pilot Study of Smartphone-Based Endoscopic Visual Inspection After Acetic Acid Using Machine Learning Techniques |
title_full | Quantitative Screening of Cervical Cancers for Low-Resource Settings: Pilot Study of Smartphone-Based Endoscopic Visual Inspection After Acetic Acid Using Machine Learning Techniques |
title_fullStr | Quantitative Screening of Cervical Cancers for Low-Resource Settings: Pilot Study of Smartphone-Based Endoscopic Visual Inspection After Acetic Acid Using Machine Learning Techniques |
title_full_unstemmed | Quantitative Screening of Cervical Cancers for Low-Resource Settings: Pilot Study of Smartphone-Based Endoscopic Visual Inspection After Acetic Acid Using Machine Learning Techniques |
title_short | Quantitative Screening of Cervical Cancers for Low-Resource Settings: Pilot Study of Smartphone-Based Endoscopic Visual Inspection After Acetic Acid Using Machine Learning Techniques |
title_sort | quantitative screening of cervical cancers for low-resource settings: pilot study of smartphone-based endoscopic visual inspection after acetic acid using machine learning techniques |
topic | Original Paper |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7097827/ https://www.ncbi.nlm.nih.gov/pubmed/32159521 http://dx.doi.org/10.2196/16467 |
work_keys_str_mv | AT baejungkweon quantitativescreeningofcervicalcancersforlowresourcesettingspilotstudyofsmartphonebasedendoscopicvisualinspectionafteraceticacidusingmachinelearningtechniques AT rohhyunjin quantitativescreeningofcervicalcancersforlowresourcesettingspilotstudyofsmartphonebasedendoscopicvisualinspectionafteraceticacidusingmachinelearningtechniques AT youjoons quantitativescreeningofcervicalcancersforlowresourcesettingspilotstudyofsmartphonebasedendoscopicvisualinspectionafteraceticacidusingmachinelearningtechniques AT kimkyungbin quantitativescreeningofcervicalcancersforlowresourcesettingspilotstudyofsmartphonebasedendoscopicvisualinspectionafteraceticacidusingmachinelearningtechniques AT ahnyujin quantitativescreeningofcervicalcancersforlowresourcesettingspilotstudyofsmartphonebasedendoscopicvisualinspectionafteraceticacidusingmachinelearningtechniques AT askarulysanzhar quantitativescreeningofcervicalcancersforlowresourcesettingspilotstudyofsmartphonebasedendoscopicvisualinspectionafteraceticacidusingmachinelearningtechniques AT parkkibeom quantitativescreeningofcervicalcancersforlowresourcesettingspilotstudyofsmartphonebasedendoscopicvisualinspectionafteraceticacidusingmachinelearningtechniques AT yanghyunmo quantitativescreeningofcervicalcancersforlowresourcesettingspilotstudyofsmartphonebasedendoscopicvisualinspectionafteraceticacidusingmachinelearningtechniques AT janggiljin quantitativescreeningofcervicalcancersforlowresourcesettingspilotstudyofsmartphonebasedendoscopicvisualinspectionafteraceticacidusingmachinelearningtechniques AT moonkyunghyun quantitativescreeningofcervicalcancersforlowresourcesettingspilotstudyofsmartphonebasedendoscopicvisualinspectionafteraceticacidusingmachinelearningtechniques AT jungwoonggyu quantitativescreeningofcervicalcancersforlowresourcesettingspilotstudyofsmartphonebasedendoscopicvisualinspectionafteraceticacidusingmachinelearningtechniques |