Comparison of Ocular Biometry and Refractive Outcomes Using IOL Master 500, IOL Master 700, and Lenstar LS900

PURPOSE: To evaluate the agreement in ocular biometry outcomes measured by three different devices, the IOL Master 500, IOL Master 700, and Lenstar LS900, and compare the refractive outcomes after cataract surgery obtained using those three devices. METHODS: Medical records of 178 eyes of 89 patient...

Descripción completa

Detalles Bibliográficos
Autores principales: Song, Jae Shin, Yoon, Do Yeh, Hyon, Joon Young, Jeon, Hyun Sun
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Korean Ophthalmological Society 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7105785/
https://www.ncbi.nlm.nih.gov/pubmed/32233146
http://dx.doi.org/10.3341/kjo.2019.0102
_version_ 1783512472584454144
author Song, Jae Shin
Yoon, Do Yeh
Hyon, Joon Young
Jeon, Hyun Sun
author_facet Song, Jae Shin
Yoon, Do Yeh
Hyon, Joon Young
Jeon, Hyun Sun
author_sort Song, Jae Shin
collection PubMed
description PURPOSE: To evaluate the agreement in ocular biometry outcomes measured by three different devices, the IOL Master 500, IOL Master 700, and Lenstar LS900, and compare the refractive outcomes after cataract surgery obtained using those three devices. METHODS: Medical records of 178 eyes of 89 patients who underwent ocular biometry with the three devices were retrospectively reviewed, and 124 eyes met the inclusion criteria. Paired comparisons were performed for axial length (AL), mean keratometry (Km), and anterior chamber depth and quantified their agreement using Bland-Altman plots. Subgroup analyses were done according to the AL and the Km. Refractive outcomes were compared with respect to absolute prediction errors after cataract surgery in 54 eyes. RESULTS: Among 124 eyes, 12, 3, and 5 eyes failed to be measured of AL by IOL Master 500, IOL Master 700 and Lenstar LS900, respectively. The AL measured by Lenstar LS900 was longer than that measured by IOL Master 500 and IOL Master 700 (p < 0.001, p = 0.002, respectively). Subgroup analysis revealed that these results were statistically significant only in long eyes (AL >25.5 mm). Km measured using the IOL Master 500 was steeper than that measured with the IOL Master 700 or Lenstar LS900 (p = 0.001, p < 0.001, respectively). anterior chamber depth measured by IOL Master 500 was shallower than that measured by IOL Master 700 or Lenstar LS900 (p < 0.001, p < 0.001, respectively). Ocular biometry measurements by the three devices showed high agreement with narrow 95% limits of agreement. Absolute prediction errors from the 3 devices showed no statistically significant differences after cataract surgery. CONCLUSIONS: The IOL Master 700 and Lenstar LS900 demonstrated superior acquisition rates of biometric measurements compared with the IOL Master 500. Ocular biometry using the 3 different devices showed high agreement, although statistically significant differences were observed; however, since there was no difference in predicting the refractive outcomes, those differences are clinically negligible.
format Online
Article
Text
id pubmed-7105785
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher The Korean Ophthalmological Society
record_format MEDLINE/PubMed
spelling pubmed-71057852020-04-09 Comparison of Ocular Biometry and Refractive Outcomes Using IOL Master 500, IOL Master 700, and Lenstar LS900 Song, Jae Shin Yoon, Do Yeh Hyon, Joon Young Jeon, Hyun Sun Korean J Ophthalmol Original Article PURPOSE: To evaluate the agreement in ocular biometry outcomes measured by three different devices, the IOL Master 500, IOL Master 700, and Lenstar LS900, and compare the refractive outcomes after cataract surgery obtained using those three devices. METHODS: Medical records of 178 eyes of 89 patients who underwent ocular biometry with the three devices were retrospectively reviewed, and 124 eyes met the inclusion criteria. Paired comparisons were performed for axial length (AL), mean keratometry (Km), and anterior chamber depth and quantified their agreement using Bland-Altman plots. Subgroup analyses were done according to the AL and the Km. Refractive outcomes were compared with respect to absolute prediction errors after cataract surgery in 54 eyes. RESULTS: Among 124 eyes, 12, 3, and 5 eyes failed to be measured of AL by IOL Master 500, IOL Master 700 and Lenstar LS900, respectively. The AL measured by Lenstar LS900 was longer than that measured by IOL Master 500 and IOL Master 700 (p < 0.001, p = 0.002, respectively). Subgroup analysis revealed that these results were statistically significant only in long eyes (AL >25.5 mm). Km measured using the IOL Master 500 was steeper than that measured with the IOL Master 700 or Lenstar LS900 (p = 0.001, p < 0.001, respectively). anterior chamber depth measured by IOL Master 500 was shallower than that measured by IOL Master 700 or Lenstar LS900 (p < 0.001, p < 0.001, respectively). Ocular biometry measurements by the three devices showed high agreement with narrow 95% limits of agreement. Absolute prediction errors from the 3 devices showed no statistically significant differences after cataract surgery. CONCLUSIONS: The IOL Master 700 and Lenstar LS900 demonstrated superior acquisition rates of biometric measurements compared with the IOL Master 500. Ocular biometry using the 3 different devices showed high agreement, although statistically significant differences were observed; however, since there was no difference in predicting the refractive outcomes, those differences are clinically negligible. The Korean Ophthalmological Society 2020-04 2020-03-27 /pmc/articles/PMC7105785/ /pubmed/32233146 http://dx.doi.org/10.3341/kjo.2019.0102 Text en © 2020 The Korean Ophthalmological Society http://creativecommons.org/licenses/by-nc/3.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Song, Jae Shin
Yoon, Do Yeh
Hyon, Joon Young
Jeon, Hyun Sun
Comparison of Ocular Biometry and Refractive Outcomes Using IOL Master 500, IOL Master 700, and Lenstar LS900
title Comparison of Ocular Biometry and Refractive Outcomes Using IOL Master 500, IOL Master 700, and Lenstar LS900
title_full Comparison of Ocular Biometry and Refractive Outcomes Using IOL Master 500, IOL Master 700, and Lenstar LS900
title_fullStr Comparison of Ocular Biometry and Refractive Outcomes Using IOL Master 500, IOL Master 700, and Lenstar LS900
title_full_unstemmed Comparison of Ocular Biometry and Refractive Outcomes Using IOL Master 500, IOL Master 700, and Lenstar LS900
title_short Comparison of Ocular Biometry and Refractive Outcomes Using IOL Master 500, IOL Master 700, and Lenstar LS900
title_sort comparison of ocular biometry and refractive outcomes using iol master 500, iol master 700, and lenstar ls900
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7105785/
https://www.ncbi.nlm.nih.gov/pubmed/32233146
http://dx.doi.org/10.3341/kjo.2019.0102
work_keys_str_mv AT songjaeshin comparisonofocularbiometryandrefractiveoutcomesusingiolmaster500iolmaster700andlenstarls900
AT yoondoyeh comparisonofocularbiometryandrefractiveoutcomesusingiolmaster500iolmaster700andlenstarls900
AT hyonjoonyoung comparisonofocularbiometryandrefractiveoutcomesusingiolmaster500iolmaster700andlenstarls900
AT jeonhyunsun comparisonofocularbiometryandrefractiveoutcomesusingiolmaster500iolmaster700andlenstarls900