Purification of infectious bronchitis coronavirus by Sephacryl S-1000 gel chromatography

A procedure was developed to purify infectious bronchitis virus (IBV) by gel chromatography (GC) with a Sephacryl S-1000 column. Virus samples concentrated by centrifugation were applied to a Sephacryl S-1000 column and eluted by 0.02 M phosphate buffer (pH 7.2) containing 0.15 M NaCl. Virus particl...

Descripción completa

Detalles Bibliográficos
Autores principales: Nagano, H., Yagyu, K., Ohta, S.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Published by Elsevier B.V. 1989
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7117437/
https://www.ncbi.nlm.nih.gov/pubmed/2558440
http://dx.doi.org/10.1016/0378-1135(89)90023-0
Descripción
Sumario:A procedure was developed to purify infectious bronchitis virus (IBV) by gel chromatography (GC) with a Sephacryl S-1000 column. Virus samples concentrated by centrifugation were applied to a Sephacryl S-1000 column and eluted by 0.02 M phosphate buffer (pH 7.2) containing 0.15 M NaCl. Virus particles were recovered mainly in the first peak. Purity of the samples was evaluated by both sodium dodecyl sulfate-polyacrylamide gel electrophoresis and electron microscopy. Using electron microscopy, it was found that there were more spike-rich particles in the virus samples purified by GC than in those purified by sucrose density gradient centrifugation (SDGC). In addition, the hemagglutination unit [log(10) (infectivity titer/hemagglutination titer)] of GC-purified virus samples was ∼ 10 times lower than that of SDGC-purified virus samples. These results indicate that Sephacryl S-1000 gel chromatography is useful for purification of IBV.