Cargando…

Implementation of the cobas Liat influenza point-of-care test into an emergency department during a high-incidence season: a retrospective evaluation following real-world implementation

The cobas Liat influenza A/B and respiratory syncytial virus (RSV) assay (Liat) was used in the adult emergency department of a large London hospital from 21(st) January 2018 to 14(th) April 2018. Influenza was detected in 308 of 1027 (30%) samples tested; influenza A in 157 (15.3%), influenza B in...

Descripción completa

Detalles Bibliográficos
Autores principales: Youngs, J., Iqbal, Y., Glass, S., Riley, P., Pope, C., Planche, T., Carrington, D.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Published by Elsevier Ltd on behalf of The Healthcare Infection Society. 2019
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7124296/
https://www.ncbi.nlm.nih.gov/pubmed/30562558
http://dx.doi.org/10.1016/j.jhin.2018.12.008
Descripción
Sumario:The cobas Liat influenza A/B and respiratory syncytial virus (RSV) assay (Liat) was used in the adult emergency department of a large London hospital from 21(st) January 2018 to 14(th) April 2018. Influenza was detected in 308 of 1027 (30%) samples tested; influenza A in 157 (15.3%), influenza B in 149 (14.5%) and RSV in 28 (2.7%). When compared against Fast Track Diagnostics Respiratory Pathogens 21 multiplex polymerase chain reaction and Cepheid Xpert Xpress Flu/RSV assay, Liat performance for the detection of influenza A or B was: sensitivity 85% [95% confidence interval (CI) 76–92)], specificity 98% (95% CI 97–99), negative predictive value 94% (95% CI 92–96) and positive predictive value 95% (95% CI 91–97).