Cargando…
Implementation of the cobas Liat influenza point-of-care test into an emergency department during a high-incidence season: a retrospective evaluation following real-world implementation
The cobas Liat influenza A/B and respiratory syncytial virus (RSV) assay (Liat) was used in the adult emergency department of a large London hospital from 21(st) January 2018 to 14(th) April 2018. Influenza was detected in 308 of 1027 (30%) samples tested; influenza A in 157 (15.3%), influenza B in...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Published by Elsevier Ltd on behalf of The Healthcare Infection Society.
2019
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7124296/ https://www.ncbi.nlm.nih.gov/pubmed/30562558 http://dx.doi.org/10.1016/j.jhin.2018.12.008 |
Sumario: | The cobas Liat influenza A/B and respiratory syncytial virus (RSV) assay (Liat) was used in the adult emergency department of a large London hospital from 21(st) January 2018 to 14(th) April 2018. Influenza was detected in 308 of 1027 (30%) samples tested; influenza A in 157 (15.3%), influenza B in 149 (14.5%) and RSV in 28 (2.7%). When compared against Fast Track Diagnostics Respiratory Pathogens 21 multiplex polymerase chain reaction and Cepheid Xpert Xpress Flu/RSV assay, Liat performance for the detection of influenza A or B was: sensitivity 85% [95% confidence interval (CI) 76–92)], specificity 98% (95% CI 97–99), negative predictive value 94% (95% CI 92–96) and positive predictive value 95% (95% CI 91–97). |
---|