Cargando…
Assessment of Visual Quality in Eyes with Forme Fruste Keratoconus and Mild and Moderate Keratoconus Based on Optical Quality Analysis System II Parameters
PURPOSE: The study aimed to evaluate the visual quality of forme fruste keratoconus (FFK) and mild and moderate keratoconus by using an optical quality analysis system II (OQAS-II) and to explore the correlation between optical quality parameters and the disease progression. METHODS: Twenty-one norm...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Hindawi
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7125474/ https://www.ncbi.nlm.nih.gov/pubmed/32280532 http://dx.doi.org/10.1155/2020/7505016 |
_version_ | 1783515952727457792 |
---|---|
author | Ren, Zhonghao Xu, Liyan Fan, Qi Yang, Kaili Ren, Shengwei Zhao, Dongqing |
author_facet | Ren, Zhonghao Xu, Liyan Fan, Qi Yang, Kaili Ren, Shengwei Zhao, Dongqing |
author_sort | Ren, Zhonghao |
collection | PubMed |
description | PURPOSE: The study aimed to evaluate the visual quality of forme fruste keratoconus (FFK) and mild and moderate keratoconus by using an optical quality analysis system II (OQAS-II) and to explore the correlation between optical quality parameters and the disease progression. METHODS: Twenty-one normal eyes, twenty-one FFK eyes, twenty-one mild keratoconus eyes, and twenty-one moderate keratoconus eyes were included in this prospective study. The optical quality parameters, such as object scatter index (OSI), modulation transfer function cutoff (MTF cutoff), strehl ratio (SR), and OQAS-II values at contrasts of 100% (OV-100), 20% (OV-20), and 9% (OV-9), were measured by OQAS-II. The repeatability of these parameters was analyzed by intraclass correlation coefficient (ICC), repeatability coefficient (RC), and coefficient of variation (CV(w)). Correlations between optical quality parameters and mean central keratometry readings (K(mean)) were evaluated. The sensitivity and specificity of the parameters were analyzed using the receiver operating characteristic (ROC). RESULTS: All the optical quality parameters among four groups showed good repeatability (all ICC≥0.75). The MTF cutoff, SR, OV-100, OV-20, OV-9 in FFK, mild and moderate keratoconus eyes were significantly lower than those in the normal group (all P < 0.05). The ROC analyses of the MTF cutoff, SR, OV-100, OV-20, and OV-9 showed significant area under the curve (AUC) in discriminating FFK form normal, mild keratoconus from FFK, and moderate keratoconus from mild keratoconus (all P < 0.05). The ROC analyses of the MTF cutoff, SR, OV-100, OV-20, and OV-9 showed significant area under the curve (AUC) in discriminating FFK form normal, mild keratoconus from FFK, and moderate keratoconus from mild keratoconus (all P < 0.05). The ROC analyses of the MTF cutoff, SR, OV-100, OV-20, and OV-9 showed significant area under the curve (AUC) in discriminating FFK form normal, mild keratoconus from FFK, and moderate keratoconus from mild keratoconus (all P < 0.05). The ROC analyses of the MTF cutoff, SR, OV-100, OV-20, and OV-9 showed significant area under the curve (AUC) in discriminating FFK form normal, mild keratoconus from FFK, and moderate keratoconus from mild keratoconus (all P < 0.05). The ROC analyses of the MTF cutoff, SR, OV-100, OV-20, and OV-9 showed significant area under the curve (AUC) in discriminating FFK form normal, mild keratoconus from FFK, and moderate keratoconus from mild keratoconus (all K(mean)) were evaluated. The sensitivity and specificity of the parameters were analyzed using the receiver operating characteristic (ROC). r = −0.710, P < 0.05). The ROC analyses of the MTF cutoff, SR, OV-100, OV-20, and OV-9 showed significant area under the curve (AUC) in discriminating FFK form normal, mild keratoconus from FFK, and moderate keratoconus from mild keratoconus (all CONCLUSION: The repeatability of OQAS-II is good in measuring visual quality of normal as well as FFK, mild, and moderate keratoconus. The visual quality of the FFK, mild, and moderate keratoconus is worse than that in normal eyes. The OQAS-II has the potential value in screening FFK from normal eyes and might be a useful tool for evaluating the progression of keratoconus. |
format | Online Article Text |
id | pubmed-7125474 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | Hindawi |
record_format | MEDLINE/PubMed |
spelling | pubmed-71254742020-04-10 Assessment of Visual Quality in Eyes with Forme Fruste Keratoconus and Mild and Moderate Keratoconus Based on Optical Quality Analysis System II Parameters Ren, Zhonghao Xu, Liyan Fan, Qi Yang, Kaili Ren, Shengwei Zhao, Dongqing J Ophthalmol Research Article PURPOSE: The study aimed to evaluate the visual quality of forme fruste keratoconus (FFK) and mild and moderate keratoconus by using an optical quality analysis system II (OQAS-II) and to explore the correlation between optical quality parameters and the disease progression. METHODS: Twenty-one normal eyes, twenty-one FFK eyes, twenty-one mild keratoconus eyes, and twenty-one moderate keratoconus eyes were included in this prospective study. The optical quality parameters, such as object scatter index (OSI), modulation transfer function cutoff (MTF cutoff), strehl ratio (SR), and OQAS-II values at contrasts of 100% (OV-100), 20% (OV-20), and 9% (OV-9), were measured by OQAS-II. The repeatability of these parameters was analyzed by intraclass correlation coefficient (ICC), repeatability coefficient (RC), and coefficient of variation (CV(w)). Correlations between optical quality parameters and mean central keratometry readings (K(mean)) were evaluated. The sensitivity and specificity of the parameters were analyzed using the receiver operating characteristic (ROC). RESULTS: All the optical quality parameters among four groups showed good repeatability (all ICC≥0.75). The MTF cutoff, SR, OV-100, OV-20, OV-9 in FFK, mild and moderate keratoconus eyes were significantly lower than those in the normal group (all P < 0.05). The ROC analyses of the MTF cutoff, SR, OV-100, OV-20, and OV-9 showed significant area under the curve (AUC) in discriminating FFK form normal, mild keratoconus from FFK, and moderate keratoconus from mild keratoconus (all P < 0.05). The ROC analyses of the MTF cutoff, SR, OV-100, OV-20, and OV-9 showed significant area under the curve (AUC) in discriminating FFK form normal, mild keratoconus from FFK, and moderate keratoconus from mild keratoconus (all P < 0.05). The ROC analyses of the MTF cutoff, SR, OV-100, OV-20, and OV-9 showed significant area under the curve (AUC) in discriminating FFK form normal, mild keratoconus from FFK, and moderate keratoconus from mild keratoconus (all P < 0.05). The ROC analyses of the MTF cutoff, SR, OV-100, OV-20, and OV-9 showed significant area under the curve (AUC) in discriminating FFK form normal, mild keratoconus from FFK, and moderate keratoconus from mild keratoconus (all P < 0.05). The ROC analyses of the MTF cutoff, SR, OV-100, OV-20, and OV-9 showed significant area under the curve (AUC) in discriminating FFK form normal, mild keratoconus from FFK, and moderate keratoconus from mild keratoconus (all K(mean)) were evaluated. The sensitivity and specificity of the parameters were analyzed using the receiver operating characteristic (ROC). r = −0.710, P < 0.05). The ROC analyses of the MTF cutoff, SR, OV-100, OV-20, and OV-9 showed significant area under the curve (AUC) in discriminating FFK form normal, mild keratoconus from FFK, and moderate keratoconus from mild keratoconus (all CONCLUSION: The repeatability of OQAS-II is good in measuring visual quality of normal as well as FFK, mild, and moderate keratoconus. The visual quality of the FFK, mild, and moderate keratoconus is worse than that in normal eyes. The OQAS-II has the potential value in screening FFK from normal eyes and might be a useful tool for evaluating the progression of keratoconus. Hindawi 2020-02-29 /pmc/articles/PMC7125474/ /pubmed/32280532 http://dx.doi.org/10.1155/2020/7505016 Text en Copyright © 2020 Zhonghao Ren et al. http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Research Article Ren, Zhonghao Xu, Liyan Fan, Qi Yang, Kaili Ren, Shengwei Zhao, Dongqing Assessment of Visual Quality in Eyes with Forme Fruste Keratoconus and Mild and Moderate Keratoconus Based on Optical Quality Analysis System II Parameters |
title | Assessment of Visual Quality in Eyes with Forme Fruste Keratoconus and Mild and Moderate Keratoconus Based on Optical Quality Analysis System II Parameters |
title_full | Assessment of Visual Quality in Eyes with Forme Fruste Keratoconus and Mild and Moderate Keratoconus Based on Optical Quality Analysis System II Parameters |
title_fullStr | Assessment of Visual Quality in Eyes with Forme Fruste Keratoconus and Mild and Moderate Keratoconus Based on Optical Quality Analysis System II Parameters |
title_full_unstemmed | Assessment of Visual Quality in Eyes with Forme Fruste Keratoconus and Mild and Moderate Keratoconus Based on Optical Quality Analysis System II Parameters |
title_short | Assessment of Visual Quality in Eyes with Forme Fruste Keratoconus and Mild and Moderate Keratoconus Based on Optical Quality Analysis System II Parameters |
title_sort | assessment of visual quality in eyes with forme fruste keratoconus and mild and moderate keratoconus based on optical quality analysis system ii parameters |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7125474/ https://www.ncbi.nlm.nih.gov/pubmed/32280532 http://dx.doi.org/10.1155/2020/7505016 |
work_keys_str_mv | AT renzhonghao assessmentofvisualqualityineyeswithformefrustekeratoconusandmildandmoderatekeratoconusbasedonopticalqualityanalysissystemiiparameters AT xuliyan assessmentofvisualqualityineyeswithformefrustekeratoconusandmildandmoderatekeratoconusbasedonopticalqualityanalysissystemiiparameters AT fanqi assessmentofvisualqualityineyeswithformefrustekeratoconusandmildandmoderatekeratoconusbasedonopticalqualityanalysissystemiiparameters AT yangkaili assessmentofvisualqualityineyeswithformefrustekeratoconusandmildandmoderatekeratoconusbasedonopticalqualityanalysissystemiiparameters AT renshengwei assessmentofvisualqualityineyeswithformefrustekeratoconusandmildandmoderatekeratoconusbasedonopticalqualityanalysissystemiiparameters AT zhaodongqing assessmentofvisualqualityineyeswithformefrustekeratoconusandmildandmoderatekeratoconusbasedonopticalqualityanalysissystemiiparameters |