Emergency physicians cannot inflate or estimate endotracheal tube cuff pressure using standard techniques

Study objectives: Tracheal necrosis and stenosis may result from an overinflated endotracheal tube cuff. Safe, appropriate pressure in endotracheal tube cuffs is considered to be between 15 and 25 cm H(2)O, pressures below normal capillary perfusion pressure. We seek to determine the ability of emer...

Descripción completa

Detalles Bibliográficos
Autores principales: Hoffman, R.J., Parwani, V., Hsu, B., Hahn, I.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: American College of Emergency Physicians. Published by Mosby, Inc. 2004
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7135376/
http://dx.doi.org/10.1016/j.annemergmed.2004.07.383
_version_ 1783518042529988608
author Hoffman, R.J.
Parwani, V.
Hsu, B.
Hahn, I.
author_facet Hoffman, R.J.
Parwani, V.
Hsu, B.
Hahn, I.
author_sort Hoffman, R.J.
collection PubMed
description Study objectives: Tracheal necrosis and stenosis may result from an overinflated endotracheal tube cuff. Safe, appropriate pressure in endotracheal tube cuffs is considered to be between 15 and 25 cm H(2)O, pressures below normal capillary perfusion pressure. We seek to determine the ability of emergency medicine residents and attending physicians in accredited emergency medicine residency training programs to inflate an endotracheal tube cuff to appropriate pressure using standard syringe technique and assess appropriateness of pressure of previously inflated endotracheal tube cuffs by palpating the pilot balloon. Methods: This institutional review board–approved descriptive survey of resident and attending physicians in accredited emergency medicine residency training programs in New York City used a previously tested, tracheal simulation model with a 7.5-mm endotracheal tube with a high-volume low-pressure cuff (Mallinkrodt, St. Louis, MO). Using their choice of a 5-mL or 10-mL plastic syringe with standard Luer Lock (Beckton-Dickson, Franklin Lakes, NJ), participants inflated the endotracheal tube cuff by standard method of injecting air as they deemed appropriate in conjunction with palpating the pilot balloon to estimate cuff pressure. Subsequently, the endotracheal tube cuff pressure was measured using a highly sensitive and accurate analog manometer (Boehringer Laboratories, Norristown, PA). Later, participants palpated the pilot balloon of 9 endotracheal tubes with cuffs previously inflated to known pressures ranging from 0 to 120 cm H(2)O and reported whether the pressure was low, appropriate, or high. Results: Twenty-five resident physicians and 42 attending physicians from 5 emergency medicine residency training programs were surveyed. Only 0.4% (n=3) of participants inflated the cuff to a safe pressure; all were attending physicians. The average cuff pressure generated by emergency medicine attending physicians was greater than 98 cm H(2)O (attending physicians >93 cm H(2)O, residents >106 cm H(2)O). The true mean could not be determined because 57% (n=38) inflated to pressures greater than the upper limit of manometer sensitivity (>120 cm H(2)O). Using palpation, participants were only 33% sensitive detecting inappropriately inflated endotracheal tube cuffs (attending physicians 22% sensitive, residents 53% sensitive), and they were only 26% sensitive in detecting overinflated endotracheal tube cuffs (attending physicians 22%, residents 33%). Average experience as an attending emergency physician was 9 years; average experience as a resident was 2.1 years. Conclusion: This group of emergency physicians had little ability to inflate an endotracheal tube cuff to safe pressure, little ability to accurately estimate pressure of a previously inflated cuff using standard technique, and minimal ability to detect overinflated endotracheal tube cuffs. Nearly all inflated the cuff to dangerously high pressures. Clinicians should consider using devices that permit safe and accurate inflation and measurement of endotracheal tube cuff pressure rather than relying on standard palpation technique, which is potentially unsafe and highly inaccurate.
format Online
Article
Text
id pubmed-7135376
institution National Center for Biotechnology Information
language English
publishDate 2004
publisher American College of Emergency Physicians. Published by Mosby, Inc.
record_format MEDLINE/PubMed
spelling pubmed-71353762020-04-08 Emergency physicians cannot inflate or estimate endotracheal tube cuff pressure using standard techniques Hoffman, R.J. Parwani, V. Hsu, B. Hahn, I. Ann Emerg Med Article Study objectives: Tracheal necrosis and stenosis may result from an overinflated endotracheal tube cuff. Safe, appropriate pressure in endotracheal tube cuffs is considered to be between 15 and 25 cm H(2)O, pressures below normal capillary perfusion pressure. We seek to determine the ability of emergency medicine residents and attending physicians in accredited emergency medicine residency training programs to inflate an endotracheal tube cuff to appropriate pressure using standard syringe technique and assess appropriateness of pressure of previously inflated endotracheal tube cuffs by palpating the pilot balloon. Methods: This institutional review board–approved descriptive survey of resident and attending physicians in accredited emergency medicine residency training programs in New York City used a previously tested, tracheal simulation model with a 7.5-mm endotracheal tube with a high-volume low-pressure cuff (Mallinkrodt, St. Louis, MO). Using their choice of a 5-mL or 10-mL plastic syringe with standard Luer Lock (Beckton-Dickson, Franklin Lakes, NJ), participants inflated the endotracheal tube cuff by standard method of injecting air as they deemed appropriate in conjunction with palpating the pilot balloon to estimate cuff pressure. Subsequently, the endotracheal tube cuff pressure was measured using a highly sensitive and accurate analog manometer (Boehringer Laboratories, Norristown, PA). Later, participants palpated the pilot balloon of 9 endotracheal tubes with cuffs previously inflated to known pressures ranging from 0 to 120 cm H(2)O and reported whether the pressure was low, appropriate, or high. Results: Twenty-five resident physicians and 42 attending physicians from 5 emergency medicine residency training programs were surveyed. Only 0.4% (n=3) of participants inflated the cuff to a safe pressure; all were attending physicians. The average cuff pressure generated by emergency medicine attending physicians was greater than 98 cm H(2)O (attending physicians >93 cm H(2)O, residents >106 cm H(2)O). The true mean could not be determined because 57% (n=38) inflated to pressures greater than the upper limit of manometer sensitivity (>120 cm H(2)O). Using palpation, participants were only 33% sensitive detecting inappropriately inflated endotracheal tube cuffs (attending physicians 22% sensitive, residents 53% sensitive), and they were only 26% sensitive in detecting overinflated endotracheal tube cuffs (attending physicians 22%, residents 33%). Average experience as an attending emergency physician was 9 years; average experience as a resident was 2.1 years. Conclusion: This group of emergency physicians had little ability to inflate an endotracheal tube cuff to safe pressure, little ability to accurately estimate pressure of a previously inflated cuff using standard technique, and minimal ability to detect overinflated endotracheal tube cuffs. Nearly all inflated the cuff to dangerously high pressures. Clinicians should consider using devices that permit safe and accurate inflation and measurement of endotracheal tube cuff pressure rather than relying on standard palpation technique, which is potentially unsafe and highly inaccurate. American College of Emergency Physicians. Published by Mosby, Inc. 2004-10 2004-09-25 /pmc/articles/PMC7135376/ http://dx.doi.org/10.1016/j.annemergmed.2004.07.383 Text en Copyright © 2004 American College of Emergency Physicians. Published by Mosby, Inc. All rights reserved. Since January 2020 Elsevier has created a COVID-19 resource centre with free information in English and Mandarin on the novel coronavirus COVID-19. The COVID-19 resource centre is hosted on Elsevier Connect, the company's public news and information website. Elsevier hereby grants permission to make all its COVID-19-related research that is available on the COVID-19 resource centre - including this research content - immediately available in PubMed Central and other publicly funded repositories, such as the WHO COVID database with rights for unrestricted research re-use and analyses in any form or by any means with acknowledgement of the original source. These permissions are granted for free by Elsevier for as long as the COVID-19 resource centre remains active.
spellingShingle Article
Hoffman, R.J.
Parwani, V.
Hsu, B.
Hahn, I.
Emergency physicians cannot inflate or estimate endotracheal tube cuff pressure using standard techniques
title Emergency physicians cannot inflate or estimate endotracheal tube cuff pressure using standard techniques
title_full Emergency physicians cannot inflate or estimate endotracheal tube cuff pressure using standard techniques
title_fullStr Emergency physicians cannot inflate or estimate endotracheal tube cuff pressure using standard techniques
title_full_unstemmed Emergency physicians cannot inflate or estimate endotracheal tube cuff pressure using standard techniques
title_short Emergency physicians cannot inflate or estimate endotracheal tube cuff pressure using standard techniques
title_sort emergency physicians cannot inflate or estimate endotracheal tube cuff pressure using standard techniques
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7135376/
http://dx.doi.org/10.1016/j.annemergmed.2004.07.383
work_keys_str_mv AT hoffmanrj emergencyphysicianscannotinflateorestimateendotrachealtubecuffpressureusingstandardtechniques
AT parwaniv emergencyphysicianscannotinflateorestimateendotrachealtubecuffpressureusingstandardtechniques
AT hsub emergencyphysicianscannotinflateorestimateendotrachealtubecuffpressureusingstandardtechniques
AT hahni emergencyphysicianscannotinflateorestimateendotrachealtubecuffpressureusingstandardtechniques