Removal of a Trigger Cord Stuck between Bands during Endoscopic Multiple-Band Ligation for Treating Esophageal Variceal Hemorrhage

Endoscopic variceal ligation is the preferred endoscopic treatment method for esophageal variceal bleeding. The incidence of complications such as chest pain, bleeding, stricture formation, and aspiration pneumonia is low. We report a case wherein a malfunctioning multiple-band ligator could have po...

Descripción completa

Detalles Bibliográficos
Autores principales: Ham, Nam Seok, Lee, Danbi, Won, Sung Hyun, Kim, Jeongseok, Jo, Seokjung, Yi, Sangyoung, So, Seol
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Korean Society of Gastrointestinal Endoscopy 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7137571/
https://www.ncbi.nlm.nih.gov/pubmed/31337195
http://dx.doi.org/10.5946/ce.2019.076
Descripción
Sumario:Endoscopic variceal ligation is the preferred endoscopic treatment method for esophageal variceal bleeding. The incidence of complications such as chest pain, bleeding, stricture formation, and aspiration pneumonia is low. We report a case wherein a malfunctioning multiple-band ligator could have potentially caused damage to the esophageal varices and massive bleeding. The equipment was safely removed using scissors and forceps. To the best of our knowledge, this is the first published report detailing the management of a case of esophageal variceal bleeding.