Limited data is available concerning the safety of active middle ear implants (AMEI) during Magnetic Resonance Imaging (MRI). Measurements in temporal bones are the gold standard for preclinical assessment of device safety. In this study the coupling stability of an actuator as used in a fully impla...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2020
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7144975/ https://www.ncbi.nlm.nih.gov/pubmed/32271819 http://dx.doi.org/10.1371/journal.pone.0231213 |
_version_ | 1783519917596737536 |
---|---|
author | Fierens, Guy Verhaert, Nicolas Benoudiba, Farida Bellin, Marie-France Ducreux, Dennis Papon, Jean-François Nevoux, Jérôme |
author_facet | Fierens, Guy Verhaert, Nicolas Benoudiba, Farida Bellin, Marie-France Ducreux, Dennis Papon, Jean-François Nevoux, Jérôme |
author_sort | Fierens, Guy |
collection | PubMed |
description | Limited data is available concerning the safety of active middle ear implants (AMEI) during Magnetic Resonance Imaging (MRI). Measurements in temporal bones are the gold standard for preclinical assessment of device safety. In this study the coupling stability of an actuator as used in a fully implantable AMEI was determined in temporal bones. Eleven temporal bones were implanted with the actuator according to the manufacturer’s surgical guidelines. The actuator was coupled on the incus short process as recommended for sensorineural hearing loss. Temporal bones were exposed 10 times to the MRI magnetic field by entering the MRI suite in a clinically relevant way. Computed Tomography (CT) images were acquired before and after the experiment to investigate the risk of actuator dislocation. Based on the electrical impedance of the actuator, the loading of the actuator to the incus was confirmed. Relative actuator displacement was determined on the CT images by comparing the initial with the final actuator position in 3D space. Impedance curves were analyzed after each exposure to check the loading of the actuator to the ossicles. Analysis of CT images with a 0.30.6 mm in-plane resolution indicate no actuator displacement. The maximum detected change in impedance for all actuators was 8.43 Ω at the actuator’s resonance frequency. Impedance curves measured when the actuator was retracted from the short process after the experiment still indicate the presence of a clear resonance peak. No actuator displacement or dislocation could be detected in the analysis of CT images and the measured impedance curves. Impedance curves obtained when the actuator was retracted from the incus short process still show a clear resonance peak, indicating the device is still functional after the MRI exposures. |
format | Online Article Text |
id | pubmed-7144975 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2020 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-71449752020-04-10 Stability of the standard incus coupling of the Carina middle ear actuator after 1.5T MRI Fierens, Guy Verhaert, Nicolas Benoudiba, Farida Bellin, Marie-France Ducreux, Dennis Papon, Jean-François Nevoux, Jérôme PLoS One Research Article Limited data is available concerning the safety of active middle ear implants (AMEI) during Magnetic Resonance Imaging (MRI). Measurements in temporal bones are the gold standard for preclinical assessment of device safety. In this study the coupling stability of an actuator as used in a fully implantable AMEI was determined in temporal bones. Eleven temporal bones were implanted with the actuator according to the manufacturer’s surgical guidelines. The actuator was coupled on the incus short process as recommended for sensorineural hearing loss. Temporal bones were exposed 10 times to the MRI magnetic field by entering the MRI suite in a clinically relevant way. Computed Tomography (CT) images were acquired before and after the experiment to investigate the risk of actuator dislocation. Based on the electrical impedance of the actuator, the loading of the actuator to the incus was confirmed. Relative actuator displacement was determined on the CT images by comparing the initial with the final actuator position in 3D space. Impedance curves were analyzed after each exposure to check the loading of the actuator to the ossicles. Analysis of CT images with a 0.30.6 mm in-plane resolution indicate no actuator displacement. The maximum detected change in impedance for all actuators was 8.43 Ω at the actuator’s resonance frequency. Impedance curves measured when the actuator was retracted from the short process after the experiment still indicate the presence of a clear resonance peak. No actuator displacement or dislocation could be detected in the analysis of CT images and the measured impedance curves. Impedance curves obtained when the actuator was retracted from the incus short process still show a clear resonance peak, indicating the device is still functional after the MRI exposures. Public Library of Science 2020-04-09 /pmc/articles/PMC7144975/ /pubmed/32271819 http://dx.doi.org/10.1371/journal.pone.0231213 Text en © 2020 Fierens et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited. |
spellingShingle | Research Article Fierens, Guy Verhaert, Nicolas Benoudiba, Farida Bellin, Marie-France Ducreux, Dennis Papon, Jean-François Nevoux, Jérôme Stability of the standard incus coupling of the Carina middle ear actuator after 1.5T MRI |
title | Stability of the standard incus coupling of the Carina middle ear actuator after 1.5T MRI |
title_full | Stability of the standard incus coupling of the Carina middle ear actuator after 1.5T MRI |
title_fullStr | Stability of the standard incus coupling of the Carina middle ear actuator after 1.5T MRI |
title_full_unstemmed | Stability of the standard incus coupling of the Carina middle ear actuator after 1.5T MRI |
title_short | Stability of the standard incus coupling of the Carina middle ear actuator after 1.5T MRI |
title_sort | stability of the standard incus coupling of the carina middle ear actuator after 1.5t mri |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7144975/ https://www.ncbi.nlm.nih.gov/pubmed/32271819 http://dx.doi.org/10.1371/journal.pone.0231213 |
work_keys_str_mv | AT fierensguy stabilityofthestandardincuscouplingofthecarinamiddleearactuatorafter15tmri AT verhaertnicolas stabilityofthestandardincuscouplingofthecarinamiddleearactuatorafter15tmri AT benoudibafarida stabilityofthestandardincuscouplingofthecarinamiddleearactuatorafter15tmri AT bellinmariefrance stabilityofthestandardincuscouplingofthecarinamiddleearactuatorafter15tmri AT ducreuxdennis stabilityofthestandardincuscouplingofthecarinamiddleearactuatorafter15tmri AT paponjeanfrancois stabilityofthestandardincuscouplingofthecarinamiddleearactuatorafter15tmri AT nevouxjerome stabilityofthestandardincuscouplingofthecarinamiddleearactuatorafter15tmri |