Stability of the standard incus coupling of the Carina middle ear actuator after 1.5T MRI

Limited data is available concerning the safety of active middle ear implants (AMEI) during Magnetic Resonance Imaging (MRI). Measurements in temporal bones are the gold standard for preclinical assessment of device safety. In this study the coupling stability of an actuator as used in a fully impla...

Descripción completa

Detalles Bibliográficos
Autores principales: Fierens, Guy, Verhaert, Nicolas, Benoudiba, Farida, Bellin, Marie-France, Ducreux, Dennis, Papon, Jean-François, Nevoux, Jérôme
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7144975/
https://www.ncbi.nlm.nih.gov/pubmed/32271819
http://dx.doi.org/10.1371/journal.pone.0231213
_version_ 1783519917596737536
author Fierens, Guy
Verhaert, Nicolas
Benoudiba, Farida
Bellin, Marie-France
Ducreux, Dennis
Papon, Jean-François
Nevoux, Jérôme
author_facet Fierens, Guy
Verhaert, Nicolas
Benoudiba, Farida
Bellin, Marie-France
Ducreux, Dennis
Papon, Jean-François
Nevoux, Jérôme
author_sort Fierens, Guy
collection PubMed
description Limited data is available concerning the safety of active middle ear implants (AMEI) during Magnetic Resonance Imaging (MRI). Measurements in temporal bones are the gold standard for preclinical assessment of device safety. In this study the coupling stability of an actuator as used in a fully implantable AMEI was determined in temporal bones. Eleven temporal bones were implanted with the actuator according to the manufacturer’s surgical guidelines. The actuator was coupled on the incus short process as recommended for sensorineural hearing loss. Temporal bones were exposed 10 times to the MRI magnetic field by entering the MRI suite in a clinically relevant way. Computed Tomography (CT) images were acquired before and after the experiment to investigate the risk of actuator dislocation. Based on the electrical impedance of the actuator, the loading of the actuator to the incus was confirmed. Relative actuator displacement was determined on the CT images by comparing the initial with the final actuator position in 3D space. Impedance curves were analyzed after each exposure to check the loading of the actuator to the ossicles. Analysis of CT images with a 0.30.6 mm in-plane resolution indicate no actuator displacement. The maximum detected change in impedance for all actuators was 8.43 Ω at the actuator’s resonance frequency. Impedance curves measured when the actuator was retracted from the short process after the experiment still indicate the presence of a clear resonance peak. No actuator displacement or dislocation could be detected in the analysis of CT images and the measured impedance curves. Impedance curves obtained when the actuator was retracted from the incus short process still show a clear resonance peak, indicating the device is still functional after the MRI exposures.
format Online
Article
Text
id pubmed-7144975
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-71449752020-04-10 Stability of the standard incus coupling of the Carina middle ear actuator after 1.5T MRI Fierens, Guy Verhaert, Nicolas Benoudiba, Farida Bellin, Marie-France Ducreux, Dennis Papon, Jean-François Nevoux, Jérôme PLoS One Research Article Limited data is available concerning the safety of active middle ear implants (AMEI) during Magnetic Resonance Imaging (MRI). Measurements in temporal bones are the gold standard for preclinical assessment of device safety. In this study the coupling stability of an actuator as used in a fully implantable AMEI was determined in temporal bones. Eleven temporal bones were implanted with the actuator according to the manufacturer’s surgical guidelines. The actuator was coupled on the incus short process as recommended for sensorineural hearing loss. Temporal bones were exposed 10 times to the MRI magnetic field by entering the MRI suite in a clinically relevant way. Computed Tomography (CT) images were acquired before and after the experiment to investigate the risk of actuator dislocation. Based on the electrical impedance of the actuator, the loading of the actuator to the incus was confirmed. Relative actuator displacement was determined on the CT images by comparing the initial with the final actuator position in 3D space. Impedance curves were analyzed after each exposure to check the loading of the actuator to the ossicles. Analysis of CT images with a 0.30.6 mm in-plane resolution indicate no actuator displacement. The maximum detected change in impedance for all actuators was 8.43 Ω at the actuator’s resonance frequency. Impedance curves measured when the actuator was retracted from the short process after the experiment still indicate the presence of a clear resonance peak. No actuator displacement or dislocation could be detected in the analysis of CT images and the measured impedance curves. Impedance curves obtained when the actuator was retracted from the incus short process still show a clear resonance peak, indicating the device is still functional after the MRI exposures. Public Library of Science 2020-04-09 /pmc/articles/PMC7144975/ /pubmed/32271819 http://dx.doi.org/10.1371/journal.pone.0231213 Text en © 2020 Fierens et al http://creativecommons.org/licenses/by/4.0/ This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0/) , which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
spellingShingle Research Article
Fierens, Guy
Verhaert, Nicolas
Benoudiba, Farida
Bellin, Marie-France
Ducreux, Dennis
Papon, Jean-François
Nevoux, Jérôme
Stability of the standard incus coupling of the Carina middle ear actuator after 1.5T MRI
title Stability of the standard incus coupling of the Carina middle ear actuator after 1.5T MRI
title_full Stability of the standard incus coupling of the Carina middle ear actuator after 1.5T MRI
title_fullStr Stability of the standard incus coupling of the Carina middle ear actuator after 1.5T MRI
title_full_unstemmed Stability of the standard incus coupling of the Carina middle ear actuator after 1.5T MRI
title_short Stability of the standard incus coupling of the Carina middle ear actuator after 1.5T MRI
title_sort stability of the standard incus coupling of the carina middle ear actuator after 1.5t mri
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7144975/
https://www.ncbi.nlm.nih.gov/pubmed/32271819
http://dx.doi.org/10.1371/journal.pone.0231213
work_keys_str_mv AT fierensguy stabilityofthestandardincuscouplingofthecarinamiddleearactuatorafter15tmri
AT verhaertnicolas stabilityofthestandardincuscouplingofthecarinamiddleearactuatorafter15tmri
AT benoudibafarida stabilityofthestandardincuscouplingofthecarinamiddleearactuatorafter15tmri
AT bellinmariefrance stabilityofthestandardincuscouplingofthecarinamiddleearactuatorafter15tmri
AT ducreuxdennis stabilityofthestandardincuscouplingofthecarinamiddleearactuatorafter15tmri
AT paponjeanfrancois stabilityofthestandardincuscouplingofthecarinamiddleearactuatorafter15tmri
AT nevouxjerome stabilityofthestandardincuscouplingofthecarinamiddleearactuatorafter15tmri