Cargando…

Assessment of the risk of infectious aerosols leaking to the environment from BSL-3 laboratory HEPA air filtration systems using model bacterial aerosols

To assess the risk of infectious bacterial aerosols leaking to the environment, the filtration efficiency of a biosafety level 3 (BSL-3) laboratory high-efficiency particulate (HEPA) filter was investigated using the aerosolized bacteria Serratia marcescens. The aerosol size was measured using an An...

Descripción completa

Detalles Bibliográficos
Autores principales: Wen, Zhanbo, Yang, Wenhui, Li, Na, Wang, Jie, Hu, Lingfei, Li, Jinsong, Yin, Zhe, Zhang, Ke, Dong, Xiaokai
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Chinese Society of Particuology and Institute of Process Engineering, Chinese Academy of Sciences. Published by Elsevier B.V. 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7148691/
http://dx.doi.org/10.1016/j.partic.2012.11.009
_version_ 1783520646265831424
author Wen, Zhanbo
Yang, Wenhui
Li, Na
Wang, Jie
Hu, Lingfei
Li, Jinsong
Yin, Zhe
Zhang, Ke
Dong, Xiaokai
author_facet Wen, Zhanbo
Yang, Wenhui
Li, Na
Wang, Jie
Hu, Lingfei
Li, Jinsong
Yin, Zhe
Zhang, Ke
Dong, Xiaokai
author_sort Wen, Zhanbo
collection PubMed
description To assess the risk of infectious bacterial aerosols leaking to the environment, the filtration efficiency of a biosafety level 3 (BSL-3) laboratory high-efficiency particulate (HEPA) filter was investigated using the aerosolized bacteria Serratia marcescens. The aerosol size was measured using an Andersen sampler. Eight first stage HEPA filters (numbered 1–8) were distributed in contaminated labs and exhausts from each of the first stage HEPA filters were aggregated and filtered through one second stage HEPA filter before being released to the environment. In total, 8 first-stage and 1 second-stage HEPA filters from the BSL-3 air purification system were analyzed. No S. marcescens was detected in first stage filters 1, 2, 4, 5, 7 and 8 and the second stage HEPA filter. The filtration efficiencies against aerosolized S. marcescens were >99.9999%. First stage filter numbers 3 and 6 had filtration efficiencies of 99.9825% and 99.9906%, respectively. When filter number 3 was replaced by a new filter and the bracket for filter number 6 was sealed, no aerosolized S. marcescens was detected in the filtered air. Our work suggests that the BSL-3 laboratory HEPA filter air purification system is effective against bacterial aerosols, with little to no bacterial leakage into the environment.
format Online
Article
Text
id pubmed-7148691
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher Chinese Society of Particuology and Institute of Process Engineering, Chinese Academy of Sciences. Published by Elsevier B.V.
record_format MEDLINE/PubMed
spelling pubmed-71486912020-04-13 Assessment of the risk of infectious aerosols leaking to the environment from BSL-3 laboratory HEPA air filtration systems using model bacterial aerosols Wen, Zhanbo Yang, Wenhui Li, Na Wang, Jie Hu, Lingfei Li, Jinsong Yin, Zhe Zhang, Ke Dong, Xiaokai Particuology Article To assess the risk of infectious bacterial aerosols leaking to the environment, the filtration efficiency of a biosafety level 3 (BSL-3) laboratory high-efficiency particulate (HEPA) filter was investigated using the aerosolized bacteria Serratia marcescens. The aerosol size was measured using an Andersen sampler. Eight first stage HEPA filters (numbered 1–8) were distributed in contaminated labs and exhausts from each of the first stage HEPA filters were aggregated and filtered through one second stage HEPA filter before being released to the environment. In total, 8 first-stage and 1 second-stage HEPA filters from the BSL-3 air purification system were analyzed. No S. marcescens was detected in first stage filters 1, 2, 4, 5, 7 and 8 and the second stage HEPA filter. The filtration efficiencies against aerosolized S. marcescens were >99.9999%. First stage filter numbers 3 and 6 had filtration efficiencies of 99.9825% and 99.9906%, respectively. When filter number 3 was replaced by a new filter and the bracket for filter number 6 was sealed, no aerosolized S. marcescens was detected in the filtered air. Our work suggests that the BSL-3 laboratory HEPA filter air purification system is effective against bacterial aerosols, with little to no bacterial leakage into the environment. Chinese Society of Particuology and Institute of Process Engineering, Chinese Academy of Sciences. Published by Elsevier B.V. 2014-04 2013-06-15 /pmc/articles/PMC7148691/ http://dx.doi.org/10.1016/j.partic.2012.11.009 Text en Copyright © 2013 Chinese Society of Particuology and Institute of Process Engineering, Chinese Academy of Sciences. Published by Elsevier B.V. All rights reserved. Since January 2020 Elsevier has created a COVID-19 resource centre with free information in English and Mandarin on the novel coronavirus COVID-19. The COVID-19 resource centre is hosted on Elsevier Connect, the company's public news and information website. Elsevier hereby grants permission to make all its COVID-19-related research that is available on the COVID-19 resource centre - including this research content - immediately available in PubMed Central and other publicly funded repositories, such as the WHO COVID database with rights for unrestricted research re-use and analyses in any form or by any means with acknowledgement of the original source. These permissions are granted for free by Elsevier for as long as the COVID-19 resource centre remains active.
spellingShingle Article
Wen, Zhanbo
Yang, Wenhui
Li, Na
Wang, Jie
Hu, Lingfei
Li, Jinsong
Yin, Zhe
Zhang, Ke
Dong, Xiaokai
Assessment of the risk of infectious aerosols leaking to the environment from BSL-3 laboratory HEPA air filtration systems using model bacterial aerosols
title Assessment of the risk of infectious aerosols leaking to the environment from BSL-3 laboratory HEPA air filtration systems using model bacterial aerosols
title_full Assessment of the risk of infectious aerosols leaking to the environment from BSL-3 laboratory HEPA air filtration systems using model bacterial aerosols
title_fullStr Assessment of the risk of infectious aerosols leaking to the environment from BSL-3 laboratory HEPA air filtration systems using model bacterial aerosols
title_full_unstemmed Assessment of the risk of infectious aerosols leaking to the environment from BSL-3 laboratory HEPA air filtration systems using model bacterial aerosols
title_short Assessment of the risk of infectious aerosols leaking to the environment from BSL-3 laboratory HEPA air filtration systems using model bacterial aerosols
title_sort assessment of the risk of infectious aerosols leaking to the environment from bsl-3 laboratory hepa air filtration systems using model bacterial aerosols
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7148691/
http://dx.doi.org/10.1016/j.partic.2012.11.009
work_keys_str_mv AT wenzhanbo assessmentoftheriskofinfectiousaerosolsleakingtotheenvironmentfrombsl3laboratoryhepaairfiltrationsystemsusingmodelbacterialaerosols
AT yangwenhui assessmentoftheriskofinfectiousaerosolsleakingtotheenvironmentfrombsl3laboratoryhepaairfiltrationsystemsusingmodelbacterialaerosols
AT lina assessmentoftheriskofinfectiousaerosolsleakingtotheenvironmentfrombsl3laboratoryhepaairfiltrationsystemsusingmodelbacterialaerosols
AT wangjie assessmentoftheriskofinfectiousaerosolsleakingtotheenvironmentfrombsl3laboratoryhepaairfiltrationsystemsusingmodelbacterialaerosols
AT hulingfei assessmentoftheriskofinfectiousaerosolsleakingtotheenvironmentfrombsl3laboratoryhepaairfiltrationsystemsusingmodelbacterialaerosols
AT lijinsong assessmentoftheriskofinfectiousaerosolsleakingtotheenvironmentfrombsl3laboratoryhepaairfiltrationsystemsusingmodelbacterialaerosols
AT yinzhe assessmentoftheriskofinfectiousaerosolsleakingtotheenvironmentfrombsl3laboratoryhepaairfiltrationsystemsusingmodelbacterialaerosols
AT zhangke assessmentoftheriskofinfectiousaerosolsleakingtotheenvironmentfrombsl3laboratoryhepaairfiltrationsystemsusingmodelbacterialaerosols
AT dongxiaokai assessmentoftheriskofinfectiousaerosolsleakingtotheenvironmentfrombsl3laboratoryhepaairfiltrationsystemsusingmodelbacterialaerosols