Scanning Electron Microscopy Study of Retrieved Implants Suggests a Ratcheting Mechanism Behind Medial Migration in Cephalomedullary Nailing of Hip Fractures

INTRODUCTION: Medial migration is the paradoxical migration of the femoral neck element (FNE) superomedially against gravity with respect to the intramedullary component of the cephalomedullary device, increasingly seen in the management of pertrochanteric hip fractures with the intramedullary nail....

Descripción completa

Detalles Bibliográficos
Autores principales: Law, GW, Koh, JSB, Yew, AKS, Howe, TS
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Malaysian Orthopaedic Association 2020
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7156168/
https://www.ncbi.nlm.nih.gov/pubmed/32296476
http://dx.doi.org/10.5704/MOJ.2003.002
_version_ 1783522163078201344
author Law, GW
Koh, JSB
Yew, AKS
Howe, TS
author_facet Law, GW
Koh, JSB
Yew, AKS
Howe, TS
author_sort Law, GW
collection PubMed
description INTRODUCTION: Medial migration is the paradoxical migration of the femoral neck element (FNE) superomedially against gravity with respect to the intramedullary component of the cephalomedullary device, increasingly seen in the management of pertrochanteric hip fractures with the intramedullary nail. We postulate that the peculiar anti-gravity movement of the FNE in the medial migration phenomenon stems from a ratcheting mechanism at the intramedullary nail-FNE interface, which should inadvertently produce unique wear patterns on the FNE that can be seen with high-powered microscopy. By examining the wear patterns on retrieved implants from patients with medial migration, our study aims to draw clinical correlations to the ratcheting mechanism hypothesis. MATERIAL AND METHODS: Four FNEs were retrieved from revision surgeries of four patients with prior intramedullary nail fixation of their pertrochanteric hip fractures complicated by femoral head perforation. The FNEs were divided into two groups based on whether or not there was radiographic evidence of medial migration prior to the revisions. Wear patterns on the FNEs were then assessed using both scanning electron microscopy and light microscopy. RESULTS: Repetitive, linearly-arranged, regularly-spaced, unique transverse scratch marks were found only in the group with medial migration, corresponding to the specific segment of the FNE that passed through the intramedullary component of the PFNA during medial migration. These scratch marks were absent in the group without medial migration. CONCLUSION: Our findings are in support of a ratcheting mechanism behind the medial migration phenomenon with repetitive toggling at the intramedullary nail-FNE interface and progressive propagation of the FNE against gravity.
format Online
Article
Text
id pubmed-7156168
institution National Center for Biotechnology Information
language English
publishDate 2020
publisher Malaysian Orthopaedic Association
record_format MEDLINE/PubMed
spelling pubmed-71561682020-04-15 Scanning Electron Microscopy Study of Retrieved Implants Suggests a Ratcheting Mechanism Behind Medial Migration in Cephalomedullary Nailing of Hip Fractures Law, GW Koh, JSB Yew, AKS Howe, TS Malays Orthop J Original Article INTRODUCTION: Medial migration is the paradoxical migration of the femoral neck element (FNE) superomedially against gravity with respect to the intramedullary component of the cephalomedullary device, increasingly seen in the management of pertrochanteric hip fractures with the intramedullary nail. We postulate that the peculiar anti-gravity movement of the FNE in the medial migration phenomenon stems from a ratcheting mechanism at the intramedullary nail-FNE interface, which should inadvertently produce unique wear patterns on the FNE that can be seen with high-powered microscopy. By examining the wear patterns on retrieved implants from patients with medial migration, our study aims to draw clinical correlations to the ratcheting mechanism hypothesis. MATERIAL AND METHODS: Four FNEs were retrieved from revision surgeries of four patients with prior intramedullary nail fixation of their pertrochanteric hip fractures complicated by femoral head perforation. The FNEs were divided into two groups based on whether or not there was radiographic evidence of medial migration prior to the revisions. Wear patterns on the FNEs were then assessed using both scanning electron microscopy and light microscopy. RESULTS: Repetitive, linearly-arranged, regularly-spaced, unique transverse scratch marks were found only in the group with medial migration, corresponding to the specific segment of the FNE that passed through the intramedullary component of the PFNA during medial migration. These scratch marks were absent in the group without medial migration. CONCLUSION: Our findings are in support of a ratcheting mechanism behind the medial migration phenomenon with repetitive toggling at the intramedullary nail-FNE interface and progressive propagation of the FNE against gravity. Malaysian Orthopaedic Association 2020-03 /pmc/articles/PMC7156168/ /pubmed/32296476 http://dx.doi.org/10.5704/MOJ.2003.002 Text en © 2020 Malaysian Orthopaedic Association (MOA). All Rights Reserved http://creativecommons.org/licenses/by/3.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited
spellingShingle Original Article
Law, GW
Koh, JSB
Yew, AKS
Howe, TS
Scanning Electron Microscopy Study of Retrieved Implants Suggests a Ratcheting Mechanism Behind Medial Migration in Cephalomedullary Nailing of Hip Fractures
title Scanning Electron Microscopy Study of Retrieved Implants Suggests a Ratcheting Mechanism Behind Medial Migration in Cephalomedullary Nailing of Hip Fractures
title_full Scanning Electron Microscopy Study of Retrieved Implants Suggests a Ratcheting Mechanism Behind Medial Migration in Cephalomedullary Nailing of Hip Fractures
title_fullStr Scanning Electron Microscopy Study of Retrieved Implants Suggests a Ratcheting Mechanism Behind Medial Migration in Cephalomedullary Nailing of Hip Fractures
title_full_unstemmed Scanning Electron Microscopy Study of Retrieved Implants Suggests a Ratcheting Mechanism Behind Medial Migration in Cephalomedullary Nailing of Hip Fractures
title_short Scanning Electron Microscopy Study of Retrieved Implants Suggests a Ratcheting Mechanism Behind Medial Migration in Cephalomedullary Nailing of Hip Fractures
title_sort scanning electron microscopy study of retrieved implants suggests a ratcheting mechanism behind medial migration in cephalomedullary nailing of hip fractures
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7156168/
https://www.ncbi.nlm.nih.gov/pubmed/32296476
http://dx.doi.org/10.5704/MOJ.2003.002
work_keys_str_mv AT lawgw scanningelectronmicroscopystudyofretrievedimplantssuggestsaratchetingmechanismbehindmedialmigrationincephalomedullarynailingofhipfractures
AT kohjsb scanningelectronmicroscopystudyofretrievedimplantssuggestsaratchetingmechanismbehindmedialmigrationincephalomedullarynailingofhipfractures
AT yewaks scanningelectronmicroscopystudyofretrievedimplantssuggestsaratchetingmechanismbehindmedialmigrationincephalomedullarynailingofhipfractures
AT howets scanningelectronmicroscopystudyofretrievedimplantssuggestsaratchetingmechanismbehindmedialmigrationincephalomedullarynailingofhipfractures